Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light

INTRODUCTION: Pseudoexfoliation syndrome (PEX) is a systemic disease involving the accumulation of pathological material deposits in the tissues of the anterior segment of the eye. The problem of modern ophthalmology is a quantitative assessment of the severity of PEX in the diagnosis and evaluation...

Descripción completa

Detalles Bibliográficos
Autores principales: Koprowski, Robert, Wróbel, Zygmunt, Nowinska, Anna, Wylegala, Edward
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3726400/
https://www.ncbi.nlm.nih.gov/pubmed/23879837
http://dx.doi.org/10.1186/1475-925X-12-74
_version_ 1782278633482616832
author Koprowski, Robert
Wróbel, Zygmunt
Nowinska, Anna
Wylegala, Edward
author_facet Koprowski, Robert
Wróbel, Zygmunt
Nowinska, Anna
Wylegala, Edward
author_sort Koprowski, Robert
collection PubMed
description INTRODUCTION: Pseudoexfoliation syndrome (PEX) is a systemic disease involving the accumulation of pathological material deposits in the tissues of the anterior segment of the eye. The problem of modern ophthalmology is a quantitative assessment of the severity of PEX in the diagnosis and evaluation of the treatment progress in patients. MATERIAL AND METHOD: For the purposes of this study, 52 images of the anterior segment of the eye with the resolution of M × N = 1280 × 960 pixels were obtained in jpg format using the slit lamp CSO 450-SL. The patients were aged 50–80 and were recruited from Poland. All patients who participated in the study provided written informed consent after explanation of the nature and possible consequences of the study. The image analysis method proposed by the authors contains the calculation of the direction field, setting a straight perpendicular line passing through each pixel of the edge of the pupil, the calculation of the intersection of straight lines in order to determine the central point of the pupil position, the detection of the contour of PEX and the outer border of the iris with the use of the polar coordinate system. All analyzed parameters were set automatically with one exception parameter chosen manually depending on the slit lamp type. RESULTS: A fully automatic measurement of PEX was carried out with the proposed method. Quantitative results enable to perform reproducible tests independently of the research centre. Owing to the image analysis method proposed by the authors, it is possible to obtain results in no more than 1 second on the Intel Core 2 Quad CPU 2.50 GHz with a measurement error below 3%. Other known methods of image analysis and processing that are compared in this paper give results with a greater error (4-35%) which depends on the degree of magnification (×6, ×16, ×20) and are not fully automatic. CONCLUSIONS: The methods of image analysis and processing enable a quantitative, repeatable and automatic measurement of the severity and progress of PEX syndrome. They support medical diagnosis and automatic archiving of results.
format Online
Article
Text
id pubmed-3726400
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-37264002013-07-31 Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light Koprowski, Robert Wróbel, Zygmunt Nowinska, Anna Wylegala, Edward Biomed Eng Online Research INTRODUCTION: Pseudoexfoliation syndrome (PEX) is a systemic disease involving the accumulation of pathological material deposits in the tissues of the anterior segment of the eye. The problem of modern ophthalmology is a quantitative assessment of the severity of PEX in the diagnosis and evaluation of the treatment progress in patients. MATERIAL AND METHOD: For the purposes of this study, 52 images of the anterior segment of the eye with the resolution of M × N = 1280 × 960 pixels were obtained in jpg format using the slit lamp CSO 450-SL. The patients were aged 50–80 and were recruited from Poland. All patients who participated in the study provided written informed consent after explanation of the nature and possible consequences of the study. The image analysis method proposed by the authors contains the calculation of the direction field, setting a straight perpendicular line passing through each pixel of the edge of the pupil, the calculation of the intersection of straight lines in order to determine the central point of the pupil position, the detection of the contour of PEX and the outer border of the iris with the use of the polar coordinate system. All analyzed parameters were set automatically with one exception parameter chosen manually depending on the slit lamp type. RESULTS: A fully automatic measurement of PEX was carried out with the proposed method. Quantitative results enable to perform reproducible tests independently of the research centre. Owing to the image analysis method proposed by the authors, it is possible to obtain results in no more than 1 second on the Intel Core 2 Quad CPU 2.50 GHz with a measurement error below 3%. Other known methods of image analysis and processing that are compared in this paper give results with a greater error (4-35%) which depends on the degree of magnification (×6, ×16, ×20) and are not fully automatic. CONCLUSIONS: The methods of image analysis and processing enable a quantitative, repeatable and automatic measurement of the severity and progress of PEX syndrome. They support medical diagnosis and automatic archiving of results. BioMed Central 2013-07-24 /pmc/articles/PMC3726400/ /pubmed/23879837 http://dx.doi.org/10.1186/1475-925X-12-74 Text en Copyright © 2013 Koprowski et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Koprowski, Robert
Wróbel, Zygmunt
Nowinska, Anna
Wylegala, Edward
Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title_full Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title_fullStr Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title_full_unstemmed Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title_short Quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
title_sort quantitative measurement of pseudoexfoliation in the anterior segment of the eye performed in visible light
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3726400/
https://www.ncbi.nlm.nih.gov/pubmed/23879837
http://dx.doi.org/10.1186/1475-925X-12-74
work_keys_str_mv AT koprowskirobert quantitativemeasurementofpseudoexfoliationintheanteriorsegmentoftheeyeperformedinvisiblelight
AT wrobelzygmunt quantitativemeasurementofpseudoexfoliationintheanteriorsegmentoftheeyeperformedinvisiblelight
AT nowinskaanna quantitativemeasurementofpseudoexfoliationintheanteriorsegmentoftheeyeperformedinvisiblelight
AT wylegalaedward quantitativemeasurementofpseudoexfoliationintheanteriorsegmentoftheeyeperformedinvisiblelight