Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method
BACKGROUND: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts. METHODS: The sensitivit...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3733653/ https://www.ncbi.nlm.nih.gov/pubmed/23875774 http://dx.doi.org/10.1186/1532-429X-15-63 |
Sumario: | BACKGROUND: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts. METHODS: The sensitivity of T1-estimates based on the SSFP based MOLLI sequence to errors in center frequency are calculated by means of a Bloch simulation and validated by phantom measurements. Typical off-resonance errors following local cardiac shimming are determined by field mapping at both 1.5 and 3.0T. In vivo examples demonstrate the artifactual appearance of T1-maps in the presence of off-resonance variation. RESULTS: Off-resonance varied 61.8 ± 15.5 Hz (mean ± SD, n = 18) across the heart at 1.5T and 125.0 ± 40.6 Hz (mean ± SD, n = 18) at 3.0T. For T1 = 1000 ms, the variation in T1 due to off-resonance variation was approximately 20 ms at 62 Hz, and > 50 ms at 125 Hz. CONCLUSIONS: Regional variations due to the inability to completely shim the B0-field variation around the heart appear as regional variation in T1, which is artifactual. |
---|