BACKGROUND: In our emergency department, we collect blood in Rapid Serum Tubes (RSTs; Becton Dickinson, Franklin Lakes, NJ), in which clotting times are reduced. We investigated the influence of RST use on cardiac troponin T (hs-cTnT) and creatine kinase-MB (CK-MB) test results, in comparison with t...

Descripción completa

Detalles Bibliográficos
Autores principales: Budak, Yasemin U, Huysal, Kagan, Bulut, Mehtap, Polat, Murat
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735479/
https://www.ncbi.nlm.nih.gov/pubmed/23763845
http://dx.doi.org/10.1186/1472-6890-13-20
_version_ 1782279651215802368
author Budak, Yasemin U
Huysal, Kagan
Bulut, Mehtap
Polat, Murat
author_facet Budak, Yasemin U
Huysal, Kagan
Bulut, Mehtap
Polat, Murat
author_sort Budak, Yasemin U
collection PubMed
description BACKGROUND: In our emergency department, we collect blood in Rapid Serum Tubes (RSTs; Becton Dickinson, Franklin Lakes, NJ), in which clotting times are reduced. We investigated the influence of RST use on cardiac troponin T (hs-cTnT) and creatine kinase-MB (CK-MB) test results, in comparison with the use of tubes featuring a separator gel containing a clotting activator (SSTs; Green-vac, Yongin, Korea). METHODS: Samples from 60 patients were divided into equal aliquots and placed into RSTs and SSTs; hs-cTnT and CK-MB concentrations were determined using an autoanalyzer (Elecsys 2010) running commercial assays (Roche Diagnostics, Penzberg, Germany). Between-tube differences in CK-MB and hs-cTnT values were compared using the paired t-test, and correlations among variables were evaluated by calculation of Spearman correlation coefficients (r values). Deming regression analysis was performed and Bland-Altman plots were constructed. RESULTS: The hs-cTnT and CK-MB test results obtained from samples placed into RSTs and SSTs did not differ (p > 0.1). The correlations between the concentrations of hs-cTnT and CK-MB in samples placed into RSTs and SSTs were good; both r values were unity (p < 0.001). Deming regression analysis yielded the equation: RST [hs-cTnT] = 0.98 SST [hs-cTnT] + 0.69 pg/ml; and RST [CK-MB] = 0.95 SST [CK-MB]–0.09 ng/ml. The biases of 1.4 pg/ml (95% CI: minus 8.1–10.7 pg/ml) for hs-cTnT levels and 0.249 ng/ml (95% CI: minus 0.682–1.681 ng/ml) for CK-MB levels assayed using either tube was acceptable. CONCLUSION: The hs-cTnT and CK-MB test results did not significantly differ when either tube was used. RST tube use was associated with a short clotting time; this was an advantage in an emergency laboratory setting.
format Online
Article
Text
id pubmed-3735479
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-37354792013-08-07 Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses Budak, Yasemin U Huysal, Kagan Bulut, Mehtap Polat, Murat BMC Clin Pathol Research Article BACKGROUND: In our emergency department, we collect blood in Rapid Serum Tubes (RSTs; Becton Dickinson, Franklin Lakes, NJ), in which clotting times are reduced. We investigated the influence of RST use on cardiac troponin T (hs-cTnT) and creatine kinase-MB (CK-MB) test results, in comparison with the use of tubes featuring a separator gel containing a clotting activator (SSTs; Green-vac, Yongin, Korea). METHODS: Samples from 60 patients were divided into equal aliquots and placed into RSTs and SSTs; hs-cTnT and CK-MB concentrations were determined using an autoanalyzer (Elecsys 2010) running commercial assays (Roche Diagnostics, Penzberg, Germany). Between-tube differences in CK-MB and hs-cTnT values were compared using the paired t-test, and correlations among variables were evaluated by calculation of Spearman correlation coefficients (r values). Deming regression analysis was performed and Bland-Altman plots were constructed. RESULTS: The hs-cTnT and CK-MB test results obtained from samples placed into RSTs and SSTs did not differ (p > 0.1). The correlations between the concentrations of hs-cTnT and CK-MB in samples placed into RSTs and SSTs were good; both r values were unity (p < 0.001). Deming regression analysis yielded the equation: RST [hs-cTnT] = 0.98 SST [hs-cTnT] + 0.69 pg/ml; and RST [CK-MB] = 0.95 SST [CK-MB]–0.09 ng/ml. The biases of 1.4 pg/ml (95% CI: minus 8.1–10.7 pg/ml) for hs-cTnT levels and 0.249 ng/ml (95% CI: minus 0.682–1.681 ng/ml) for CK-MB levels assayed using either tube was acceptable. CONCLUSION: The hs-cTnT and CK-MB test results did not significantly differ when either tube was used. RST tube use was associated with a short clotting time; this was an advantage in an emergency laboratory setting. BioMed Central 2013-06-13 /pmc/articles/PMC3735479/ /pubmed/23763845 http://dx.doi.org/10.1186/1472-6890-13-20 Text en Copyright © 2013 Budak et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Budak, Yasemin U
Huysal, Kagan
Bulut, Mehtap
Polat, Murat
Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title_full Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title_fullStr Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title_full_unstemmed Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title_short Evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-cTnT and CK-MB analyses
title_sort evaluation in an emergency department of rapid separator tubes containing thrombin for serum preparation prior to hs-ctnt and ck-mb analyses
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735479/
https://www.ncbi.nlm.nih.gov/pubmed/23763845
http://dx.doi.org/10.1186/1472-6890-13-20
work_keys_str_mv AT budakyaseminu evaluationinanemergencydepartmentofrapidseparatortubescontainingthrombinforserumpreparationpriortohsctntandckmbanalyses
AT huysalkagan evaluationinanemergencydepartmentofrapidseparatortubescontainingthrombinforserumpreparationpriortohsctntandckmbanalyses
AT bulutmehtap evaluationinanemergencydepartmentofrapidseparatortubescontainingthrombinforserumpreparationpriortohsctntandckmbanalyses
AT polatmurat evaluationinanemergencydepartmentofrapidseparatortubescontainingthrombinforserumpreparationpriortohsctntandckmbanalyses