Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia

The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using t...

Descripción completa

Detalles Bibliográficos
Autores principales: MENG, XIANGHU, ZANG, GUANGHUI, FAN, LONGCHANG, ZHENG, LEI, DAI, JINZHEN, WANG, XUEREN, XIA, WEI, LIU, JIHONG, ZHANG, CHUANHAN
Formato: Online Artículo Texto
Lenguaje:English
Publicado: D.A. Spandidos 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735910/
https://www.ncbi.nlm.nih.gov/pubmed/23935710
http://dx.doi.org/10.3892/etm.2013.1121
Descripción
Sumario:The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended.