Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using t...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
D.A. Spandidos
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735910/ https://www.ncbi.nlm.nih.gov/pubmed/23935710 http://dx.doi.org/10.3892/etm.2013.1121 |
_version_ | 1782279709147529216 |
---|---|
author | MENG, XIANGHU ZANG, GUANGHUI FAN, LONGCHANG ZHENG, LEI DAI, JINZHEN WANG, XUEREN XIA, WEI LIU, JIHONG ZHANG, CHUANHAN |
author_facet | MENG, XIANGHU ZANG, GUANGHUI FAN, LONGCHANG ZHENG, LEI DAI, JINZHEN WANG, XUEREN XIA, WEI LIU, JIHONG ZHANG, CHUANHAN |
author_sort | MENG, XIANGHU |
collection | PubMed |
description | The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended. |
format | Online Article Text |
id | pubmed-3735910 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | D.A. Spandidos |
record_format | MEDLINE/PubMed |
spelling | pubmed-37359102013-08-09 Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia MENG, XIANGHU ZANG, GUANGHUI FAN, LONGCHANG ZHENG, LEI DAI, JINZHEN WANG, XUEREN XIA, WEI LIU, JIHONG ZHANG, CHUANHAN Exp Ther Med Articles The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended. D.A. Spandidos 2013-07 2013-05-17 /pmc/articles/PMC3735910/ /pubmed/23935710 http://dx.doi.org/10.3892/etm.2013.1121 Text en Copyright © 2013, Spandidos Publications http://creativecommons.org/licenses/by/3.0 This is an open-access article licensed under a Creative Commons Attribution-NonCommercial 3.0 Unported License. The article may be redistributed, reproduced, and reused for non-commercial purposes, provided the original source is properly cited. |
spellingShingle | Articles MENG, XIANGHU ZANG, GUANGHUI FAN, LONGCHANG ZHENG, LEI DAI, JINZHEN WANG, XUEREN XIA, WEI LIU, JIHONG ZHANG, CHUANHAN Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title | Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title_full | Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title_fullStr | Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title_full_unstemmed | Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title_short | Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
title_sort | non-invasive monitoring of blood pressure using the philips intellivue mp50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia |
topic | Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735910/ https://www.ncbi.nlm.nih.gov/pubmed/23935710 http://dx.doi.org/10.3892/etm.2013.1121 |
work_keys_str_mv | AT mengxianghu noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT zangguanghui noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT fanlongchang noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT zhenglei noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT daijinzhen noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT wangxueren noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT xiawei noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT liujihong noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia AT zhangchuanhan noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia |