Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia

The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using t...

Descripción completa

Detalles Bibliográficos
Autores principales: MENG, XIANGHU, ZANG, GUANGHUI, FAN, LONGCHANG, ZHENG, LEI, DAI, JINZHEN, WANG, XUEREN, XIA, WEI, LIU, JIHONG, ZHANG, CHUANHAN
Formato: Online Artículo Texto
Lenguaje:English
Publicado: D.A. Spandidos 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735910/
https://www.ncbi.nlm.nih.gov/pubmed/23935710
http://dx.doi.org/10.3892/etm.2013.1121
_version_ 1782279709147529216
author MENG, XIANGHU
ZANG, GUANGHUI
FAN, LONGCHANG
ZHENG, LEI
DAI, JINZHEN
WANG, XUEREN
XIA, WEI
LIU, JIHONG
ZHANG, CHUANHAN
author_facet MENG, XIANGHU
ZANG, GUANGHUI
FAN, LONGCHANG
ZHENG, LEI
DAI, JINZHEN
WANG, XUEREN
XIA, WEI
LIU, JIHONG
ZHANG, CHUANHAN
author_sort MENG, XIANGHU
collection PubMed
description The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended.
format Online
Article
Text
id pubmed-3735910
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher D.A. Spandidos
record_format MEDLINE/PubMed
spelling pubmed-37359102013-08-09 Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia MENG, XIANGHU ZANG, GUANGHUI FAN, LONGCHANG ZHENG, LEI DAI, JINZHEN WANG, XUEREN XIA, WEI LIU, JIHONG ZHANG, CHUANHAN Exp Ther Med Articles The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended. D.A. Spandidos 2013-07 2013-05-17 /pmc/articles/PMC3735910/ /pubmed/23935710 http://dx.doi.org/10.3892/etm.2013.1121 Text en Copyright © 2013, Spandidos Publications http://creativecommons.org/licenses/by/3.0 This is an open-access article licensed under a Creative Commons Attribution-NonCommercial 3.0 Unported License. The article may be redistributed, reproduced, and reused for non-commercial purposes, provided the original source is properly cited.
spellingShingle Articles
MENG, XIANGHU
ZANG, GUANGHUI
FAN, LONGCHANG
ZHENG, LEI
DAI, JINZHEN
WANG, XUEREN
XIA, WEI
LIU, JIHONG
ZHANG, CHUANHAN
Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title_full Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title_fullStr Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title_full_unstemmed Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title_short Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
title_sort non-invasive monitoring of blood pressure using the philips intellivue mp50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3735910/
https://www.ncbi.nlm.nih.gov/pubmed/23935710
http://dx.doi.org/10.3892/etm.2013.1121
work_keys_str_mv AT mengxianghu noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT zangguanghui noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT fanlongchang noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT zhenglei noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT daijinzhen noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT wangxueren noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT xiawei noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT liujihong noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia
AT zhangchuanhan noninvasivemonitoringofbloodpressureusingthephilipsintellivuemp50monitorcannotreplaceinvasivebloodpressuretechniquesinsurgerypatientsundergeneralanesthesia