Cargando…

Outcome of Transcatheter Closure of Oval Shaped Atrial Septal Defect with Amplatzer Septal Occluder

PURPOSE: For the successful completion of transcatheter closure of atrial septal defects with the Amplatzer septal occluder, shape of the defects should be considered prior to selecting the device. The purpose of this study is to evaluate the results of a transcatheter closure of oval shaped atrial...

Descripción completa

Detalles Bibliográficos
Autores principales: Song, Jinyoung, Lee, Sang Yoon, Baek, Jae Sook, Shim, Woo Seub, Choi, Eun Young
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Yonsei University College of Medicine 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3743189/
https://www.ncbi.nlm.nih.gov/pubmed/23918558
http://dx.doi.org/10.3349/ymj.2013.54.5.1104
Descripción
Sumario:PURPOSE: For the successful completion of transcatheter closure of atrial septal defects with the Amplatzer septal occluder, shape of the defects should be considered prior to selecting the device. The purpose of this study is to evaluate the results of a transcatheter closure of oval shaped atrial septal defect. MATERIALS AND METHODS: From November 2009 until November 2011, cardiac computed tomography was performed on 69 patients who needed a transcatheter closure of atrial septal defect. We defined an oval shaped atrial septal defect as the ratio of the shortest diameter to the longest diameter ≤0.75 measured using computed tomography. A trans-thoracic echocardiogram was performed one day after and six months after. RESULTS: The transcatheter closure of atrial septal defect was performed successfully in 24 patients in the ovoid group and 45 patients in the circular group. There were no serious complications in both groups and the complete closure rate at 6 months later was 92.3% in the ovoid group and 93.1% in the circular group (p>0.05). The differences between the device size to the longest diameter of the defect and the ratios of the device size to the longest diameter were significantly smaller in the ovoid group (1.8±2.8 vs. 3.7±2.6 and 1.1±0.1 vs. 1.2±0.2). CONCLUSION: Transcatheter closure of an oval shaped atrial septal defect was found to be safe with the smaller Amplatzer septal occluder device when compared with circular atrial septal defects.