Cargando…

aSi-EPID transit signal calibration for dynamic beams: a needful step for the IMRT in vivo dosimetry

This work reports a method based on correlation functions to convert EPID transit signals into in vivo dose values at the isocenter point, D (iso), of dynamic IMRT beams supplied by Varian linac. Dose reconstruction for intensity-modulated beams required significant corrections of EPID response, due...

Descripción completa

Detalles Bibliográficos
Autores principales: Greco, Francesca, Piermattei, Angelo, Azario, Luigi, Placidi, Lorenzo, Cilla, Savino, Caivano, Rocchina, Fusco, Vincenzo, Fidanzio, Andrea
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3751319/
https://www.ncbi.nlm.nih.gov/pubmed/23835663
http://dx.doi.org/10.1007/s11517-013-1094-x
Descripción
Sumario:This work reports a method based on correlation functions to convert EPID transit signals into in vivo dose values at the isocenter point, D (iso), of dynamic IMRT beams supplied by Varian linac. Dose reconstruction for intensity-modulated beams required significant corrections of EPID response, due to the X-ray component transmitted through multileaf collimator. The algorithm was formulated using a set of simulated IMRT beams. The beams were parameterized by means of a fluence inhomogeneity index, FI, introduced to describe the degree of beam modulation with respect to open beams. This way, all dosimetric parameters involved in D (iso) reconstruction algorithm, such as the correlation functions, the correction factor for EPID to phantom distance and the modulated tissue maximum ratios, were determined as a function of the FI index. Clinical IMRT beams were used to irradiate a homogeneous phantom, and for each beam, the agreement between the reconstructed dose, D (iso), and the dose computed by TPS, D (iso,TPS), was well within 5 %. Moreover, the average ratios, R, between the D (iso), and D (iso,TPS), resulted equal to 1.002 ± 0.030. Thirty-five IMRT fields of 5 different patients undergoing radiotherapy for head–neck tumors were tested and the results were displayed on a computer screen after 2 min from the end of the treatment. However, 350 in vivo tests supplied an average ratio R equal to 1.004 ± 0.040. The in vivo dosimetry procedure here presented is among the objectives of a National Project financially supported by the Istituto Nazionale di Fisica Nucleare for the development of in vivo dosimetry procedures (Piermattei et al. in Nucl Instrum Methods Phys Res B 274:42–50, 2012) connected to the Record–Verify system of the radiotherapy center.