Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam

PURPOSE: The purpose of this study was to compare keratometry and anterior chamber depth (ACD) measurements from the Lenstar LS 900 (Haag-Streit AG, Switzerland) and the Pentacam (Oculus, Weltzar, Germany), and compare the keratometry readings of these two systems to a manual keratometer (MK), (Haag...

Descripción completa

Detalles Bibliográficos
Autores principales: Uçakhan, Ömür Ö., Akbel, Veysel, Bıyıklı, Zeynep, Kanpolat, Ayfer
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3757627/
https://www.ncbi.nlm.nih.gov/pubmed/24014981
http://dx.doi.org/10.4103/0974-9233.114791
Descripción
Sumario:PURPOSE: The purpose of this study was to compare keratometry and anterior chamber depth (ACD) measurements from the Lenstar LS 900 (Haag-Streit AG, Switzerland) and the Pentacam (Oculus, Weltzar, Germany), and compare the keratometry readings of these two systems to a manual keratometer (MK), (Haag-Streit, Switzerland). MATERIALS AND METHODS: In this prospective study, keratometry and ACD measurements were obtained in 50 eyes of 50 normal subjects with the Lenstar and the Pentacam. Keratometry was also measured using a MK. Correlation, comparison, and interdevice agreement was evaluated using intraclass correlation analysis, Wilcoxon test, and Bland-Altman plots. RESULTS: The keratometry and ACD measurements obtained from the Lenstar and the Pentacam showed excellent correlation. The mean interdevice differences in mean keratometry (Km) for the Lenstar and the Pentacam, the Lenstar and the MK, and the Pentacam and the MK were 0.39 D, 0.10 D, and 0.30 D respectively; and the 95% limits of agreement (LoA) for Km were 0.04-0.82 D; −1.90-2.10 D; and −2.30-1.70 D, respectively. The mean interdevice difference in ACD for the Lenstar versus the Pentacam was 0.09 mm, with 95% LoA of 0.23-0.05 mm. CONCLUSIONS: The ACD measurements obtained using the Lenstar and the Pentacam seem to be interchangeable, whereas, the keratometry measurements obtained using the Lenstar, Pentacam, and MK differ considerably, and are not interchangeable.