Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam

PURPOSE: The purpose of this study was to compare keratometry and anterior chamber depth (ACD) measurements from the Lenstar LS 900 (Haag-Streit AG, Switzerland) and the Pentacam (Oculus, Weltzar, Germany), and compare the keratometry readings of these two systems to a manual keratometer (MK), (Haag...

Descripción completa

Detalles Bibliográficos
Autores principales: Uçakhan, Ömür Ö., Akbel, Veysel, Bıyıklı, Zeynep, Kanpolat, Ayfer
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3757627/
https://www.ncbi.nlm.nih.gov/pubmed/24014981
http://dx.doi.org/10.4103/0974-9233.114791
_version_ 1782282238691377152
author Uçakhan, Ömür Ö.
Akbel, Veysel
Bıyıklı, Zeynep
Kanpolat, Ayfer
author_facet Uçakhan, Ömür Ö.
Akbel, Veysel
Bıyıklı, Zeynep
Kanpolat, Ayfer
author_sort Uçakhan, Ömür Ö.
collection PubMed
description PURPOSE: The purpose of this study was to compare keratometry and anterior chamber depth (ACD) measurements from the Lenstar LS 900 (Haag-Streit AG, Switzerland) and the Pentacam (Oculus, Weltzar, Germany), and compare the keratometry readings of these two systems to a manual keratometer (MK), (Haag-Streit, Switzerland). MATERIALS AND METHODS: In this prospective study, keratometry and ACD measurements were obtained in 50 eyes of 50 normal subjects with the Lenstar and the Pentacam. Keratometry was also measured using a MK. Correlation, comparison, and interdevice agreement was evaluated using intraclass correlation analysis, Wilcoxon test, and Bland-Altman plots. RESULTS: The keratometry and ACD measurements obtained from the Lenstar and the Pentacam showed excellent correlation. The mean interdevice differences in mean keratometry (Km) for the Lenstar and the Pentacam, the Lenstar and the MK, and the Pentacam and the MK were 0.39 D, 0.10 D, and 0.30 D respectively; and the 95% limits of agreement (LoA) for Km were 0.04-0.82 D; −1.90-2.10 D; and −2.30-1.70 D, respectively. The mean interdevice difference in ACD for the Lenstar versus the Pentacam was 0.09 mm, with 95% LoA of 0.23-0.05 mm. CONCLUSIONS: The ACD measurements obtained using the Lenstar and the Pentacam seem to be interchangeable, whereas, the keratometry measurements obtained using the Lenstar, Pentacam, and MK differ considerably, and are not interchangeable.
format Online
Article
Text
id pubmed-3757627
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-37576272013-09-06 Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam Uçakhan, Ömür Ö. Akbel, Veysel Bıyıklı, Zeynep Kanpolat, Ayfer Middle East Afr J Ophthalmol Original Article PURPOSE: The purpose of this study was to compare keratometry and anterior chamber depth (ACD) measurements from the Lenstar LS 900 (Haag-Streit AG, Switzerland) and the Pentacam (Oculus, Weltzar, Germany), and compare the keratometry readings of these two systems to a manual keratometer (MK), (Haag-Streit, Switzerland). MATERIALS AND METHODS: In this prospective study, keratometry and ACD measurements were obtained in 50 eyes of 50 normal subjects with the Lenstar and the Pentacam. Keratometry was also measured using a MK. Correlation, comparison, and interdevice agreement was evaluated using intraclass correlation analysis, Wilcoxon test, and Bland-Altman plots. RESULTS: The keratometry and ACD measurements obtained from the Lenstar and the Pentacam showed excellent correlation. The mean interdevice differences in mean keratometry (Km) for the Lenstar and the Pentacam, the Lenstar and the MK, and the Pentacam and the MK were 0.39 D, 0.10 D, and 0.30 D respectively; and the 95% limits of agreement (LoA) for Km were 0.04-0.82 D; −1.90-2.10 D; and −2.30-1.70 D, respectively. The mean interdevice difference in ACD for the Lenstar versus the Pentacam was 0.09 mm, with 95% LoA of 0.23-0.05 mm. CONCLUSIONS: The ACD measurements obtained using the Lenstar and the Pentacam seem to be interchangeable, whereas, the keratometry measurements obtained using the Lenstar, Pentacam, and MK differ considerably, and are not interchangeable. Medknow Publications & Media Pvt Ltd 2013 /pmc/articles/PMC3757627/ /pubmed/24014981 http://dx.doi.org/10.4103/0974-9233.114791 Text en Copyright: © Middle East African Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Uçakhan, Ömür Ö.
Akbel, Veysel
Bıyıklı, Zeynep
Kanpolat, Ayfer
Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title_full Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title_fullStr Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title_full_unstemmed Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title_short Comparison of Corneal Curvature and Anterior Chamber Depth Measurements Using the Manual Keratometer, Lenstar LS 900 and the Pentacam
title_sort comparison of corneal curvature and anterior chamber depth measurements using the manual keratometer, lenstar ls 900 and the pentacam
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3757627/
https://www.ncbi.nlm.nih.gov/pubmed/24014981
http://dx.doi.org/10.4103/0974-9233.114791
work_keys_str_mv AT ucakhanomuro comparisonofcornealcurvatureandanteriorchamberdepthmeasurementsusingthemanualkeratometerlenstarls900andthepentacam
AT akbelveysel comparisonofcornealcurvatureandanteriorchamberdepthmeasurementsusingthemanualkeratometerlenstarls900andthepentacam
AT bıyıklızeynep comparisonofcornealcurvatureandanteriorchamberdepthmeasurementsusingthemanualkeratometerlenstarls900andthepentacam
AT kanpolatayfer comparisonofcornealcurvatureandanteriorchamberdepthmeasurementsusingthemanualkeratometerlenstarls900andthepentacam