Good accuracy of the Phase III Oxford Mobile Bearing Unicompartmental Knee Instrumentation

BACKGROUND AND PURPOSE: — Unicompartmental knee arthroplasty (UKA) needs careful balancing of flexion/extension (F/E) gaps to prevent dislocation of the mobile meniscal bearing. Assessment of gaps is based on the surgeon’s subjective insertion force of a feeler gauge with different thicknesses and/o...

Descripción completa

Detalles Bibliográficos
Autores principales: Bitsch, Rudi G, von Keudell, Arvind, Losina, Elena, Fitz, Wolfgang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Informa Healthcare 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3768042/
https://www.ncbi.nlm.nih.gov/pubmed/23799349
http://dx.doi.org/10.3109/17453674.2013.813801
Descripción
Sumario:BACKGROUND AND PURPOSE: — Unicompartmental knee arthroplasty (UKA) needs careful balancing of flexion/extension (F/E) gaps to prevent dislocation of the mobile meniscal bearing. Assessment of gaps is based on the surgeon’s subjective insertion force of a feeler gauge with different thicknesses and/or the lift-off of a trial meniscal bearing. However, the accuracy of this method remains unclear. We assessed the accuracy of the technique. PATIENTS AND METHODS: — A consecutive series of 33 UKAs in 32 patients (mean age 64 years, 24 women) were balanced using the Oxford Phase III (OP III) Instrumentation. The recommended technique for F/E gap assessment was performed using different feeler gauges with 1-mm increments and the meniscal bearing lift-off tests according to surgical technique. A tensiometer was inserted and both gaps were maximally distracted by hand. Measurements in mm were recorded and analyzed with a reading of 90 N for both gaps in 20 and 90 degrees of flexion. RESULTS: — The gaps measured were 12 (11–18) mm in extension and 13 (11–18) mm in 90 degrees of flexion. The difference between the gaps was 0.4 (–0.5 to 1.0) mm (p < 0.001). There were no statistically significant gender differences regarding composite implant thickness, laxity, flexion gap, extension gap, or gap difference. INTERPRETATION: — OP III instrumentation using feeler gauges and the lift-off test provides accurate balancing of F/E gaps with an accuracy of less than 1 mm.