Cargando…

Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging

While channel-cut crystals, in which the diffracting surfaces in an asymmetric cut are kept parallel, can provide beam collimation and spectral beam shaping, they can in addition provide beam compression or expansion if the cut is V-shaped. The compression/expansion ratio depends in this case on the...

Descripción completa

Detalles Bibliográficos
Autores principales: Korytár, D., Vagovič, P., Végsö, K., Šiffalovič, P., Dobročka, E., Jark, W., Áč, V., Zápražný, Z., Ferrari, C., Cecilia, A., Hamann, E., Mikulík, P., Baumbach, T., Fiederle, M., Jergel, M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Union of Crystallography 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3769071/
https://www.ncbi.nlm.nih.gov/pubmed/24046503
http://dx.doi.org/10.1107/S0021889813006122
_version_ 1782283934600527872
author Korytár, D.
Vagovič, P.
Végsö, K.
Šiffalovič, P.
Dobročka, E.
Jark, W.
Áč, V.
Zápražný, Z.
Ferrari, C.
Cecilia, A.
Hamann, E.
Mikulík, P.
Baumbach, T.
Fiederle, M.
Jergel, M.
author_facet Korytár, D.
Vagovič, P.
Végsö, K.
Šiffalovič, P.
Dobročka, E.
Jark, W.
Áč, V.
Zápražný, Z.
Ferrari, C.
Cecilia, A.
Hamann, E.
Mikulík, P.
Baumbach, T.
Fiederle, M.
Jergel, M.
author_sort Korytár, D.
collection PubMed
description While channel-cut crystals, in which the diffracting surfaces in an asymmetric cut are kept parallel, can provide beam collimation and spectral beam shaping, they can in addition provide beam compression or expansion if the cut is V-shaped. The compression/expansion ratio depends in this case on the total asymmetry factor. If the Ge(220) diffraction planes and a total asymmetry factor in excess of 10 are used, the rocking curves of two diffractors will have a sufficient overlap only if the second diffractor is tuned slightly with respect to the first one. This study compares and analyses several ways of overcoming this mismatch, which is due to refraction, when the Cu Kα(1) beam is compressed 21-fold in a V21 monochromator. A more than sixfold intensity increase was obtained if the matching was improved either by a compositional variation or by a thermal deformation. This provided an intensity gain compared with the use of a simple slit in a symmetrical channel-cut monochromator. The first attempt to overcome the mismatch by introducing different types of X-ray prisms for the required beam deflection is described as well. The performance of the V-shaped monochromators is demonstrated in two applications. A narrow collimated monochromatic beam obtained in the beam compressing mode was used for high-resolution grazing-incidence small-angle X-ray scattering measurements of a silicon sample with corrupted surface. In addition, a two-dimensional Bragg magnifier, based on two crossed V15 channel monochromators in beam expansion mode and tuned by means of unequal asymmetries, was successfully applied to high-resolution imaging of test structures in combination with a Medipix detector.
format Online
Article
Text
id pubmed-3769071
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher International Union of Crystallography
record_format MEDLINE/PubMed
spelling pubmed-37690712013-09-17 Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging Korytár, D. Vagovič, P. Végsö, K. Šiffalovič, P. Dobročka, E. Jark, W. Áč, V. Zápražný, Z. Ferrari, C. Cecilia, A. Hamann, E. Mikulík, P. Baumbach, T. Fiederle, M. Jergel, M. J Appl Crystallogr X-Ray Diffraction and Imaging While channel-cut crystals, in which the diffracting surfaces in an asymmetric cut are kept parallel, can provide beam collimation and spectral beam shaping, they can in addition provide beam compression or expansion if the cut is V-shaped. The compression/expansion ratio depends in this case on the total asymmetry factor. If the Ge(220) diffraction planes and a total asymmetry factor in excess of 10 are used, the rocking curves of two diffractors will have a sufficient overlap only if the second diffractor is tuned slightly with respect to the first one. This study compares and analyses several ways of overcoming this mismatch, which is due to refraction, when the Cu Kα(1) beam is compressed 21-fold in a V21 monochromator. A more than sixfold intensity increase was obtained if the matching was improved either by a compositional variation or by a thermal deformation. This provided an intensity gain compared with the use of a simple slit in a symmetrical channel-cut monochromator. The first attempt to overcome the mismatch by introducing different types of X-ray prisms for the required beam deflection is described as well. The performance of the V-shaped monochromators is demonstrated in two applications. A narrow collimated monochromatic beam obtained in the beam compressing mode was used for high-resolution grazing-incidence small-angle X-ray scattering measurements of a silicon sample with corrupted surface. In addition, a two-dimensional Bragg magnifier, based on two crossed V15 channel monochromators in beam expansion mode and tuned by means of unequal asymmetries, was successfully applied to high-resolution imaging of test structures in combination with a Medipix detector. International Union of Crystallography 2013-08-01 2013-06-07 /pmc/articles/PMC3769071/ /pubmed/24046503 http://dx.doi.org/10.1107/S0021889813006122 Text en © D. Korytár et al. 2013 http://creativecommons.org/licenses/by/2.0/uk/ This is an open-access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.
spellingShingle X-Ray Diffraction and Imaging
Korytár, D.
Vagovič, P.
Végsö, K.
Šiffalovič, P.
Dobročka, E.
Jark, W.
Áč, V.
Zápražný, Z.
Ferrari, C.
Cecilia, A.
Hamann, E.
Mikulík, P.
Baumbach, T.
Fiederle, M.
Jergel, M.
Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title_full Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title_fullStr Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title_full_unstemmed Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title_short Potential use of V-channel Ge(220) monochromators in X-ray metrology and imaging
title_sort potential use of v-channel ge(220) monochromators in x-ray metrology and imaging
topic X-Ray Diffraction and Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3769071/
https://www.ncbi.nlm.nih.gov/pubmed/24046503
http://dx.doi.org/10.1107/S0021889813006122
work_keys_str_mv AT korytard potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT vagovicp potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT vegsok potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT siffalovicp potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT dobrockae potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT jarkw potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT acv potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT zapraznyz potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT ferraric potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT ceciliaa potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT hamanne potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT mikulikp potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT baumbacht potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT fiederlem potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging
AT jergelm potentialuseofvchannelge220monochromatorsinxraymetrologyandimaging