New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients

A pilot study has been conducted to validate the Breath Motion Detecting System (BMDS), a new concept using Passive Infrared (PIR) technology for a contactless detection of respiratory movements. The primary objective of the study was to show if movements detected during sleep by the BMDS were indee...

Descripción completa

Detalles Bibliográficos
Autores principales: Hers, V., Corbugy, D., Joslet, I., Hermant, P., Demarteau, J., Delhougne, B., Vandermoten, G., Hermanne, J. P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3778891/
https://www.ncbi.nlm.nih.gov/pubmed/23549646
http://dx.doi.org/10.1007/s10877-013-9457-2
_version_ 1782285178962444288
author Hers, V.
Corbugy, D.
Joslet, I.
Hermant, P.
Demarteau, J.
Delhougne, B.
Vandermoten, G.
Hermanne, J. P.
author_facet Hers, V.
Corbugy, D.
Joslet, I.
Hermant, P.
Demarteau, J.
Delhougne, B.
Vandermoten, G.
Hermanne, J. P.
author_sort Hers, V.
collection PubMed
description A pilot study has been conducted to validate the Breath Motion Detecting System (BMDS), a new concept using Passive Infrared (PIR) technology for a contactless detection of respiratory movements. The primary objective of the study was to show if movements detected during sleep by the BMDS were indeed related to breathing. This medical device is not intended to measure the respiratory rate, but in a second step, it will be able to detect pathological central apnea in adults. One hundred and sixty-nine adult patients underwent a full polysomnography in which each respiratory movement was recorded concomitantly through the BMDS. Curves obtained by the BMDS were compared to those of thoracic movements recorded by classical piezoelectric belts and of pressure obtained with nasal cannula. The correlations between the PIR sensors were highly indicative of respiratory movement detection. Since PIR sensors are sensitive only to the exemplification of the rib cage, they did not detect obstructive apnea. Unfortunately, only a few patients in the studied population had a central apnea. Moreover as our sleep laboratory was equipped only with piezoelectric bands, the central apnea respiratory effort data are not a validated signal to be used during sleep recordings. The data recorded by the BMDS demonstrate the ability of the PIR technology to detect respiratory movements in adults. The concept is practical, inexpensive and safe for the patient. Further studies with respiratory inductive plethysmography are needed to investigate the potential of BMDS to detect central apneas.
format Online
Article
Text
id pubmed-3778891
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-37788912013-09-25 New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients Hers, V. Corbugy, D. Joslet, I. Hermant, P. Demarteau, J. Delhougne, B. Vandermoten, G. Hermanne, J. P. J Clin Monit Comput Original Research A pilot study has been conducted to validate the Breath Motion Detecting System (BMDS), a new concept using Passive Infrared (PIR) technology for a contactless detection of respiratory movements. The primary objective of the study was to show if movements detected during sleep by the BMDS were indeed related to breathing. This medical device is not intended to measure the respiratory rate, but in a second step, it will be able to detect pathological central apnea in adults. One hundred and sixty-nine adult patients underwent a full polysomnography in which each respiratory movement was recorded concomitantly through the BMDS. Curves obtained by the BMDS were compared to those of thoracic movements recorded by classical piezoelectric belts and of pressure obtained with nasal cannula. The correlations between the PIR sensors were highly indicative of respiratory movement detection. Since PIR sensors are sensitive only to the exemplification of the rib cage, they did not detect obstructive apnea. Unfortunately, only a few patients in the studied population had a central apnea. Moreover as our sleep laboratory was equipped only with piezoelectric bands, the central apnea respiratory effort data are not a validated signal to be used during sleep recordings. The data recorded by the BMDS demonstrate the ability of the PIR technology to detect respiratory movements in adults. The concept is practical, inexpensive and safe for the patient. Further studies with respiratory inductive plethysmography are needed to investigate the potential of BMDS to detect central apneas. Springer Netherlands 2013-04-03 2013 /pmc/articles/PMC3778891/ /pubmed/23549646 http://dx.doi.org/10.1007/s10877-013-9457-2 Text en © The Author(s) 2013 https://creativecommons.org/licenses/by/2.0/ Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
spellingShingle Original Research
Hers, V.
Corbugy, D.
Joslet, I.
Hermant, P.
Demarteau, J.
Delhougne, B.
Vandermoten, G.
Hermanne, J. P.
New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title_full New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title_fullStr New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title_full_unstemmed New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title_short New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
title_sort new concept using passive infrared (pir) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3778891/
https://www.ncbi.nlm.nih.gov/pubmed/23549646
http://dx.doi.org/10.1007/s10877-013-9457-2
work_keys_str_mv AT hersv newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT corbugyd newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT josleti newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT hermantp newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT demarteauj newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT delhougneb newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT vandermoteng newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients
AT hermannejp newconceptusingpassiveinfraredpirtechnologyforacontactlessdetectionofbreathingmovementapilotstudyinvolvingacohortof169adultpatients