Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium,...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medicina Oral S.L.
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3790656/ https://www.ncbi.nlm.nih.gov/pubmed/23722125 http://dx.doi.org/10.4317/medoral.18674 |
_version_ | 1782286618829258752 |
---|---|
author | Walter, André Winsauer, Heinz Marcé-Nogué, Jordi Mojal, Sergi Puigdollers, Andreu |
author_facet | Walter, André Winsauer, Heinz Marcé-Nogué, Jordi Mojal, Sergi Puigdollers, Andreu |
author_sort | Walter, André |
collection | PubMed |
description | Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium, 4 large) from 8 manufacturers were tested for their primary stability in simulated medium-high cancellous bone and the risk to fracture in high-density methacrylate blocks. For the assessments of the maximum insertion torque (IT) and torsional fracture (TF) 5 of each OMI were used and for the pull-out strength (POS) 10. The OMIs were inserted with a torque screwdriver (12 sec/360°) until the bottom at 8 mm depth was reached. OMI designs were analyzed with a scan electron microscope (SEM). Results: SEM images revealed a great variation in product refinement. In the whole sample, a cylindrical OMI shape was associated with higher POS (p<0.001) but lower IT (p=0.002) values. The outer and inner OMI diameters were design characteristics well correlated with POS, IT and TF values (ranging from 0.601 to 0.961). Greater thread depth was related to greater POS values (r= 0.628), although OMIs with similar POS values may have different IT values. Thread depth and pitch had some impact on POS. TF depended mainly on the OMI inner (r= 0.961) and outer diameters (r=0.892). A thread depth to outer diameter ratio close to 40% increased TF risk. Conclusions: Although at the same insertion depth the OMI outer and inner diameters are the most important factors for primary stability, other OMI design characteristics (cylindrical vs. conical, thread design) may significantly affect primary stability and torsional fracture. This needs to be considered when selecting the appropriate OMI for the desired orthodontic procedures. Key words:Orthodontic mini-implants, primary stability, insertion torque, pullout strength, torsional fracture. |
format | Online Article Text |
id | pubmed-3790656 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Medicina Oral S.L. |
record_format | MEDLINE/PubMed |
spelling | pubmed-37906562013-10-21 Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies Walter, André Winsauer, Heinz Marcé-Nogué, Jordi Mojal, Sergi Puigdollers, Andreu Med Oral Patol Oral Cir Bucal Research-Article Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium, 4 large) from 8 manufacturers were tested for their primary stability in simulated medium-high cancellous bone and the risk to fracture in high-density methacrylate blocks. For the assessments of the maximum insertion torque (IT) and torsional fracture (TF) 5 of each OMI were used and for the pull-out strength (POS) 10. The OMIs were inserted with a torque screwdriver (12 sec/360°) until the bottom at 8 mm depth was reached. OMI designs were analyzed with a scan electron microscope (SEM). Results: SEM images revealed a great variation in product refinement. In the whole sample, a cylindrical OMI shape was associated with higher POS (p<0.001) but lower IT (p=0.002) values. The outer and inner OMI diameters were design characteristics well correlated with POS, IT and TF values (ranging from 0.601 to 0.961). Greater thread depth was related to greater POS values (r= 0.628), although OMIs with similar POS values may have different IT values. Thread depth and pitch had some impact on POS. TF depended mainly on the OMI inner (r= 0.961) and outer diameters (r=0.892). A thread depth to outer diameter ratio close to 40% increased TF risk. Conclusions: Although at the same insertion depth the OMI outer and inner diameters are the most important factors for primary stability, other OMI design characteristics (cylindrical vs. conical, thread design) may significantly affect primary stability and torsional fracture. This needs to be considered when selecting the appropriate OMI for the desired orthodontic procedures. Key words:Orthodontic mini-implants, primary stability, insertion torque, pullout strength, torsional fracture. Medicina Oral S.L. 2013-09 2013-05-31 /pmc/articles/PMC3790656/ /pubmed/23722125 http://dx.doi.org/10.4317/medoral.18674 Text en Copyright: © 2013 Medicina Oral S.L. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research-Article Walter, André Winsauer, Heinz Marcé-Nogué, Jordi Mojal, Sergi Puigdollers, Andreu Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title | Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title_full | Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title_fullStr | Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title_full_unstemmed | Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title_short | Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies |
title_sort | design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies |
topic | Research-Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3790656/ https://www.ncbi.nlm.nih.gov/pubmed/23722125 http://dx.doi.org/10.4317/medoral.18674 |
work_keys_str_mv | AT walterandre designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies AT winsauerheinz designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies AT marcenoguejordi designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies AT mojalsergi designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies AT puigdollersandreu designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies |