Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies

Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium,...

Descripción completa

Detalles Bibliográficos
Autores principales: Walter, André, Winsauer, Heinz, Marcé-Nogué, Jordi, Mojal, Sergi, Puigdollers, Andreu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medicina Oral S.L. 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3790656/
https://www.ncbi.nlm.nih.gov/pubmed/23722125
http://dx.doi.org/10.4317/medoral.18674
_version_ 1782286618829258752
author Walter, André
Winsauer, Heinz
Marcé-Nogué, Jordi
Mojal, Sergi
Puigdollers, Andreu
author_facet Walter, André
Winsauer, Heinz
Marcé-Nogué, Jordi
Mojal, Sergi
Puigdollers, Andreu
author_sort Walter, André
collection PubMed
description Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium, 4 large) from 8 manufacturers were tested for their primary stability in simulated medium-high cancellous bone and the risk to fracture in high-density methacrylate blocks. For the assessments of the maximum insertion torque (IT) and torsional fracture (TF) 5 of each OMI were used and for the pull-out strength (POS) 10. The OMIs were inserted with a torque screwdriver (12 sec/360°) until the bottom at 8 mm depth was reached. OMI designs were analyzed with a scan electron microscope (SEM). Results: SEM images revealed a great variation in product refinement. In the whole sample, a cylindrical OMI shape was associated with higher POS (p<0.001) but lower IT (p=0.002) values. The outer and inner OMI diameters were design characteristics well correlated with POS, IT and TF values (ranging from 0.601 to 0.961). Greater thread depth was related to greater POS values (r= 0.628), although OMIs with similar POS values may have different IT values. Thread depth and pitch had some impact on POS. TF depended mainly on the OMI inner (r= 0.961) and outer diameters (r=0.892). A thread depth to outer diameter ratio close to 40% increased TF risk. Conclusions: Although at the same insertion depth the OMI outer and inner diameters are the most important factors for primary stability, other OMI design characteristics (cylindrical vs. conical, thread design) may significantly affect primary stability and torsional fracture. This needs to be considered when selecting the appropriate OMI for the desired orthodontic procedures. Key words:Orthodontic mini-implants, primary stability, insertion torque, pullout strength, torsional fracture.
format Online
Article
Text
id pubmed-3790656
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Medicina Oral S.L.
record_format MEDLINE/PubMed
spelling pubmed-37906562013-10-21 Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies Walter, André Winsauer, Heinz Marcé-Nogué, Jordi Mojal, Sergi Puigdollers, Andreu Med Oral Patol Oral Cir Bucal Research-Article Objectives: Orthodontic mini-implants (OMIs) are increasingly used in orthodontics but can fail for various reasons. This study investigates the effects of OMI design characteristics on the mechanical properties in artificial bone. Material and Methods: Twelve self-drilling OMIs (2 small, 6 medium, 4 large) from 8 manufacturers were tested for their primary stability in simulated medium-high cancellous bone and the risk to fracture in high-density methacrylate blocks. For the assessments of the maximum insertion torque (IT) and torsional fracture (TF) 5 of each OMI were used and for the pull-out strength (POS) 10. The OMIs were inserted with a torque screwdriver (12 sec/360°) until the bottom at 8 mm depth was reached. OMI designs were analyzed with a scan electron microscope (SEM). Results: SEM images revealed a great variation in product refinement. In the whole sample, a cylindrical OMI shape was associated with higher POS (p<0.001) but lower IT (p=0.002) values. The outer and inner OMI diameters were design characteristics well correlated with POS, IT and TF values (ranging from 0.601 to 0.961). Greater thread depth was related to greater POS values (r= 0.628), although OMIs with similar POS values may have different IT values. Thread depth and pitch had some impact on POS. TF depended mainly on the OMI inner (r= 0.961) and outer diameters (r=0.892). A thread depth to outer diameter ratio close to 40% increased TF risk. Conclusions: Although at the same insertion depth the OMI outer and inner diameters are the most important factors for primary stability, other OMI design characteristics (cylindrical vs. conical, thread design) may significantly affect primary stability and torsional fracture. This needs to be considered when selecting the appropriate OMI for the desired orthodontic procedures. Key words:Orthodontic mini-implants, primary stability, insertion torque, pullout strength, torsional fracture. Medicina Oral S.L. 2013-09 2013-05-31 /pmc/articles/PMC3790656/ /pubmed/23722125 http://dx.doi.org/10.4317/medoral.18674 Text en Copyright: © 2013 Medicina Oral S.L. http://creativecommons.org/licenses/by/2.5/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research-Article
Walter, André
Winsauer, Heinz
Marcé-Nogué, Jordi
Mojal, Sergi
Puigdollers, Andreu
Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title_full Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title_fullStr Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title_full_unstemmed Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title_short Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: Pilot scan electron microscope and mechanical studies
title_sort design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies
topic Research-Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3790656/
https://www.ncbi.nlm.nih.gov/pubmed/23722125
http://dx.doi.org/10.4317/medoral.18674
work_keys_str_mv AT walterandre designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies
AT winsauerheinz designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies
AT marcenoguejordi designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies
AT mojalsergi designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies
AT puigdollersandreu designcharacteristicsprimarystabilityandriskoffractureoforthodonticminiimplantspilotscanelectronmicroscopeandmechanicalstudies