Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?

PURPOSE: The aim of the study was to verify dose distribution parameters for multi-lumen, and artificially created single-lumen balloon applicator used for the same patient with two optimization algorithms: inverse planning simulated annealing (IPSA) and dose point optimization with distance option....

Descripción completa

Detalles Bibliográficos
Autores principales: Skowronek, Janusz, Bielęda, Grzegorz, Łaski, Piotr, Kycler, Witold
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3797409/
https://www.ncbi.nlm.nih.gov/pubmed/24143147
http://dx.doi.org/10.5114/jcb.2013.37776
_version_ 1782287610403618816
author Skowronek, Janusz
Bielęda, Grzegorz
Łaski, Piotr
Kycler, Witold
author_facet Skowronek, Janusz
Bielęda, Grzegorz
Łaski, Piotr
Kycler, Witold
author_sort Skowronek, Janusz
collection PubMed
description PURPOSE: The aim of the study was to verify dose distribution parameters for multi-lumen, and artificially created single-lumen balloon applicator used for the same patient with two optimization algorithms: inverse planning simulated annealing (IPSA) and dose point optimization with distance option. MATERIAL AND METHODS: Group of 24 patients with multi-lumen balloon applied were investigated. Each patient received 10 fractions of 3.4 Gy (2 fractions daily). For every patient, four treatment plans were prepared. Firstly, for five-lumen balloon optimized with IPSA algorithm and optimization parameters adjusted for each case. Secondly, for the same applicator optimized with dose point optimization and distant option. Two other plans were prepared for single-lumen applicator, created by removing four peripheral lumens, optimized with both algorithms. RESULTS: The highest D(95) parameter was obtained for plans optimized with IPSA algorithm, mean value 99.3 percent of prescribed dose, and it was significantly higher than plans optimized with dose point algorithm (mean = 83.50%, p < 0.0001), IPSA single-lumen balloon plan (mean = 83.50%, p = 0.0037) and optimized to dose point single-lumen balloon (mean = 85.51%, p < 0.0001). There were no statistically significant differences concerning maximum doses distributed to skin surface for neither application nor optimization method. Volumes receiving 200% of prescribed dose in PTV were higher for multi-lumen balloon dose point optimized plans (mean = 8.78%), than for other plans (IPSA multi-lumen balloon plan: mean = 7.37%, p < 0.0001, single-lumen IPSA: mean = 7.20%, p < 0.0001, single-lumen dose point: mean = 7.19%, p < 0.0001). CONCLUSIONS: Basing on performed survey, better dose distribution parameters are obtained for patients with multi-lumen balloon applied and optimized using IPSA algorithm with individualized optimization parameters.
format Online
Article
Text
id pubmed-3797409
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-37974092013-10-18 Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation? Skowronek, Janusz Bielęda, Grzegorz Łaski, Piotr Kycler, Witold J Contemp Brachytherapy Original Paper PURPOSE: The aim of the study was to verify dose distribution parameters for multi-lumen, and artificially created single-lumen balloon applicator used for the same patient with two optimization algorithms: inverse planning simulated annealing (IPSA) and dose point optimization with distance option. MATERIAL AND METHODS: Group of 24 patients with multi-lumen balloon applied were investigated. Each patient received 10 fractions of 3.4 Gy (2 fractions daily). For every patient, four treatment plans were prepared. Firstly, for five-lumen balloon optimized with IPSA algorithm and optimization parameters adjusted for each case. Secondly, for the same applicator optimized with dose point optimization and distant option. Two other plans were prepared for single-lumen applicator, created by removing four peripheral lumens, optimized with both algorithms. RESULTS: The highest D(95) parameter was obtained for plans optimized with IPSA algorithm, mean value 99.3 percent of prescribed dose, and it was significantly higher than plans optimized with dose point algorithm (mean = 83.50%, p < 0.0001), IPSA single-lumen balloon plan (mean = 83.50%, p = 0.0037) and optimized to dose point single-lumen balloon (mean = 85.51%, p < 0.0001). There were no statistically significant differences concerning maximum doses distributed to skin surface for neither application nor optimization method. Volumes receiving 200% of prescribed dose in PTV were higher for multi-lumen balloon dose point optimized plans (mean = 8.78%), than for other plans (IPSA multi-lumen balloon plan: mean = 7.37%, p < 0.0001, single-lumen IPSA: mean = 7.20%, p < 0.0001, single-lumen dose point: mean = 7.19%, p < 0.0001). CONCLUSIONS: Basing on performed survey, better dose distribution parameters are obtained for patients with multi-lumen balloon applied and optimized using IPSA algorithm with individualized optimization parameters. Termedia Publishing House 2013-09-20 2013-09 /pmc/articles/PMC3797409/ /pubmed/24143147 http://dx.doi.org/10.5114/jcb.2013.37776 Text en Copyright © 2013 Termedia http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial 3.0 Unported License, permitting all non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Paper
Skowronek, Janusz
Bielęda, Grzegorz
Łaski, Piotr
Kycler, Witold
Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title_full Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title_fullStr Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title_full_unstemmed Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title_short Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?
title_sort can we improve the dose distribution for single or multi-lumen breast balloons used for accelerated partial breast irradiation?
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3797409/
https://www.ncbi.nlm.nih.gov/pubmed/24143147
http://dx.doi.org/10.5114/jcb.2013.37776
work_keys_str_mv AT skowronekjanusz canweimprovethedosedistributionforsingleormultilumenbreastballoonsusedforacceleratedpartialbreastirradiation
AT bieledagrzegorz canweimprovethedosedistributionforsingleormultilumenbreastballoonsusedforacceleratedpartialbreastirradiation
AT łaskipiotr canweimprovethedosedistributionforsingleormultilumenbreastballoonsusedforacceleratedpartialbreastirradiation
AT kyclerwitold canweimprovethedosedistributionforsingleormultilumenbreastballoonsusedforacceleratedpartialbreastirradiation