Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength
PURPOSE: Cervical disc prostheses induce significant amount of artifact in magnetic resonance imaging which may complicate radiologic follow-up after surgery. The purpose of this study was to investigate as to what extent the artifact, induced by the frequently used Discover(®) cervical disc prosthe...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3804698/ https://www.ncbi.nlm.nih.gov/pubmed/24061966 http://dx.doi.org/10.1007/s00586-013-2994-z |
_version_ | 1782288201415655424 |
---|---|
author | Sundseth, Jarle Jacobsen, Eva A. Kolstad, Frode Nygaard, Oystein P. Zwart, John A. Hol, Per K. |
author_facet | Sundseth, Jarle Jacobsen, Eva A. Kolstad, Frode Nygaard, Oystein P. Zwart, John A. Hol, Per K. |
author_sort | Sundseth, Jarle |
collection | PubMed |
description | PURPOSE: Cervical disc prostheses induce significant amount of artifact in magnetic resonance imaging which may complicate radiologic follow-up after surgery. The purpose of this study was to investigate as to what extent the artifact, induced by the frequently used Discover(®) cervical disc prosthesis, impedes interpretation of the MR images at operated and adjacent levels in 1.5 and 3 Tesla MR. METHODS: Ten subsequent patients were investigated in both 1.5 and 3 Tesla MR with standard image sequences one year following anterior cervical discectomy with arthroplasty. OUTCOME MEASURES: Two neuroradiologists evaluated the images by consensus. Emphasis was made on signal changes in medulla at all levels and visualization of root canals at operated and adjacent levels. A “blur artifact ratio” was calculated and defined as the height of the artifact on T1 sagittal images related to the operated level. RESULTS: The artifacts induced in 1.5 and 3 Tesla MR were of entirely different character and evaluation of the spinal cord at operated level was impossible in both magnets. Artifacts also made the root canals difficult to assess at operated level and more pronounced in the 3 Tesla MR. At the adjacent levels however, the spinal cord and root canals were completely visualized in all patients. The “blur artifact” induced at operated level was also more pronounced in the 3 Tesla MR. CONCLUSIONS: The artifact induced by the Discover(®) titanium disc prosthesis in both 1.5 and 3 Tesla MR, makes interpretation of the spinal cord impossible and visualization of the root canals difficult at operated level. Adjusting the MR sequences to produce the least amount of artifact is important. |
format | Online Article Text |
id | pubmed-3804698 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-38046982013-10-29 Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength Sundseth, Jarle Jacobsen, Eva A. Kolstad, Frode Nygaard, Oystein P. Zwart, John A. Hol, Per K. Eur Spine J Original Article PURPOSE: Cervical disc prostheses induce significant amount of artifact in magnetic resonance imaging which may complicate radiologic follow-up after surgery. The purpose of this study was to investigate as to what extent the artifact, induced by the frequently used Discover(®) cervical disc prosthesis, impedes interpretation of the MR images at operated and adjacent levels in 1.5 and 3 Tesla MR. METHODS: Ten subsequent patients were investigated in both 1.5 and 3 Tesla MR with standard image sequences one year following anterior cervical discectomy with arthroplasty. OUTCOME MEASURES: Two neuroradiologists evaluated the images by consensus. Emphasis was made on signal changes in medulla at all levels and visualization of root canals at operated and adjacent levels. A “blur artifact ratio” was calculated and defined as the height of the artifact on T1 sagittal images related to the operated level. RESULTS: The artifacts induced in 1.5 and 3 Tesla MR were of entirely different character and evaluation of the spinal cord at operated level was impossible in both magnets. Artifacts also made the root canals difficult to assess at operated level and more pronounced in the 3 Tesla MR. At the adjacent levels however, the spinal cord and root canals were completely visualized in all patients. The “blur artifact” induced at operated level was also more pronounced in the 3 Tesla MR. CONCLUSIONS: The artifact induced by the Discover(®) titanium disc prosthesis in both 1.5 and 3 Tesla MR, makes interpretation of the spinal cord impossible and visualization of the root canals difficult at operated level. Adjusting the MR sequences to produce the least amount of artifact is important. Springer Berlin Heidelberg 2013-09-06 2013-10 /pmc/articles/PMC3804698/ /pubmed/24061966 http://dx.doi.org/10.1007/s00586-013-2994-z Text en © The Author(s) 2013 https://creativecommons.org/licenses/by/2.0/ Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. |
spellingShingle | Original Article Sundseth, Jarle Jacobsen, Eva A. Kolstad, Frode Nygaard, Oystein P. Zwart, John A. Hol, Per K. Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title | Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title_full | Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title_fullStr | Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title_full_unstemmed | Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title_short | Magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 Tesla magnet strength |
title_sort | magnetic resonance imaging evaluation after implantation of a titanium cervical disc prosthesis: a comparison of 1.5 and 3 tesla magnet strength |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3804698/ https://www.ncbi.nlm.nih.gov/pubmed/24061966 http://dx.doi.org/10.1007/s00586-013-2994-z |
work_keys_str_mv | AT sundsethjarle magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength AT jacobsenevaa magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength AT kolstadfrode magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength AT nygaardoysteinp magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength AT zwartjohna magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength AT holperk magneticresonanceimagingevaluationafterimplantationofatitaniumcervicaldiscprosthesisacomparisonof15and3teslamagnetstrength |