Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure

Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP. Methods. Forty-six eyes of healthy volunteers were measured by two...

Descripción completa

Detalles Bibliográficos
Autores principales: Fujimura, Fusako, Kamiya, Kazutaka, Fujiwara, Kazuko, Shoji, Nobuyuki, Shimizu, Kimiya
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3816021/
https://www.ncbi.nlm.nih.gov/pubmed/24222904
http://dx.doi.org/10.1155/2013/370592
_version_ 1782289477905940480
author Fujimura, Fusako
Kamiya, Kazutaka
Fujiwara, Kazuko
Shoji, Nobuyuki
Shimizu, Kimiya
author_facet Fujimura, Fusako
Kamiya, Kazutaka
Fujiwara, Kazuko
Shoji, Nobuyuki
Shimizu, Kimiya
author_sort Fujimura, Fusako
collection PubMed
description Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP. Methods. Forty-six eyes of healthy volunteers were measured by two examiners. Three consecutive measurements per eye were performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P. Results. The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were −0.17 ± 1.42 mmHg (range: −2.95 to 2.61 mmHg) for IOP, −0.93 ± 4.37 μm (range: −9.50 to 7.64 μm) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (P = 0.010.3). The noncorrected IOP significantly correlated with CCT (r = −0.4883, P = 0.0006). The corrected IOP showed no significant correlation with CCT (r = −0.0285, P = 0.8509). Conclusions. NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.
format Online
Article
Text
id pubmed-3816021
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-38160212013-11-11 Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure Fujimura, Fusako Kamiya, Kazutaka Fujiwara, Kazuko Shoji, Nobuyuki Shimizu, Kimiya Biomed Res Int Research Article Purpose. To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP. Methods. Forty-six eyes of healthy volunteers were measured by two examiners. Three consecutive measurements per eye were performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P. Results. The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were −0.17 ± 1.42 mmHg (range: −2.95 to 2.61 mmHg) for IOP, −0.93 ± 4.37 μm (range: −9.50 to 7.64 μm) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (P = 0.010.3). The noncorrected IOP significantly correlated with CCT (r = −0.4883, P = 0.0006). The corrected IOP showed no significant correlation with CCT (r = −0.0285, P = 0.8509). Conclusions. NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness. Hindawi Publishing Corporation 2013 2013-10-08 /pmc/articles/PMC3816021/ /pubmed/24222904 http://dx.doi.org/10.1155/2013/370592 Text en Copyright © 2013 Fusako Fujimura et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Fujimura, Fusako
Kamiya, Kazutaka
Fujiwara, Kazuko
Shoji, Nobuyuki
Shimizu, Kimiya
Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title_full Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title_fullStr Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title_full_unstemmed Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title_short Repeatability and Reproducibility of Measurements Using a NT-530P Noncontact Tono/Pachymeter and Correlation of Central Corneal Thickness with Intraocular Pressure
title_sort repeatability and reproducibility of measurements using a nt-530p noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3816021/
https://www.ncbi.nlm.nih.gov/pubmed/24222904
http://dx.doi.org/10.1155/2013/370592
work_keys_str_mv AT fujimurafusako repeatabilityandreproducibilityofmeasurementsusingant530pnoncontacttonopachymeterandcorrelationofcentralcornealthicknesswithintraocularpressure
AT kamiyakazutaka repeatabilityandreproducibilityofmeasurementsusingant530pnoncontacttonopachymeterandcorrelationofcentralcornealthicknesswithintraocularpressure
AT fujiwarakazuko repeatabilityandreproducibilityofmeasurementsusingant530pnoncontacttonopachymeterandcorrelationofcentralcornealthicknesswithintraocularpressure
AT shojinobuyuki repeatabilityandreproducibilityofmeasurementsusingant530pnoncontacttonopachymeterandcorrelationofcentralcornealthicknesswithintraocularpressure
AT shimizukimiya repeatabilityandreproducibilityofmeasurementsusingant530pnoncontacttonopachymeterandcorrelationofcentralcornealthicknesswithintraocularpressure