Cargando…

Temperature Stability of the Sky Quality Meter

The stability of radiance measurements taken by the Sky Quality Meter (SQM) was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the...

Descripción completa

Detalles Bibliográficos
Autores principales: Schnitt, Sabrina, Ruhtz, Thomas, Fischer, Jürgen, Hölker, Franz, Kyba, Christopher C.M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3821345/
https://www.ncbi.nlm.nih.gov/pubmed/24030682
http://dx.doi.org/10.3390/s130912166
Descripción
Sumario:The stability of radiance measurements taken by the Sky Quality Meter (SQM) was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the initial reported radiance over a temperature range of −15°C to 35°C, and during temperature changes of −33°C/h and +70°C/h. This is smaller than the manufacturer's quoted unit-to-unit systematic uncertainty of 10%, indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.