Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom

We validated 3D radiochromic film dosimetry for volumetric modulated arc therapy (VMAT) using a newly developed spiral water phantom. The phantom consists of a main body and an insert box, each of which has an acrylic wall thickness of 3 mm and is filled with water. The insert box includes a spiral...

Descripción completa

Detalles Bibliográficos
Autores principales: Tanooka, Masao, Doi, Hiroshi, Miura, Hideharu, Inoue, Hiroyuki, Niwa, Yasue, Takada, Yasuhiro, Fujiwara, Masayuki, Sakai, Toshiyuki, Sakamoto, Kiyoshi, Kamikonya, Norihiko, Hirota, Shozo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3823780/
https://www.ncbi.nlm.nih.gov/pubmed/23685667
http://dx.doi.org/10.1093/jrr/rrt059
_version_ 1782290611942981632
author Tanooka, Masao
Doi, Hiroshi
Miura, Hideharu
Inoue, Hiroyuki
Niwa, Yasue
Takada, Yasuhiro
Fujiwara, Masayuki
Sakai, Toshiyuki
Sakamoto, Kiyoshi
Kamikonya, Norihiko
Hirota, Shozo
author_facet Tanooka, Masao
Doi, Hiroshi
Miura, Hideharu
Inoue, Hiroyuki
Niwa, Yasue
Takada, Yasuhiro
Fujiwara, Masayuki
Sakai, Toshiyuki
Sakamoto, Kiyoshi
Kamikonya, Norihiko
Hirota, Shozo
author_sort Tanooka, Masao
collection PubMed
description We validated 3D radiochromic film dosimetry for volumetric modulated arc therapy (VMAT) using a newly developed spiral water phantom. The phantom consists of a main body and an insert box, each of which has an acrylic wall thickness of 3 mm and is filled with water. The insert box includes a spiral film box used for dose-distribution measurement, and a film holder for positioning a radiochromic film. The film holder has two parallel walls whose facing inner surfaces are equipped with spiral grooves in a mirrored configuration. The film is inserted into the spiral grooves by its side edges and runs along them to be positioned on a spiral plane. Dose calculation was performed by applying clinical VMAT plans to the spiral water phantom using a commercial Monte Carlo-based treatment-planning system, Monaco, whereas dose was measured by delivering the VMAT beams to the phantom. The calculated dose distributions were resampled on the spiral plane, and the dose distributions recorded on the film were scanned. Comparisons between the calculated and measured dose distributions yielded an average gamma-index pass rate of 87.0% (range, 91.2–84.6%) in nine prostate VMAT plans under 3 mm/3% criteria with a dose-calculation grid size of 2 mm. The pass rates were increased beyond 90% (average, 91.1%; range, 90.1–92.0%) when the dose-calculation grid size was decreased to 1 mm. We have confirmed that 3D radiochromic film dosimetry using the spiral water phantom is a simple and cost-effective approach to VMAT dose verification.
format Online
Article
Text
id pubmed-3823780
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-38237802013-11-12 Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom Tanooka, Masao Doi, Hiroshi Miura, Hideharu Inoue, Hiroyuki Niwa, Yasue Takada, Yasuhiro Fujiwara, Masayuki Sakai, Toshiyuki Sakamoto, Kiyoshi Kamikonya, Norihiko Hirota, Shozo J Radiat Res Technology We validated 3D radiochromic film dosimetry for volumetric modulated arc therapy (VMAT) using a newly developed spiral water phantom. The phantom consists of a main body and an insert box, each of which has an acrylic wall thickness of 3 mm and is filled with water. The insert box includes a spiral film box used for dose-distribution measurement, and a film holder for positioning a radiochromic film. The film holder has two parallel walls whose facing inner surfaces are equipped with spiral grooves in a mirrored configuration. The film is inserted into the spiral grooves by its side edges and runs along them to be positioned on a spiral plane. Dose calculation was performed by applying clinical VMAT plans to the spiral water phantom using a commercial Monte Carlo-based treatment-planning system, Monaco, whereas dose was measured by delivering the VMAT beams to the phantom. The calculated dose distributions were resampled on the spiral plane, and the dose distributions recorded on the film were scanned. Comparisons between the calculated and measured dose distributions yielded an average gamma-index pass rate of 87.0% (range, 91.2–84.6%) in nine prostate VMAT plans under 3 mm/3% criteria with a dose-calculation grid size of 2 mm. The pass rates were increased beyond 90% (average, 91.1%; range, 90.1–92.0%) when the dose-calculation grid size was decreased to 1 mm. We have confirmed that 3D radiochromic film dosimetry using the spiral water phantom is a simple and cost-effective approach to VMAT dose verification. Oxford University Press 2013-11 2013-05-17 /pmc/articles/PMC3823780/ /pubmed/23685667 http://dx.doi.org/10.1093/jrr/rrt059 Text en © The Author 2013. Published by Oxford University Press on behalf of The Japan Radiation Research Society and Japanese Society for Therapeutic Radiology and Oncology. http://creativecommons.org/licenses/by/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technology
Tanooka, Masao
Doi, Hiroshi
Miura, Hideharu
Inoue, Hiroyuki
Niwa, Yasue
Takada, Yasuhiro
Fujiwara, Masayuki
Sakai, Toshiyuki
Sakamoto, Kiyoshi
Kamikonya, Norihiko
Hirota, Shozo
Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title_full Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title_fullStr Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title_full_unstemmed Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title_short Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
title_sort three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom
topic Technology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3823780/
https://www.ncbi.nlm.nih.gov/pubmed/23685667
http://dx.doi.org/10.1093/jrr/rrt059
work_keys_str_mv AT tanookamasao threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT doihiroshi threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT miurahideharu threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT inouehiroyuki threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT niwayasue threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT takadayasuhiro threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT fujiwaramasayuki threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT sakaitoshiyuki threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT sakamotokiyoshi threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT kamikonyanorihiko threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom
AT hirotashozo threedimensionalradiochromicfilmdosimetryforvolumetricmodulatedarctherapyusingaspiralwaterphantom