Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones

OBJECTIVE: This study compared the adequacy of dental cone beam computed tomography (CBCT) and micro computed tomography (micro-CT) in evaluating the structural parameters of trabecular bones. METHODS: The cellular synthetic bones in 4 density groups (Groups 1–4: 0.12, 0.16, 0.20, and 0.32 g/cm(3))...

Descripción completa

Detalles Bibliográficos
Autores principales: Ho, Jung-Ting, Wu, Jay, Huang, Heng-Li, Chen, Michael YC, Fuh, Lih-Jyh, Hsu, Jui-Ting
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3826852/
https://www.ncbi.nlm.nih.gov/pubmed/24207062
http://dx.doi.org/10.1186/1475-925X-12-115
_version_ 1782290968995692544
author Ho, Jung-Ting
Wu, Jay
Huang, Heng-Li
Chen, Michael YC
Fuh, Lih-Jyh
Hsu, Jui-Ting
author_facet Ho, Jung-Ting
Wu, Jay
Huang, Heng-Li
Chen, Michael YC
Fuh, Lih-Jyh
Hsu, Jui-Ting
author_sort Ho, Jung-Ting
collection PubMed
description OBJECTIVE: This study compared the adequacy of dental cone beam computed tomography (CBCT) and micro computed tomography (micro-CT) in evaluating the structural parameters of trabecular bones. METHODS: The cellular synthetic bones in 4 density groups (Groups 1–4: 0.12, 0.16, 0.20, and 0.32 g/cm(3)) were used in this study. Each group comprised 8 experimental specimens that were approximately 1 cm(3). Dental CBCT and micro-CT scans were conducted on each specimen to obtain independent measurements of the following 4 trabecular bone structural parameters: bone volume fraction (BV/TV), specific bone surface (BS/BV), trabecular thickness (Tb.Th.), and trabecular separation (Tb.Sp.). Wilcoxon signed ranks tests were used to compare the measurement variations between the dental CBCT and micro-CT scans. A Spearman analysis was conducted to calculate the correlation coefficients (r) of the dental CBCT and micro-CT measurements. RESULTS AND CONCLUSION: Of the 4 groups, the BV/TV and Tb.Th. measured using dental CBCT were larger compared with those measured using micro-CT. By contrast, the BS/BV measured using dental CBCT was significantly less compared with those measured using micro-CT. Furthermore, in the low-density groups (Groups 1 and 2), the Tb.Sp. measured using dental CBCT was smaller compared with those measured using micro-CT. However, the Tb.Sp. measured using dental CBCT was slightly larger in the high-density groups (Groups 3 and 4) than it was in the low density groups. The correlation coefficients between the BV/TV, BS/BV, Tb.Th., and Tb.Sp. values measured using dental CBCT and micro-CT were 0.9296 (p < .001), 0.8061 (p < .001), 0.9390 (p < .001), and 0.9583 (p < .001), respectively. Although the dental CBCT and micro-CT approaches exhibited high correlations, the absolute values of BV/TV, BS/BV, Tb.Th., Tb.Sp. differed significantly between these measurements. Additional studies must be conducted to evaluate using dental CBCT in clinical practice.
format Online
Article
Text
id pubmed-3826852
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-38268522013-11-14 Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones Ho, Jung-Ting Wu, Jay Huang, Heng-Li Chen, Michael YC Fuh, Lih-Jyh Hsu, Jui-Ting Biomed Eng Online Research OBJECTIVE: This study compared the adequacy of dental cone beam computed tomography (CBCT) and micro computed tomography (micro-CT) in evaluating the structural parameters of trabecular bones. METHODS: The cellular synthetic bones in 4 density groups (Groups 1–4: 0.12, 0.16, 0.20, and 0.32 g/cm(3)) were used in this study. Each group comprised 8 experimental specimens that were approximately 1 cm(3). Dental CBCT and micro-CT scans were conducted on each specimen to obtain independent measurements of the following 4 trabecular bone structural parameters: bone volume fraction (BV/TV), specific bone surface (BS/BV), trabecular thickness (Tb.Th.), and trabecular separation (Tb.Sp.). Wilcoxon signed ranks tests were used to compare the measurement variations between the dental CBCT and micro-CT scans. A Spearman analysis was conducted to calculate the correlation coefficients (r) of the dental CBCT and micro-CT measurements. RESULTS AND CONCLUSION: Of the 4 groups, the BV/TV and Tb.Th. measured using dental CBCT were larger compared with those measured using micro-CT. By contrast, the BS/BV measured using dental CBCT was significantly less compared with those measured using micro-CT. Furthermore, in the low-density groups (Groups 1 and 2), the Tb.Sp. measured using dental CBCT was smaller compared with those measured using micro-CT. However, the Tb.Sp. measured using dental CBCT was slightly larger in the high-density groups (Groups 3 and 4) than it was in the low density groups. The correlation coefficients between the BV/TV, BS/BV, Tb.Th., and Tb.Sp. values measured using dental CBCT and micro-CT were 0.9296 (p < .001), 0.8061 (p < .001), 0.9390 (p < .001), and 0.9583 (p < .001), respectively. Although the dental CBCT and micro-CT approaches exhibited high correlations, the absolute values of BV/TV, BS/BV, Tb.Th., Tb.Sp. differed significantly between these measurements. Additional studies must be conducted to evaluate using dental CBCT in clinical practice. BioMed Central 2013-11-09 /pmc/articles/PMC3826852/ /pubmed/24207062 http://dx.doi.org/10.1186/1475-925X-12-115 Text en Copyright © 2013 Ho et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Ho, Jung-Ting
Wu, Jay
Huang, Heng-Li
Chen, Michael YC
Fuh, Lih-Jyh
Hsu, Jui-Ting
Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title_full Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title_fullStr Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title_full_unstemmed Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title_short Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
title_sort trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3826852/
https://www.ncbi.nlm.nih.gov/pubmed/24207062
http://dx.doi.org/10.1186/1475-925X-12-115
work_keys_str_mv AT hojungting trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones
AT wujay trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones
AT huanghengli trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones
AT chenmichaelyc trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones
AT fuhlihjyh trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones
AT hsujuiting trabecularbonestructuralparametersevaluatedusingdentalconebeamcomputedtomographycellularsyntheticbones