Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood

Regular immunosuppressant drug monitoring is important for maintaining the drug concentrations of organ recipients within the therapeutic range. The standardized liquid chromatography-tandem mass spectrometry (LC-TMS) technique has been used for the accurate analysis of immunosuppressive drugs. In t...

Descripción completa

Detalles Bibliográficos
Autor principal: LEE, YONG-WHA
Formato: Online Artículo Texto
Lenguaje:English
Publicado: D.A. Spandidos 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3829750/
https://www.ncbi.nlm.nih.gov/pubmed/24255687
http://dx.doi.org/10.3892/etm.2013.1325
_version_ 1782291381802237952
author LEE, YONG-WHA
author_facet LEE, YONG-WHA
author_sort LEE, YONG-WHA
collection PubMed
description Regular immunosuppressant drug monitoring is important for maintaining the drug concentrations of organ recipients within the therapeutic range. The standardized liquid chromatography-tandem mass spectrometry (LC-TMS) technique has been used for the accurate analysis of immunosuppressive drugs. In the present study, the performance of the recently developed high-throughput, rapid ultra-performance liquid chromatography combined with tandem mass spectrometry (UPLC-TMS) method was validated for the simultaneous measurement of cyclosporin A and tacrolimus in whole blood. The method of measuring cyclosporin A and tacrolimus using UPLC-TMS was established and the precision, limit of detection (LOD), limit of quantitation (LOQ) and matrix effect were validated. In addition, the performance of UPLC-TMS was compared with that of a chemiluminescence immunoassay (CLIA) in >3,400 clinical specimens. The UPLC-TMS revealed a within-run and between-run precision of <8% and showed a bias of <5%. The LOD and LOQ were 2.0 and 2.5 ng/ml for cyclosporin A, and 0.3 and 0.4 ng/ml for tacrolimus, respectively. Interference from the matrix was not observed. The CLIA measurements of cyclosporin A and tacrolimus showed correlations corresponding with the formulae: Concentration((CLIA)) = 1.18 × UPLC-TMS – 5.85; [95% CI: proportional, 1.16–1.19; constant, −6.86–(−4.81)] and Concentration((CLIA)) = 1.14 × UPLC-TMS – 0.38; [(95% CI: proportional, 1.13–1.14; constant, −0.35–(−0.43)], respectively. The majority of results were higher for the immunoassay than for the UPLC-TMS. The newly developed rapid UPLC-TMS method was suitable for use with a large therapeutic concentration range of the analyzed immunosuppressive drugs. Sample preparation was simple and it was possible to detect several immunosuppressants simultaneously, thus significantly lowering the cost of analysis. In conclusion, this method may contribute to improved accuracy and may be preferred to immunoassays for the routine clinical measurement of immunosuppressive drug concentrations in whole blood.
format Online
Article
Text
id pubmed-3829750
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher D.A. Spandidos
record_format MEDLINE/PubMed
spelling pubmed-38297502013-11-19 Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood LEE, YONG-WHA Exp Ther Med Articles Regular immunosuppressant drug monitoring is important for maintaining the drug concentrations of organ recipients within the therapeutic range. The standardized liquid chromatography-tandem mass spectrometry (LC-TMS) technique has been used for the accurate analysis of immunosuppressive drugs. In the present study, the performance of the recently developed high-throughput, rapid ultra-performance liquid chromatography combined with tandem mass spectrometry (UPLC-TMS) method was validated for the simultaneous measurement of cyclosporin A and tacrolimus in whole blood. The method of measuring cyclosporin A and tacrolimus using UPLC-TMS was established and the precision, limit of detection (LOD), limit of quantitation (LOQ) and matrix effect were validated. In addition, the performance of UPLC-TMS was compared with that of a chemiluminescence immunoassay (CLIA) in >3,400 clinical specimens. The UPLC-TMS revealed a within-run and between-run precision of <8% and showed a bias of <5%. The LOD and LOQ were 2.0 and 2.5 ng/ml for cyclosporin A, and 0.3 and 0.4 ng/ml for tacrolimus, respectively. Interference from the matrix was not observed. The CLIA measurements of cyclosporin A and tacrolimus showed correlations corresponding with the formulae: Concentration((CLIA)) = 1.18 × UPLC-TMS – 5.85; [95% CI: proportional, 1.16–1.19; constant, −6.86–(−4.81)] and Concentration((CLIA)) = 1.14 × UPLC-TMS – 0.38; [(95% CI: proportional, 1.13–1.14; constant, −0.35–(−0.43)], respectively. The majority of results were higher for the immunoassay than for the UPLC-TMS. The newly developed rapid UPLC-TMS method was suitable for use with a large therapeutic concentration range of the analyzed immunosuppressive drugs. Sample preparation was simple and it was possible to detect several immunosuppressants simultaneously, thus significantly lowering the cost of analysis. In conclusion, this method may contribute to improved accuracy and may be preferred to immunoassays for the routine clinical measurement of immunosuppressive drug concentrations in whole blood. D.A. Spandidos 2013-12 2013-10-04 /pmc/articles/PMC3829750/ /pubmed/24255687 http://dx.doi.org/10.3892/etm.2013.1325 Text en Copyright © 2013, Spandidos Publications http://creativecommons.org/licenses/by/3.0 This is an open-access article licensed under a Creative Commons Attribution-NonCommercial 3.0 Unported License. The article may be redistributed, reproduced, and reused for non-commercial purposes, provided the original source is properly cited.
spellingShingle Articles
LEE, YONG-WHA
Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title_full Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title_fullStr Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title_full_unstemmed Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title_short Comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin A and tacrolimus levels in whole blood
title_sort comparison between ultra-performance liquid chromatography with tandem mass spectrometry and a chemiluminescence immunoassay in the determination of cyclosporin a and tacrolimus levels in whole blood
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3829750/
https://www.ncbi.nlm.nih.gov/pubmed/24255687
http://dx.doi.org/10.3892/etm.2013.1325
work_keys_str_mv AT leeyongwha comparisonbetweenultraperformanceliquidchromatographywithtandemmassspectrometryandachemiluminescenceimmunoassayinthedeterminationofcyclosporinaandtacrolimuslevelsinwholeblood