Cargando…
Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking
Friction-induced moments and subsequent cup loosening can be the reason for total hip joint replacement failure. The aim of this study was to measure the in vivo contact forces and friction moments during walking. Instrumented hip implants with Al(2)O(3) ceramic head and an XPE inlay were used. In v...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3832636/ https://www.ncbi.nlm.nih.gov/pubmed/24260114 http://dx.doi.org/10.1371/journal.pone.0078373 |
_version_ | 1782291715590193152 |
---|---|
author | Damm, Philipp Dymke, Joern Ackermann, Robert Bender, Alwina Graichen, Friedmar Halder, Andreas Beier, Alexander Bergmann, Georg |
author_facet | Damm, Philipp Dymke, Joern Ackermann, Robert Bender, Alwina Graichen, Friedmar Halder, Andreas Beier, Alexander Bergmann, Georg |
author_sort | Damm, Philipp |
collection | PubMed |
description | Friction-induced moments and subsequent cup loosening can be the reason for total hip joint replacement failure. The aim of this study was to measure the in vivo contact forces and friction moments during walking. Instrumented hip implants with Al(2)O(3) ceramic head and an XPE inlay were used. In vivo measurements were taken 3 months post operatively in 8 subjects. The coefficient of friction was calculated in 3D throughout the whole gait cycle, and average values of the friction-induced power dissipation in the joint were determined. On average, peak contact forces of 248% of the bodyweight and peak friction moments of 0.26% bodyweight times meter were determined. However, contact forces and friction moments varied greatly between individuals. The friction moment increased during the extension phase of the joint. The average coefficient of friction also increased during this period, from 0.04 (0.03 to 0.06) at contralateral toe off to 0.06 (0.04 to 0.08) at contralateral heel strike. During the flexion phase, the coefficient of friction increased further to 0.14 (0.09 to 0.23) at toe off. The average friction-induced power throughout the whole gait cycle was 2.3 W (1.4 W to 3.8 W). Although more parameters than only the synovia determine the friction, the wide ranges of friction coefficients and power dissipation indicate that the lubricating properties of synovia are individually very different. However, such differences may also exist in natural joints and may influence the progression of arthrosis. Furthermore, subjects with very high power dissipation may be at risk of thermally induced implant loosening. The large increase of the friction coefficient during each step could be caused by the synovia being squeezed out under load. |
format | Online Article Text |
id | pubmed-3832636 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-38326362013-11-20 Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking Damm, Philipp Dymke, Joern Ackermann, Robert Bender, Alwina Graichen, Friedmar Halder, Andreas Beier, Alexander Bergmann, Georg PLoS One Research Article Friction-induced moments and subsequent cup loosening can be the reason for total hip joint replacement failure. The aim of this study was to measure the in vivo contact forces and friction moments during walking. Instrumented hip implants with Al(2)O(3) ceramic head and an XPE inlay were used. In vivo measurements were taken 3 months post operatively in 8 subjects. The coefficient of friction was calculated in 3D throughout the whole gait cycle, and average values of the friction-induced power dissipation in the joint were determined. On average, peak contact forces of 248% of the bodyweight and peak friction moments of 0.26% bodyweight times meter were determined. However, contact forces and friction moments varied greatly between individuals. The friction moment increased during the extension phase of the joint. The average coefficient of friction also increased during this period, from 0.04 (0.03 to 0.06) at contralateral toe off to 0.06 (0.04 to 0.08) at contralateral heel strike. During the flexion phase, the coefficient of friction increased further to 0.14 (0.09 to 0.23) at toe off. The average friction-induced power throughout the whole gait cycle was 2.3 W (1.4 W to 3.8 W). Although more parameters than only the synovia determine the friction, the wide ranges of friction coefficients and power dissipation indicate that the lubricating properties of synovia are individually very different. However, such differences may also exist in natural joints and may influence the progression of arthrosis. Furthermore, subjects with very high power dissipation may be at risk of thermally induced implant loosening. The large increase of the friction coefficient during each step could be caused by the synovia being squeezed out under load. Public Library of Science 2013-11-08 /pmc/articles/PMC3832636/ /pubmed/24260114 http://dx.doi.org/10.1371/journal.pone.0078373 Text en © 2013 Damm et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Damm, Philipp Dymke, Joern Ackermann, Robert Bender, Alwina Graichen, Friedmar Halder, Andreas Beier, Alexander Bergmann, Georg Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title | Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title_full | Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title_fullStr | Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title_full_unstemmed | Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title_short | Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking |
title_sort | friction in total hip joint prosthesis measured in vivo during walking |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3832636/ https://www.ncbi.nlm.nih.gov/pubmed/24260114 http://dx.doi.org/10.1371/journal.pone.0078373 |
work_keys_str_mv | AT dammphilipp frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT dymkejoern frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT ackermannrobert frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT benderalwina frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT graichenfriedmar frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT halderandreas frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT beieralexander frictionintotalhipjointprosthesismeasuredinvivoduringwalking AT bergmanngeorg frictionintotalhipjointprosthesismeasuredinvivoduringwalking |