Cargando…

Ambulatory Holter monitoring in asymptomatic patients with DDD pacemakers – do we need ACC/AHA Guidelines revision?

INTRODUCTION: We sought to determine the usefulness of ambulatory 24-hour Holter monitoring in detecting asymptomatic pacemaker (PM) malfunction episodes in patients with dual-chamber pacemakers whose pacing and sensing parameters were proper, as seen in routine post-implantation follow-ups. MATERIA...

Descripción completa

Detalles Bibliográficos
Autores principales: Chudzik, Michal, Klimczak, Artur, Wranicz, Jerzy Krzysztof
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3832827/
https://www.ncbi.nlm.nih.gov/pubmed/24273562
http://dx.doi.org/10.5114/aoms.2013.38675
Descripción
Sumario:INTRODUCTION: We sought to determine the usefulness of ambulatory 24-hour Holter monitoring in detecting asymptomatic pacemaker (PM) malfunction episodes in patients with dual-chamber pacemakers whose pacing and sensing parameters were proper, as seen in routine post-implantation follow-ups. MATERIAL AND METHODS: Ambulatory 24-hour Holter recordings (HM) were performed in 100 patients with DDD pacemakers 1 day after the implantation. Only asymptomatic patients with proper pacing and sensing parameters (assessed on PM telemetry on the first day post-implantation) were enrolled in the study. The following parameters were assessed: failure to pace, failure to sense (both oversensing and undersensing episodes) as well as the percentage of all PM disturbances. RESULTS: Despite proper sensing and pacing parameters, HM revealed PM disturbances in 23 patients out of 100 (23%). Atrial undersensing episodes were found in 12 patients (p < 0.005) with totally 963 episodes and failure to capture in 1 patient (1%). T wave oversensing was the most common ventricular channel disorder (1316 episodes in 9 patients, p < 0.0005). Malfunction episodes occurred sporadically, leading to pauses of up to 1.6 s or temporary bradycardia, which were, nevertheless, not accompanied by clinical symptoms. No ventricular pacing disturbances were found. CONCLUSIONS: Asymptomatic pacemaker dysfunction may be observed in nearly 25% of patients with proper DDD parameters after implantation. Thus, ambulatory HM during the early post-implantation period may be a useful tool to detect the need to reprogram PM parameters.