Re-irradiation of spinal column metastases by IMRT: impact of setup errors on the dose distribution
BACKGROUND: This study investigates the impact of an automated image guided patient setup correction on the dose distribution for ten patients with in-field IMRT re-irradiation of vertebral metastases. METHODS: 10 patients with spinal column metastases who had previously been treated with 3D-conform...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3842751/ https://www.ncbi.nlm.nih.gov/pubmed/24238332 http://dx.doi.org/10.1186/1748-717X-8-269 |
Sumario: | BACKGROUND: This study investigates the impact of an automated image guided patient setup correction on the dose distribution for ten patients with in-field IMRT re-irradiation of vertebral metastases. METHODS: 10 patients with spinal column metastases who had previously been treated with 3D-conformal radiotherapy (3D-CRT) were simulated to have an in-field recurrence. IMRT plans were generated for treatment of the vertebrae sparing the spinal cord. The dose distributions were compared for a patient setup based on skin marks only and a Cone Beam CT (CBCT) based setup with translational and rotational couch corrections using an automatic robotic image guided couch top (Elekta - HexaPOD™ IGuide(®) - system). The biological equivalent dose (BED) was calculated to evaluate and rank the effects of the automatic setup correction for the dose distribution of CTV and spinal cord. RESULTS: The mean absolute value (± standard deviation) over all patients and fractions of the translational error is 6.1 mm (±4 mm) and 2.7° (±1.1 mm) for the rotational error. The dose coverage of the 95% isodose for the CTV is considerable decreased for the uncorrected table setup. This is associated with an increasing of the spinal cord dose above the tolerance dose. CONCLUSIONS: An automatic image guided table correction ensures the delivery of accurate dose distribution and reduces the risk of radiation induced myelopathy. |
---|