Cargando…
Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis
Ongoing clinical studies on patients recently implanted with the auditory midbrain implant (AMI) into the inferior colliculus (IC) for hearing restoration have shown that these patients do not achieve performance levels comparable to cochlear implant patients. The AMI consists of a single-shank arra...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3846787/ https://www.ncbi.nlm.nih.gov/pubmed/24312638 http://dx.doi.org/10.1371/journal.pone.0082148 |
_version_ | 1782293486167392256 |
---|---|
author | Calixto, Roger Salamat, Behrouz Rode, Thilo Hartmann, Tanja Volckaerts, Bart Ruther, Patrick Lenarz, Thomas Lim, Hubert H. |
author_facet | Calixto, Roger Salamat, Behrouz Rode, Thilo Hartmann, Tanja Volckaerts, Bart Ruther, Patrick Lenarz, Thomas Lim, Hubert H. |
author_sort | Calixto, Roger |
collection | PubMed |
description | Ongoing clinical studies on patients recently implanted with the auditory midbrain implant (AMI) into the inferior colliculus (IC) for hearing restoration have shown that these patients do not achieve performance levels comparable to cochlear implant patients. The AMI consists of a single-shank array (20 electrodes) for stimulation along the tonotopic axis of the IC. Recent findings suggest that one major limitation in AMI performance is the inability to sufficiently activate neurons across the three-dimensional (3-D) IC. Unfortunately, there are no currently available 3-D array technologies that can be used for clinical applications. More recently, there has been a new initiative by the European Commission to fund and develop 3-D chronic electrode arrays for science and clinical applications through the NeuroProbes project that can overcome the bulkiness and limited 3-D configurations of currently available array technologies. As part of the NeuroProbes initiative, we investigated whether their new array technology could be potentially used for future AMI patients. Since the NeuroProbes technology had not yet been tested for electrical stimulation in an in vivo animal preparation, we performed experiments in ketamine-anesthetized guinea pigs in which we inserted and stimulated a NeuroProbes array within the IC and recorded the corresponding neural activation within the auditory cortex. We used 2-D arrays for this initial feasibility study since they were already available and were sufficient to access the IC and also demonstrate effective activation of the central auditory system. Based on these encouraging results and the ability to develop customized 3-D arrays with the NeuroProbes technology, we can further investigate different stimulation patterns across the ICC to improve AMI performance. |
format | Online Article Text |
id | pubmed-3846787 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-38467872013-12-05 Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis Calixto, Roger Salamat, Behrouz Rode, Thilo Hartmann, Tanja Volckaerts, Bart Ruther, Patrick Lenarz, Thomas Lim, Hubert H. PLoS One Research Article Ongoing clinical studies on patients recently implanted with the auditory midbrain implant (AMI) into the inferior colliculus (IC) for hearing restoration have shown that these patients do not achieve performance levels comparable to cochlear implant patients. The AMI consists of a single-shank array (20 electrodes) for stimulation along the tonotopic axis of the IC. Recent findings suggest that one major limitation in AMI performance is the inability to sufficiently activate neurons across the three-dimensional (3-D) IC. Unfortunately, there are no currently available 3-D array technologies that can be used for clinical applications. More recently, there has been a new initiative by the European Commission to fund and develop 3-D chronic electrode arrays for science and clinical applications through the NeuroProbes project that can overcome the bulkiness and limited 3-D configurations of currently available array technologies. As part of the NeuroProbes initiative, we investigated whether their new array technology could be potentially used for future AMI patients. Since the NeuroProbes technology had not yet been tested for electrical stimulation in an in vivo animal preparation, we performed experiments in ketamine-anesthetized guinea pigs in which we inserted and stimulated a NeuroProbes array within the IC and recorded the corresponding neural activation within the auditory cortex. We used 2-D arrays for this initial feasibility study since they were already available and were sufficient to access the IC and also demonstrate effective activation of the central auditory system. Based on these encouraging results and the ability to develop customized 3-D arrays with the NeuroProbes technology, we can further investigate different stimulation patterns across the ICC to improve AMI performance. Public Library of Science 2013-12-02 /pmc/articles/PMC3846787/ /pubmed/24312638 http://dx.doi.org/10.1371/journal.pone.0082148 Text en © 2013 Calixto et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Calixto, Roger Salamat, Behrouz Rode, Thilo Hartmann, Tanja Volckaerts, Bart Ruther, Patrick Lenarz, Thomas Lim, Hubert H. Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title | Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title_full | Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title_fullStr | Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title_full_unstemmed | Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title_short | Investigation of a New Electrode Array Technology for a Central Auditory Prosthesis |
title_sort | investigation of a new electrode array technology for a central auditory prosthesis |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3846787/ https://www.ncbi.nlm.nih.gov/pubmed/24312638 http://dx.doi.org/10.1371/journal.pone.0082148 |
work_keys_str_mv | AT calixtoroger investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT salamatbehrouz investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT rodethilo investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT hartmanntanja investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT volckaertsbart investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT rutherpatrick investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT lenarzthomas investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis AT limhuberth investigationofanewelectrodearraytechnologyforacentralauditoryprosthesis |