Cargando…
Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance
BACKGROUND: We estimated sufficient setup margins for head-and-neck cancer (HNC) radiotherapy (RT) when 2D kV images are utilized for routine patient setup verification. As another goal we estimated a threshold for the displacements of the most important bony landmarks related to the target volumes...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3851546/ https://www.ncbi.nlm.nih.gov/pubmed/24020432 http://dx.doi.org/10.1186/1748-717X-8-212 |
_version_ | 1782294302526799872 |
---|---|
author | Kapanen, Mika Laaksomaa, Marko Tulijoki, Tapio Peltola, Seppo Wigren, Tuija Hyödynmaa, Simo Kellokumpu-Lehtinen, Pirkko-Liisa |
author_facet | Kapanen, Mika Laaksomaa, Marko Tulijoki, Tapio Peltola, Seppo Wigren, Tuija Hyödynmaa, Simo Kellokumpu-Lehtinen, Pirkko-Liisa |
author_sort | Kapanen, Mika |
collection | PubMed |
description | BACKGROUND: We estimated sufficient setup margins for head-and-neck cancer (HNC) radiotherapy (RT) when 2D kV images are utilized for routine patient setup verification. As another goal we estimated a threshold for the displacements of the most important bony landmarks related to the target volumes requiring immediate attention. METHODS: We analyzed 1491 orthogonal x-ray images utilized in RT treatment guidance for 80 HNC patients. We estimated overall setup errors and errors for four subregions to account for patient rotation and deformation: the vertebrae C1-2, C5-7, the occiput bone and the mandible. Setup margins were estimated for two 2D image guidance protocols: i) imaging at first three fractions and weekly thereafter and ii) daily imaging. Two 2D image matching principles were investigated: i) to the vertebrae in the middle of planning target volume (PTV) (MID_PTV) and ii) minimizing maximal position error for the four subregions (MIN_MAX). The threshold for the position errors was calculated with two previously unpublished methods based on the van Herk’s formula and clinical data by retaining a margin of 5 mm sufficient for each subregion. RESULTS: Sufficient setup margins to compensate the displacements of the subregions were approximately two times larger than were needed to compensate setup errors for rigid target. Adequate margins varied from 2.7 mm to 9.6 mm depending on the subregions related to the target, applied image guidance protocol and early correction of clinically important systematic 3D displacements of the subregions exceeding 4 mm. The MIN_MAX match resulted in smaller margins but caused an overall shift of 2.5 mm for the target center. Margins ≤ 5mm were sufficient with the MID_PTV match only through application of daily 2D imaging and the threshold of 4 mm to correct systematic displacement of a subregion. CONCLUSIONS: Adequate setup margins depend remarkably on the subregions related to the target volume. When the systematic 3D displacement of a subregion exceeds 4 mm, it is optimal to correct patient immobilization first. If this is not successful, adaptive replanning should be considered to retain sufficiently small margins. |
format | Online Article Text |
id | pubmed-3851546 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-38515462013-12-20 Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance Kapanen, Mika Laaksomaa, Marko Tulijoki, Tapio Peltola, Seppo Wigren, Tuija Hyödynmaa, Simo Kellokumpu-Lehtinen, Pirkko-Liisa Radiat Oncol Research BACKGROUND: We estimated sufficient setup margins for head-and-neck cancer (HNC) radiotherapy (RT) when 2D kV images are utilized for routine patient setup verification. As another goal we estimated a threshold for the displacements of the most important bony landmarks related to the target volumes requiring immediate attention. METHODS: We analyzed 1491 orthogonal x-ray images utilized in RT treatment guidance for 80 HNC patients. We estimated overall setup errors and errors for four subregions to account for patient rotation and deformation: the vertebrae C1-2, C5-7, the occiput bone and the mandible. Setup margins were estimated for two 2D image guidance protocols: i) imaging at first three fractions and weekly thereafter and ii) daily imaging. Two 2D image matching principles were investigated: i) to the vertebrae in the middle of planning target volume (PTV) (MID_PTV) and ii) minimizing maximal position error for the four subregions (MIN_MAX). The threshold for the position errors was calculated with two previously unpublished methods based on the van Herk’s formula and clinical data by retaining a margin of 5 mm sufficient for each subregion. RESULTS: Sufficient setup margins to compensate the displacements of the subregions were approximately two times larger than were needed to compensate setup errors for rigid target. Adequate margins varied from 2.7 mm to 9.6 mm depending on the subregions related to the target, applied image guidance protocol and early correction of clinically important systematic 3D displacements of the subregions exceeding 4 mm. The MIN_MAX match resulted in smaller margins but caused an overall shift of 2.5 mm for the target center. Margins ≤ 5mm were sufficient with the MID_PTV match only through application of daily 2D imaging and the threshold of 4 mm to correct systematic displacement of a subregion. CONCLUSIONS: Adequate setup margins depend remarkably on the subregions related to the target volume. When the systematic 3D displacement of a subregion exceeds 4 mm, it is optimal to correct patient immobilization first. If this is not successful, adaptive replanning should be considered to retain sufficiently small margins. BioMed Central 2013-09-10 /pmc/articles/PMC3851546/ /pubmed/24020432 http://dx.doi.org/10.1186/1748-717X-8-212 Text en Copyright © 2013 Kapanen et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Kapanen, Mika Laaksomaa, Marko Tulijoki, Tapio Peltola, Seppo Wigren, Tuija Hyödynmaa, Simo Kellokumpu-Lehtinen, Pirkko-Liisa Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title | Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title_full | Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title_fullStr | Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title_full_unstemmed | Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title_short | Estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2D image guidance |
title_sort | estimation of adequate setup margins and threshold for position errors requiring immediate attention in head and neck cancer radiotherapy based on 2d image guidance |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3851546/ https://www.ncbi.nlm.nih.gov/pubmed/24020432 http://dx.doi.org/10.1186/1748-717X-8-212 |
work_keys_str_mv | AT kapanenmika estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT laaksomaamarko estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT tulijokitapio estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT peltolaseppo estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT wigrentuija estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT hyodynmaasimo estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance AT kellokumpulehtinenpirkkoliisa estimationofadequatesetupmarginsandthresholdforpositionerrorsrequiringimmediateattentioninheadandneckcancerradiotherapybasedon2dimageguidance |