Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts

BACKGROUND: Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficie...

Descripción completa

Detalles Bibliográficos
Autores principales: Seppänen, Tiina M, Alho, Olli-Pekka, Seppänen, Tapio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3852323/
https://www.ncbi.nlm.nih.gov/pubmed/24073710
http://dx.doi.org/10.1186/1475-925X-12-97
_version_ 1782478650232274944
author Seppänen, Tiina M
Alho, Olli-Pekka
Seppänen, Tapio
author_facet Seppänen, Tiina M
Alho, Olli-Pekka
Seppänen, Tapio
author_sort Seppänen, Tiina M
collection PubMed
description BACKGROUND: Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject’s body position and breathing style. METHODS: An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature. RESULTS: Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R(2) varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods. CONCLUSIONS: Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required.
format Online
Article
Text
id pubmed-3852323
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-38523232013-12-19 Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts Seppänen, Tiina M Alho, Olli-Pekka Seppänen, Tapio Biomed Eng Online Research BACKGROUND: Respiratory effort belt measurement is a widely used method to monitor respiration. Signal waveforms of respiratory volume and flow may indicate pathological signs of several diseases and, thus, it would be highly desirable to predict them accurately. Calibrated effort belts are sufficiently accurate for estimating respiratory rate, but the respiratory volume and flow prediction accuracies degrade considerably with changes in the subject’s body position and breathing style. METHODS: An improved calibration method of respiratory effort belts is presented in this paper. It is based on an optimally trained FIR (Finite Impulse Response) filter bank constructed as a MISO system (Multiple-Input Single-Output) between respiratory effort belt signals and the spirometer in order to reduce waveform errors. Ten healthy adult volunteers were recruited. Breathing was varied between the following styles: metronome-guided controlled breathing rate of 0.1 Hz, 0.15 Hz, 0.25 Hz and 0.33 Hz, and a free rate that was felt normal by each subject. Body position was varied between supine, sitting and standing. The proposed calibration method was tested against these variations and compared with the state-of-the-art methods from the literature. RESULTS: Relative waveform error decreased 60-70% when predicting airflow under changing breathing styles. The coefficient of determination R(2) varied between 0.88-0.95 and 0.65-0.79 with the proposed and the standard method, respectively. Standard deviation of respiratory volume error decreased even 80%. The proposed method outperformed other methods. CONCLUSIONS: Results show that not only the respiratory volume can be computed more precisely from the predicted airflow, but also the flow waveforms are very accurate with the proposed method. The method is robust to breathing style changes and body position changes improving greatly the accuracy of the calibration of respiratory effort belts over the standard method. The enhanced accuracy of the belt calibration offers interesting opportunities, e.g. in pulmonary and critical care medicine when objective measurements are required. BioMed Central 2013-09-28 /pmc/articles/PMC3852323/ /pubmed/24073710 http://dx.doi.org/10.1186/1475-925X-12-97 Text en Copyright © 2013 Seppänen et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Seppänen, Tiina M
Alho, Olli-Pekka
Seppänen, Tapio
Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title_full Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title_fullStr Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title_full_unstemmed Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title_short Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
title_sort reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3852323/
https://www.ncbi.nlm.nih.gov/pubmed/24073710
http://dx.doi.org/10.1186/1475-925X-12-97
work_keys_str_mv AT seppanentiinam reducingtheairflowwaveformdistortionsfrombreathingstyleandbodypositionwithimprovedcalibrationofrespiratoryeffortbelts
AT alhoollipekka reducingtheairflowwaveformdistortionsfrombreathingstyleandbodypositionwithimprovedcalibrationofrespiratoryeffortbelts
AT seppanentapio reducingtheairflowwaveformdistortionsfrombreathingstyleandbodypositionwithimprovedcalibrationofrespiratoryeffortbelts