Cargando…
Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion
Lumbar interbody fusion is a common procedure owing to the high prevalence of degenerative spinal disorders. During such procedures, carbon fiber–reinforced polymer (CFRP) cages are frequently utilized to fill the void created between adjacent vertebral bodies, to provide mechanical stability, and t...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Georg Thieme Verlag KG
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3854590/ https://www.ncbi.nlm.nih.gov/pubmed/24436878 http://dx.doi.org/10.1055/s-0033-1343075 |
_version_ | 1782294833118838784 |
---|---|
author | Sardar, Zeeshan Jarzem, Peter |
author_facet | Sardar, Zeeshan Jarzem, Peter |
author_sort | Sardar, Zeeshan |
collection | PubMed |
description | Lumbar interbody fusion is a common procedure owing to the high prevalence of degenerative spinal disorders. During such procedures, carbon fiber–reinforced polymer (CFRP) cages are frequently utilized to fill the void created between adjacent vertebral bodies, to provide mechanical stability, and to carry graft material. Failure of such implants can lead to significant morbidity. We discuss the possible causes leading to the failure of a CFRP cage in a patient with rheumatoid arthritis. Review of a 49-year-old woman who underwent revision anterior lumbar interbody fusion 2 years after posterior instrumentation and transforaminal lumbar interbody fusion at L4–L5 and L5–S1. The patient developed pseudarthrosis at the two previously fused levels with failure of the posterior instrumentation. Revision surgery reveled failure with fragmentation of the CFRP cage at the L5–S1 level. CFRP implants can break if mechanical instability or nonunion occurs in the spinal segments, thus emphasizing the need for optimizing medical management and meticulous surgical technique in achieving stability. |
format | Online Article Text |
id | pubmed-3854590 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Georg Thieme Verlag KG |
record_format | MEDLINE/PubMed |
spelling | pubmed-38545902014-12-01 Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion Sardar, Zeeshan Jarzem, Peter Global Spine J Article Lumbar interbody fusion is a common procedure owing to the high prevalence of degenerative spinal disorders. During such procedures, carbon fiber–reinforced polymer (CFRP) cages are frequently utilized to fill the void created between adjacent vertebral bodies, to provide mechanical stability, and to carry graft material. Failure of such implants can lead to significant morbidity. We discuss the possible causes leading to the failure of a CFRP cage in a patient with rheumatoid arthritis. Review of a 49-year-old woman who underwent revision anterior lumbar interbody fusion 2 years after posterior instrumentation and transforaminal lumbar interbody fusion at L4–L5 and L5–S1. The patient developed pseudarthrosis at the two previously fused levels with failure of the posterior instrumentation. Revision surgery reveled failure with fragmentation of the CFRP cage at the L5–S1 level. CFRP implants can break if mechanical instability or nonunion occurs in the spinal segments, thus emphasizing the need for optimizing medical management and meticulous surgical technique in achieving stability. Georg Thieme Verlag KG 2013-04-15 2013-12 /pmc/articles/PMC3854590/ /pubmed/24436878 http://dx.doi.org/10.1055/s-0033-1343075 Text en © Thieme Medical Publishers |
spellingShingle | Article Sardar, Zeeshan Jarzem, Peter Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title | Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title_full | Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title_fullStr | Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title_full_unstemmed | Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title_short | Failure of a Carbon Fiber–Reinforced Polymer Implant Used for Transforaminal Lumbar Interbody Fusion |
title_sort | failure of a carbon fiber–reinforced polymer implant used for transforaminal lumbar interbody fusion |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3854590/ https://www.ncbi.nlm.nih.gov/pubmed/24436878 http://dx.doi.org/10.1055/s-0033-1343075 |
work_keys_str_mv | AT sardarzeeshan failureofacarbonfiberreinforcedpolymerimplantusedfortransforaminallumbarinterbodyfusion AT jarzempeter failureofacarbonfiberreinforcedpolymerimplantusedfortransforaminallumbarinterbodyfusion |