The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor
OBJECTIVES: The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor. MATERIAL AND METHODS: A digital photostimulable storage pho...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Stilus Optimus
2010
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3886041/ https://www.ncbi.nlm.nih.gov/pubmed/24421961 http://dx.doi.org/10.5037/jomr.2010.1105 |
_version_ | 1782478849986002944 |
---|---|
author | Vandenberghe, Bart Jacobs, Reinhilde |
author_facet | Vandenberghe, Bart Jacobs, Reinhilde |
author_sort | Vandenberghe, Bart |
collection | PubMed |
description | OBJECTIVES: The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor. MATERIAL AND METHODS: A digital photostimulable storage phosphor (PSP) sensor (Vistascan) and a multipulse X-ray generator (Prostyle Intra) with two tube voltages were used in this study. The front, premolar and molar region of two adult human cadaver skulls jaws were imaged using the X-ray tube at 63 kV and 70 kV, both at 8 mA and decreasing exposure times (160 ms, 120 ms and 80 ms). A standardized exposure protocol containing waxed occlusal keys and an aiming device ensured proper and reproducible beam alignment. Three observers assessed the digital radiographs for 31 selected periodontal bone loss sites. Radiographic measurements were compared to physical measurements (Standard). Subjective ratings of lamina dura, crater defect and furcation involvement visibility, contrast perception and bone quality were also performed. RESULTS: Multiple regression equation of the variables kV and exposure time demonstrated no significant difference for the periodontal bone level measurements (P > 0.05). In 90.3% and 96.7% of the measurements for 70 kV and 63 kV respectively, deviation was within 1 mm. The subjective ratings produced similar findings in terms of image quality for both tube voltages and the three exposure times. CONCLUSIONS: The results of the present study revealed that tube voltages of 63 kV and 70 kV provided similar accuracy and image quality for periodontal disease diagnosis. |
format | Online Article Text |
id | pubmed-3886041 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2010 |
publisher | Stilus Optimus |
record_format | MEDLINE/PubMed |
spelling | pubmed-38860412014-01-13 The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor Vandenberghe, Bart Jacobs, Reinhilde J Oral Maxillofac Res Original Paper OBJECTIVES: The purpose of the present study was to determine the measurement accuracy and subjective image quality for periodontal disease diagnosis when using two X-ray tube voltages with a digital photostimulable storage phosphor sensor. MATERIAL AND METHODS: A digital photostimulable storage phosphor (PSP) sensor (Vistascan) and a multipulse X-ray generator (Prostyle Intra) with two tube voltages were used in this study. The front, premolar and molar region of two adult human cadaver skulls jaws were imaged using the X-ray tube at 63 kV and 70 kV, both at 8 mA and decreasing exposure times (160 ms, 120 ms and 80 ms). A standardized exposure protocol containing waxed occlusal keys and an aiming device ensured proper and reproducible beam alignment. Three observers assessed the digital radiographs for 31 selected periodontal bone loss sites. Radiographic measurements were compared to physical measurements (Standard). Subjective ratings of lamina dura, crater defect and furcation involvement visibility, contrast perception and bone quality were also performed. RESULTS: Multiple regression equation of the variables kV and exposure time demonstrated no significant difference for the periodontal bone level measurements (P > 0.05). In 90.3% and 96.7% of the measurements for 70 kV and 63 kV respectively, deviation was within 1 mm. The subjective ratings produced similar findings in terms of image quality for both tube voltages and the three exposure times. CONCLUSIONS: The results of the present study revealed that tube voltages of 63 kV and 70 kV provided similar accuracy and image quality for periodontal disease diagnosis. Stilus Optimus 2010-04-01 /pmc/articles/PMC3886041/ /pubmed/24421961 http://dx.doi.org/10.5037/jomr.2010.1105 Text en Copyright © Vandenberghe B, Jacobs R. Published in the JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH (http://www.ejomr.org), 1 April 2010. http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an open-access article, first published in the JOURNAL OF ORAL & MAXILLOFACIAL RESEARCH, distributed under the terms of the Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 Unported License (http://creativecommons.org/licenses/by-nc-nd/3.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work and is properly cited. The copyright, license information and link to the original publication on http://www.ejomr.org must be included. |
spellingShingle | Original Paper Vandenberghe, Bart Jacobs, Reinhilde The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title | The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title_full | The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title_fullStr | The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title_full_unstemmed | The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title_short | The Influence of Tube Potential on Periodontal Bone Level Measurements and Subjective Image Quality Using a Digital Photostimulable Storage Phosphor Sensor |
title_sort | influence of tube potential on periodontal bone level measurements and subjective image quality using a digital photostimulable storage phosphor sensor |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3886041/ https://www.ncbi.nlm.nih.gov/pubmed/24421961 http://dx.doi.org/10.5037/jomr.2010.1105 |
work_keys_str_mv | AT vandenberghebart theinfluenceoftubepotentialonperiodontalbonelevelmeasurementsandsubjectiveimagequalityusingadigitalphotostimulablestoragephosphorsensor AT jacobsreinhilde theinfluenceoftubepotentialonperiodontalbonelevelmeasurementsandsubjectiveimagequalityusingadigitalphotostimulablestoragephosphorsensor AT vandenberghebart influenceoftubepotentialonperiodontalbonelevelmeasurementsandsubjectiveimagequalityusingadigitalphotostimulablestoragephosphorsensor AT jacobsreinhilde influenceoftubepotentialonperiodontalbonelevelmeasurementsandsubjectiveimagequalityusingadigitalphotostimulablestoragephosphorsensor |