The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory t...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3890616/ https://www.ncbi.nlm.nih.gov/pubmed/24387685 http://dx.doi.org/10.1186/1471-2253-14-2 |
_version_ | 1782299287981391872 |
---|---|
author | Bergmann, Ingo Crozier, Thomas Allen Roessler, Markus Schotola, Hanna Mansur, Ashham Büttner, Benedikt Hinz, José Maria Bauer, Martin |
author_facet | Bergmann, Ingo Crozier, Thomas Allen Roessler, Markus Schotola, Hanna Mansur, Ashham Büttner, Benedikt Hinz, José Maria Bauer, Martin |
author_sort | Bergmann, Ingo |
collection | PubMed |
description | BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory tracts and less airway morbidity if the device were first affixed and the cuff then inflated. METHODS: Our clinical review board approved the study (public registry number DRKS00003174). An LMA Supreme® was inserted into 184 patients undergoing lower limb arthroscopy in propofol-remifentanil anaesthesia who were randomly assigned to either the control (inflation then fixation; n = 92) or study group (fixation then inflation; n = 92). The cuff was inflated to 60 cmH(2)O. The patients’ lungs were ventilated in pressure-controlled mode with 5 cmH(2)O PEEP, Pmax to give 6 ml kg(-1) tidal volume, and respiratory rate adjusted to end-tidal CO(2) of 4.8 and 5.6 kPa. Correct cuff and tip position were determined by leak detection, capnometry trace, oropharyngeal leak pressure, suprasternal notch test, and lube-tube test. Bowl and cuff position and the presence of glottic narrowing were assessed by fiberscopic examination. Postoperative dysphagia, hoarseness and sore throat were assessed with a questionnaire. Ventilatory impairment was defined as a tidal volume < 6 ml kg(-1) with Pmax at oropharyngeal leak pressure, glottic narrowing was defined as an angle between the vocal cords under 16 degrees. RESULTS: The incidence of incorrect device position (18% vs. 21%), failed ventilation (10% vs. 9%), leak pressure (24.8 vs. 25.2 cmH(2)O, p = 0.63), failed lube-tube test (16.3% vs. 17.6%) and glottic narrowing (19.3% vs. 14.1%, p = 0.35) was similar in both groups (control vs. study, resp.). When glottic narrowing occurred, it was more frequently associated with ventilatory impairment in the control group (77% vs. 39%; p = 0.04). Airway morbidity was more common in the control group (33% vs. 19%; p < 0.05). CONCLUSIONS: Altering the sequence of cuff inflation and device fixation does not affect device position, oropharyngeal leak pressures or separation of gastrointestinal and respiratory tracts. It reduces the incidence of glottic narrowing with impaired ventilation and also perioperative airway morbidity. |
format | Online Article Text |
id | pubmed-3890616 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-38906162014-01-15 The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study Bergmann, Ingo Crozier, Thomas Allen Roessler, Markus Schotola, Hanna Mansur, Ashham Büttner, Benedikt Hinz, José Maria Bauer, Martin BMC Anesthesiol Research Article BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory tracts and less airway morbidity if the device were first affixed and the cuff then inflated. METHODS: Our clinical review board approved the study (public registry number DRKS00003174). An LMA Supreme® was inserted into 184 patients undergoing lower limb arthroscopy in propofol-remifentanil anaesthesia who were randomly assigned to either the control (inflation then fixation; n = 92) or study group (fixation then inflation; n = 92). The cuff was inflated to 60 cmH(2)O. The patients’ lungs were ventilated in pressure-controlled mode with 5 cmH(2)O PEEP, Pmax to give 6 ml kg(-1) tidal volume, and respiratory rate adjusted to end-tidal CO(2) of 4.8 and 5.6 kPa. Correct cuff and tip position were determined by leak detection, capnometry trace, oropharyngeal leak pressure, suprasternal notch test, and lube-tube test. Bowl and cuff position and the presence of glottic narrowing were assessed by fiberscopic examination. Postoperative dysphagia, hoarseness and sore throat were assessed with a questionnaire. Ventilatory impairment was defined as a tidal volume < 6 ml kg(-1) with Pmax at oropharyngeal leak pressure, glottic narrowing was defined as an angle between the vocal cords under 16 degrees. RESULTS: The incidence of incorrect device position (18% vs. 21%), failed ventilation (10% vs. 9%), leak pressure (24.8 vs. 25.2 cmH(2)O, p = 0.63), failed lube-tube test (16.3% vs. 17.6%) and glottic narrowing (19.3% vs. 14.1%, p = 0.35) was similar in both groups (control vs. study, resp.). When glottic narrowing occurred, it was more frequently associated with ventilatory impairment in the control group (77% vs. 39%; p = 0.04). Airway morbidity was more common in the control group (33% vs. 19%; p < 0.05). CONCLUSIONS: Altering the sequence of cuff inflation and device fixation does not affect device position, oropharyngeal leak pressures or separation of gastrointestinal and respiratory tracts. It reduces the incidence of glottic narrowing with impaired ventilation and also perioperative airway morbidity. BioMed Central 2014-01-04 /pmc/articles/PMC3890616/ /pubmed/24387685 http://dx.doi.org/10.1186/1471-2253-14-2 Text en Copyright © 2014 Bergmann et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Bergmann, Ingo Crozier, Thomas Allen Roessler, Markus Schotola, Hanna Mansur, Ashham Büttner, Benedikt Hinz, José Maria Bauer, Martin The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title | The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title_full | The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title_fullStr | The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title_full_unstemmed | The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title_short | The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
title_sort | effect of changing the sequence of cuff inflation and device fixation with the lma-supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3890616/ https://www.ncbi.nlm.nih.gov/pubmed/24387685 http://dx.doi.org/10.1186/1471-2253-14-2 |
work_keys_str_mv | AT bergmanningo theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT crozierthomasallen theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT roesslermarkus theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT schotolahanna theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT mansurashham theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT buttnerbenedikt theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT hinzjosemaria theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT bauermartin theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT bergmanningo effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT crozierthomasallen effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT roesslermarkus effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT schotolahanna effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT mansurashham effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT buttnerbenedikt effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT hinzjosemaria effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy AT bauermartin effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy |