The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study

BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory t...

Descripción completa

Detalles Bibliográficos
Autores principales: Bergmann, Ingo, Crozier, Thomas Allen, Roessler, Markus, Schotola, Hanna, Mansur, Ashham, Büttner, Benedikt, Hinz, José Maria, Bauer, Martin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3890616/
https://www.ncbi.nlm.nih.gov/pubmed/24387685
http://dx.doi.org/10.1186/1471-2253-14-2
_version_ 1782299287981391872
author Bergmann, Ingo
Crozier, Thomas Allen
Roessler, Markus
Schotola, Hanna
Mansur, Ashham
Büttner, Benedikt
Hinz, José Maria
Bauer, Martin
author_facet Bergmann, Ingo
Crozier, Thomas Allen
Roessler, Markus
Schotola, Hanna
Mansur, Ashham
Büttner, Benedikt
Hinz, José Maria
Bauer, Martin
author_sort Bergmann, Ingo
collection PubMed
description BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory tracts and less airway morbidity if the device were first affixed and the cuff then inflated. METHODS: Our clinical review board approved the study (public registry number DRKS00003174). An LMA Supreme® was inserted into 184 patients undergoing lower limb arthroscopy in propofol-remifentanil anaesthesia who were randomly assigned to either the control (inflation then fixation; n = 92) or study group (fixation then inflation; n = 92). The cuff was inflated to 60 cmH(2)O. The patients’ lungs were ventilated in pressure-controlled mode with 5 cmH(2)O PEEP, Pmax to give 6 ml kg(-1) tidal volume, and respiratory rate adjusted to end-tidal CO(2) of 4.8 and 5.6 kPa. Correct cuff and tip position were determined by leak detection, capnometry trace, oropharyngeal leak pressure, suprasternal notch test, and lube-tube test. Bowl and cuff position and the presence of glottic narrowing were assessed by fiberscopic examination. Postoperative dysphagia, hoarseness and sore throat were assessed with a questionnaire. Ventilatory impairment was defined as a tidal volume < 6 ml kg(-1) with Pmax at oropharyngeal leak pressure, glottic narrowing was defined as an angle between the vocal cords under 16 degrees. RESULTS: The incidence of incorrect device position (18% vs. 21%), failed ventilation (10% vs. 9%), leak pressure (24.8 vs. 25.2 cmH(2)O, p = 0.63), failed lube-tube test (16.3% vs. 17.6%) and glottic narrowing (19.3% vs. 14.1%, p = 0.35) was similar in both groups (control vs. study, resp.). When glottic narrowing occurred, it was more frequently associated with ventilatory impairment in the control group (77% vs. 39%; p = 0.04). Airway morbidity was more common in the control group (33% vs. 19%; p < 0.05). CONCLUSIONS: Altering the sequence of cuff inflation and device fixation does not affect device position, oropharyngeal leak pressures or separation of gastrointestinal and respiratory tracts. It reduces the incidence of glottic narrowing with impaired ventilation and also perioperative airway morbidity.
format Online
Article
Text
id pubmed-3890616
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-38906162014-01-15 The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study Bergmann, Ingo Crozier, Thomas Allen Roessler, Markus Schotola, Hanna Mansur, Ashham Büttner, Benedikt Hinz, José Maria Bauer, Martin BMC Anesthesiol Research Article BACKGROUND: The conventional sequence when using supraglottic airway devices is insertion, cuff inflation and fixation. Our hypothesis was that a tighter fit of the cuff and tip could be achieved with a consequently lower incidence of air leak, better separation of gastrointestinal and respiratory tracts and less airway morbidity if the device were first affixed and the cuff then inflated. METHODS: Our clinical review board approved the study (public registry number DRKS00003174). An LMA Supreme® was inserted into 184 patients undergoing lower limb arthroscopy in propofol-remifentanil anaesthesia who were randomly assigned to either the control (inflation then fixation; n = 92) or study group (fixation then inflation; n = 92). The cuff was inflated to 60 cmH(2)O. The patients’ lungs were ventilated in pressure-controlled mode with 5 cmH(2)O PEEP, Pmax to give 6 ml kg(-1) tidal volume, and respiratory rate adjusted to end-tidal CO(2) of 4.8 and 5.6 kPa. Correct cuff and tip position were determined by leak detection, capnometry trace, oropharyngeal leak pressure, suprasternal notch test, and lube-tube test. Bowl and cuff position and the presence of glottic narrowing were assessed by fiberscopic examination. Postoperative dysphagia, hoarseness and sore throat were assessed with a questionnaire. Ventilatory impairment was defined as a tidal volume < 6 ml kg(-1) with Pmax at oropharyngeal leak pressure, glottic narrowing was defined as an angle between the vocal cords under 16 degrees. RESULTS: The incidence of incorrect device position (18% vs. 21%), failed ventilation (10% vs. 9%), leak pressure (24.8 vs. 25.2 cmH(2)O, p = 0.63), failed lube-tube test (16.3% vs. 17.6%) and glottic narrowing (19.3% vs. 14.1%, p = 0.35) was similar in both groups (control vs. study, resp.). When glottic narrowing occurred, it was more frequently associated with ventilatory impairment in the control group (77% vs. 39%; p = 0.04). Airway morbidity was more common in the control group (33% vs. 19%; p < 0.05). CONCLUSIONS: Altering the sequence of cuff inflation and device fixation does not affect device position, oropharyngeal leak pressures or separation of gastrointestinal and respiratory tracts. It reduces the incidence of glottic narrowing with impaired ventilation and also perioperative airway morbidity. BioMed Central 2014-01-04 /pmc/articles/PMC3890616/ /pubmed/24387685 http://dx.doi.org/10.1186/1471-2253-14-2 Text en Copyright © 2014 Bergmann et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Bergmann, Ingo
Crozier, Thomas Allen
Roessler, Markus
Schotola, Hanna
Mansur, Ashham
Büttner, Benedikt
Hinz, José Maria
Bauer, Martin
The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title_full The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title_fullStr The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title_full_unstemmed The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title_short The effect of changing the sequence of cuff inflation and device fixation with the LMA-Supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
title_sort effect of changing the sequence of cuff inflation and device fixation with the lma-supreme® on device position, ventilatory complications, and airway morbidity: a clinical and fiberscopic study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3890616/
https://www.ncbi.nlm.nih.gov/pubmed/24387685
http://dx.doi.org/10.1186/1471-2253-14-2
work_keys_str_mv AT bergmanningo theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT crozierthomasallen theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT roesslermarkus theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT schotolahanna theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT mansurashham theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT buttnerbenedikt theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT hinzjosemaria theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT bauermartin theeffectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT bergmanningo effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT crozierthomasallen effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT roesslermarkus effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT schotolahanna effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT mansurashham effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT buttnerbenedikt effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT hinzjosemaria effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy
AT bauermartin effectofchangingthesequenceofcuffinflationanddevicefixationwiththelmasupremeondevicepositionventilatorycomplicationsandairwaymorbidityaclinicalandfiberscopicstudy