How accurate is image guided radiation therapy (IGRT) delivered with a micro-irradiator?

There is significant interest in delivering precisely targeted small-volume radiation treatments, in the pre-clinical setting, to study dose-volume relationships with tumor control and normal tissue damage. In this work we investigate the IGRT targeting accuracy of the XRad225Cx system from Precisio...

Descripción completa

Detalles Bibliográficos
Autores principales: Oldham, M, Newton, J, Rankine, L, Adamovics, J, Kirsch, D, Das, S
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3894105/
https://www.ncbi.nlm.nih.gov/pubmed/24454521
http://dx.doi.org/10.1088/1742-6596/444/1/012070
Descripción
Sumario:There is significant interest in delivering precisely targeted small-volume radiation treatments, in the pre-clinical setting, to study dose-volume relationships with tumor control and normal tissue damage. In this work we investigate the IGRT targeting accuracy of the XRad225Cx system from Precision x-Ray using high resolution 3D dosimetry techniques. Initial results revealed a significant targeting error of about 2.4mm. This error was reduced to within 0.5mm after the IGRT hardware and software had been recalibrated. The facility for 3D dosimetry was essential to gain a comprehensive understanding of the targeting error in 3D.