Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics

PURPOSE: The purpose of this study was to evaluate the biocompatibility, local tissue effects and performance of a synthetic long-term resorbable test mesh (TIGR(®) Matrix Surgical Mesh) compared to a non-resorbable polypropylene control mesh following implantation in a sheep model. METHODS: Full-th...

Descripción completa

Detalles Bibliográficos
Autores principales: Hjort, H., Mathisen, T., Alves, A., Clermont, G., Boutrand, J. P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer-Verlag 2011
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3895198/
https://www.ncbi.nlm.nih.gov/pubmed/21972049
http://dx.doi.org/10.1007/s10029-011-0885-y
_version_ 1782299939897868288
author Hjort, H.
Mathisen, T.
Alves, A.
Clermont, G.
Boutrand, J. P.
author_facet Hjort, H.
Mathisen, T.
Alves, A.
Clermont, G.
Boutrand, J. P.
author_sort Hjort, H.
collection PubMed
description PURPOSE: The purpose of this study was to evaluate the biocompatibility, local tissue effects and performance of a synthetic long-term resorbable test mesh (TIGR(®) Matrix Surgical Mesh) compared to a non-resorbable polypropylene control mesh following implantation in a sheep model. METHODS: Full-thickness abdominal wall defects were created in 14 sheep and subsequently repaired using test or control meshes. Sacrifices were made at 4, 9, 15, 24 and 36 months and results in terms of macroscopic observations, histology and collagen analysis are described for 4, 9, 15, 24 and 36 months. RESULTS: The overall biocompatibility was good, and equivalent in the test and control meshes while the resorbable mesh was characterized by a collagen deposition more similar to native connective tissue and an increased thickness of the integrating tissue. The control polypropylene mesh provoked a typical chronic inflammation persistent over the 36-month study period. As the resorbable test mesh gradually degraded it was replaced by a newly formed collagen matrix with an increasing ratio of collagen type I/III, indicating a continuous remodeling of the collagen towards a strong connective tissue. After 36 months, the test mesh was fully resorbed and only microscopic implant residues could be found in the tissue. CONCLUSIONS: This study suggests that the concept of a long-term resorbable mesh with time-dependent mechanical characteristics offers new possibilities for soft tissue repair and reinforcement.
format Online
Article
Text
id pubmed-3895198
institution National Center for Biotechnology Information
language English
publishDate 2011
publisher Springer-Verlag
record_format MEDLINE/PubMed
spelling pubmed-38951982014-01-22 Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics Hjort, H. Mathisen, T. Alves, A. Clermont, G. Boutrand, J. P. Hernia Original Article PURPOSE: The purpose of this study was to evaluate the biocompatibility, local tissue effects and performance of a synthetic long-term resorbable test mesh (TIGR(®) Matrix Surgical Mesh) compared to a non-resorbable polypropylene control mesh following implantation in a sheep model. METHODS: Full-thickness abdominal wall defects were created in 14 sheep and subsequently repaired using test or control meshes. Sacrifices were made at 4, 9, 15, 24 and 36 months and results in terms of macroscopic observations, histology and collagen analysis are described for 4, 9, 15, 24 and 36 months. RESULTS: The overall biocompatibility was good, and equivalent in the test and control meshes while the resorbable mesh was characterized by a collagen deposition more similar to native connective tissue and an increased thickness of the integrating tissue. The control polypropylene mesh provoked a typical chronic inflammation persistent over the 36-month study period. As the resorbable test mesh gradually degraded it was replaced by a newly formed collagen matrix with an increasing ratio of collagen type I/III, indicating a continuous remodeling of the collagen towards a strong connective tissue. After 36 months, the test mesh was fully resorbed and only microscopic implant residues could be found in the tissue. CONCLUSIONS: This study suggests that the concept of a long-term resorbable mesh with time-dependent mechanical characteristics offers new possibilities for soft tissue repair and reinforcement. Springer-Verlag 2011-10-05 2012 /pmc/articles/PMC3895198/ /pubmed/21972049 http://dx.doi.org/10.1007/s10029-011-0885-y Text en © Springer-Verlag 2011
spellingShingle Original Article
Hjort, H.
Mathisen, T.
Alves, A.
Clermont, G.
Boutrand, J. P.
Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title_full Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title_fullStr Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title_full_unstemmed Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title_short Three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
title_sort three-year results from a preclinical implantation study of a long-term resorbable surgical mesh with time-dependent mechanical characteristics
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3895198/
https://www.ncbi.nlm.nih.gov/pubmed/21972049
http://dx.doi.org/10.1007/s10029-011-0885-y
work_keys_str_mv AT hjorth threeyearresultsfromapreclinicalimplantationstudyofalongtermresorbablesurgicalmeshwithtimedependentmechanicalcharacteristics
AT mathisent threeyearresultsfromapreclinicalimplantationstudyofalongtermresorbablesurgicalmeshwithtimedependentmechanicalcharacteristics
AT alvesa threeyearresultsfromapreclinicalimplantationstudyofalongtermresorbablesurgicalmeshwithtimedependentmechanicalcharacteristics
AT clermontg threeyearresultsfromapreclinicalimplantationstudyofalongtermresorbablesurgicalmeshwithtimedependentmechanicalcharacteristics
AT boutrandjp threeyearresultsfromapreclinicalimplantationstudyofalongtermresorbablesurgicalmeshwithtimedependentmechanicalcharacteristics