Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)

BACKGROUND: Vibration response imaging (VRI) is a bedside technology to monitor ventilation by detecting lung sound vibrations. It is currently unknown whether VRI is able to accurately monitor the local distribution of ventilation within the lungs. We therefore compared VRI to electrical impedance...

Descripción completa

Detalles Bibliográficos
Autores principales: Shi, Chang, Boehme, Stefan, Bentley, Alexander H., Hartmann, Erik K., Klein, Klaus U., Bodenstein, Marc, Baumgardner, James E., David, Matthias, Ullrich, Roman, Markstaller, Klaus
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3903564/
https://www.ncbi.nlm.nih.gov/pubmed/24475160
http://dx.doi.org/10.1371/journal.pone.0086638
_version_ 1782301112356831232
author Shi, Chang
Boehme, Stefan
Bentley, Alexander H.
Hartmann, Erik K.
Klein, Klaus U.
Bodenstein, Marc
Baumgardner, James E.
David, Matthias
Ullrich, Roman
Markstaller, Klaus
author_facet Shi, Chang
Boehme, Stefan
Bentley, Alexander H.
Hartmann, Erik K.
Klein, Klaus U.
Bodenstein, Marc
Baumgardner, James E.
David, Matthias
Ullrich, Roman
Markstaller, Klaus
author_sort Shi, Chang
collection PubMed
description BACKGROUND: Vibration response imaging (VRI) is a bedside technology to monitor ventilation by detecting lung sound vibrations. It is currently unknown whether VRI is able to accurately monitor the local distribution of ventilation within the lungs. We therefore compared VRI to electrical impedance tomography (EIT), an established technique used for the assessment of regional ventilation. METHODOLOGY/PRINCIPAL FINDINGS: Simultaneous EIT and VRI measurements were performed in the healthy and injured lungs (ALI; induced by saline lavage) at different PEEP levels (0, 5, 10, 15 mbar) in nine piglets. Vibration energy amplitude (VEA) by VRI, and amplitudes of relative impedance changes (rel.ΔZ) by EIT, were evaluated in seven regions of interest (ROIs). To assess the distribution of tidal volume (V(T)) by VRI and EIT, absolute values were normalized to the V(T) obtained by simultaneous spirometry measurements. Redistribution of ventilation by ALI and PEEP was detected by VRI and EIT. The linear correlation between pooled V(T) by VEA and rel.ΔZ was R(2) = 0.96. Bland-Altman analysis showed a bias of −1.07±24.71 ml and limits of agreement of −49.05 to +47.36 ml. Within the different ROIs, correlations of V(T)-distribution by EIT and VRI ranged between R(2) values of 0.29 and 0.96. ALI and PEEP did not alter the agreement of V(T) between VRI and EIT. CONCLUSIONS/SIGNIFICANCE: Measurements of regional ventilation distribution by VRI are comparable to those obtained by EIT.
format Online
Article
Text
id pubmed-3903564
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-39035642014-01-28 Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT) Shi, Chang Boehme, Stefan Bentley, Alexander H. Hartmann, Erik K. Klein, Klaus U. Bodenstein, Marc Baumgardner, James E. David, Matthias Ullrich, Roman Markstaller, Klaus PLoS One Research Article BACKGROUND: Vibration response imaging (VRI) is a bedside technology to monitor ventilation by detecting lung sound vibrations. It is currently unknown whether VRI is able to accurately monitor the local distribution of ventilation within the lungs. We therefore compared VRI to electrical impedance tomography (EIT), an established technique used for the assessment of regional ventilation. METHODOLOGY/PRINCIPAL FINDINGS: Simultaneous EIT and VRI measurements were performed in the healthy and injured lungs (ALI; induced by saline lavage) at different PEEP levels (0, 5, 10, 15 mbar) in nine piglets. Vibration energy amplitude (VEA) by VRI, and amplitudes of relative impedance changes (rel.ΔZ) by EIT, were evaluated in seven regions of interest (ROIs). To assess the distribution of tidal volume (V(T)) by VRI and EIT, absolute values were normalized to the V(T) obtained by simultaneous spirometry measurements. Redistribution of ventilation by ALI and PEEP was detected by VRI and EIT. The linear correlation between pooled V(T) by VEA and rel.ΔZ was R(2) = 0.96. Bland-Altman analysis showed a bias of −1.07±24.71 ml and limits of agreement of −49.05 to +47.36 ml. Within the different ROIs, correlations of V(T)-distribution by EIT and VRI ranged between R(2) values of 0.29 and 0.96. ALI and PEEP did not alter the agreement of V(T) between VRI and EIT. CONCLUSIONS/SIGNIFICANCE: Measurements of regional ventilation distribution by VRI are comparable to those obtained by EIT. Public Library of Science 2014-01-27 /pmc/articles/PMC3903564/ /pubmed/24475160 http://dx.doi.org/10.1371/journal.pone.0086638 Text en © 2014 Shi et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Shi, Chang
Boehme, Stefan
Bentley, Alexander H.
Hartmann, Erik K.
Klein, Klaus U.
Bodenstein, Marc
Baumgardner, James E.
David, Matthias
Ullrich, Roman
Markstaller, Klaus
Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title_full Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title_fullStr Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title_full_unstemmed Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title_short Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
title_sort assessment of regional ventilation distribution: comparison of vibration response imaging (vri) with electrical impedance tomography (eit)
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3903564/
https://www.ncbi.nlm.nih.gov/pubmed/24475160
http://dx.doi.org/10.1371/journal.pone.0086638
work_keys_str_mv AT shichang assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT boehmestefan assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT bentleyalexanderh assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT hartmannerikk assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT kleinklausu assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT bodensteinmarc assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT baumgardnerjamese assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT davidmatthias assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT ullrichroman assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit
AT markstallerklaus assessmentofregionalventilationdistributioncomparisonofvibrationresponseimagingvriwithelectricalimpedancetomographyeit