Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory

BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validatio...

Descripción completa

Detalles Bibliográficos
Autores principales: de Bitencourt, Eberson Damião dos Santos, Voegeli, Carlos Franco, Onzi, Gabriela dos Santos, Boscato, Sara Cardoso, Ghem, Carine, Munhoz, Terezinha
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Associação Brasileira de Hematologia e Hemoterapia 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3905822/
https://www.ncbi.nlm.nih.gov/pubmed/24478606
http://dx.doi.org/10.5581/1516-8484.20130121
Descripción
Sumario:BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. OBJECTIVE: To evaluate the performance and reliability of the Sysmex(®) SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. METHODS: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex(®) SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision(®) DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. RESULTS: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. CONCLUSION: The new Sysmex(®)SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts.