Cargando…
Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory
BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validatio...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Associação Brasileira de Hematologia e Hemoterapia
2013
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3905822/ https://www.ncbi.nlm.nih.gov/pubmed/24478606 http://dx.doi.org/10.5581/1516-8484.20130121 |
_version_ | 1782301387045994496 |
---|---|
author | de Bitencourt, Eberson Damião dos Santos Voegeli, Carlos Franco Onzi, Gabriela dos Santos Boscato, Sara Cardoso Ghem, Carine Munhoz, Terezinha |
author_facet | de Bitencourt, Eberson Damião dos Santos Voegeli, Carlos Franco Onzi, Gabriela dos Santos Boscato, Sara Cardoso Ghem, Carine Munhoz, Terezinha |
author_sort | de Bitencourt, Eberson Damião dos Santos |
collection | PubMed |
description | BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. OBJECTIVE: To evaluate the performance and reliability of the Sysmex(®) SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. METHODS: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex(®) SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision(®) DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. RESULTS: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. CONCLUSION: The new Sysmex(®)SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts. |
format | Online Article Text |
id | pubmed-3905822 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2013 |
publisher | Associação Brasileira de Hematologia e Hemoterapia |
record_format | MEDLINE/PubMed |
spelling | pubmed-39058222014-01-29 Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory de Bitencourt, Eberson Damião dos Santos Voegeli, Carlos Franco Onzi, Gabriela dos Santos Boscato, Sara Cardoso Ghem, Carine Munhoz, Terezinha Rev Bras Hematol Hemoter Original Articles BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process. OBJECTIVE: To evaluate the performance and reliability of the Sysmex(®) SP-1000i slide preparer-stainer incorporated into the routine of a hospital laboratory in Porto Alegre. METHODS: Peripheral blood samples of patients attending the laboratory for ambulatory exams with leukocyte counts between 7000/°L and 12,000/°L were evaluated, independent of gender and age. Two slides were prepared for each sample using the Sysmex(®) SP-1000i equipment; one of the slides was used to perform quality control tests using the CellaVision(®) DM96 device, and the other slide was used to compare pre-classification by the same device and the classification performed by a pharmacist-biochemist. RESULTS: The results of all the slides used as controls were acceptable according to the quality control test as established by the manufacturer of the device. In the comparison between the automated pre-classification and the classification made by the professional, there was an acceptable variation in the differential counts of leukocytes for 90% of the analyzed slides. Pearson correlation coefficient showed a strong correlation for band neutrophils (r = 0.802; p-value < 0.001), segmented neutrophils (r = 0.963; p-value < 0.001), eosinophils (r = 0.958; p-value < 0.001), lymphocytes (r = 0.985; p-value < 0.001) and atypical lymphocytes (r = 0.866; p-value < 0.001) using both methods. The red blood cell analysis was adequate for all slides analyzed by the equipment and by the professional. CONCLUSION: The new Sysmex(®)SP1000-i methodology was found to be reliable, fast and safe for the routines of medium and large laboratories, improving the quality of microscopic analysis in complete blood counts. Associação Brasileira de Hematologia e Hemoterapia 2013 /pmc/articles/PMC3905822/ /pubmed/24478606 http://dx.doi.org/10.5581/1516-8484.20130121 Text en http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Articles de Bitencourt, Eberson Damião dos Santos Voegeli, Carlos Franco Onzi, Gabriela dos Santos Boscato, Sara Cardoso Ghem, Carine Munhoz, Terezinha Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory |
title | Validation of the Sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
title_full | Validation of the Sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
title_fullStr | Validation of the Sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
title_full_unstemmed | Validation of the Sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
title_short | Validation of the Sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
title_sort | validation of the sysmex sp-1000i automated slide preparer-stainer in a
clinical laboratory |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3905822/ https://www.ncbi.nlm.nih.gov/pubmed/24478606 http://dx.doi.org/10.5581/1516-8484.20130121 |
work_keys_str_mv | AT debitencourtebersondamiaodossantos validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory AT voegelicarlosfranco validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory AT onzigabrieladossantos validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory AT boscatosaracardoso validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory AT ghemcarine validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory AT munhozterezinha validationofthesysmexsp1000iautomatedslidepreparerstainerinaclinicallaboratory |