Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy

PURPOSE: To recommend the optimal plan parameter set of grid size and angular increment for dose calculations in treatment planning for lung stereotactic body radiation therapy (SBRT) using dynamic conformal arc therapy (DCAT) considering both accuracy and computational efficiency. MATERIALS AND MET...

Descripción completa

Detalles Bibliográficos
Autores principales: Park, Ji-Yeon, Kim, Siyong, Park, Hae-Jin, Lee, Jeong-Woo, Kim, Yeon-Sil, Suh, Tae-Suk
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3909323/
https://www.ncbi.nlm.nih.gov/pubmed/24387278
http://dx.doi.org/10.1186/1748-717X-9-5
_version_ 1782301829246222336
author Park, Ji-Yeon
Kim, Siyong
Park, Hae-Jin
Lee, Jeong-Woo
Kim, Yeon-Sil
Suh, Tae-Suk
author_facet Park, Ji-Yeon
Kim, Siyong
Park, Hae-Jin
Lee, Jeong-Woo
Kim, Yeon-Sil
Suh, Tae-Suk
author_sort Park, Ji-Yeon
collection PubMed
description PURPOSE: To recommend the optimal plan parameter set of grid size and angular increment for dose calculations in treatment planning for lung stereotactic body radiation therapy (SBRT) using dynamic conformal arc therapy (DCAT) considering both accuracy and computational efficiency. MATERIALS AND METHODS: Dose variations with varying grid sizes (2, 3, and 4 mm) and angular increments (2°, 4°, 6°, and 10°) were analyzed in a thorax phantom for 3 spherical target volumes and in 9 patient cases. A 2-mm grid size and 2° angular increment are assumed sufficient to serve as reference values. The dosimetric effect was evaluated using dose–volume histograms, monitor units (MUs), and dose to organs at risk (OARs) for a definite volume corresponding to the dose–volume constraint in lung SBRT. The times required for dose calculations using each parameter set were compared for clinical practicality. RESULTS: Larger grid sizes caused a dose increase to the structures and required higher MUs to achieve the target coverage. The discrete beam arrangements at each angular increment led to over- and under-estimated OARs doses due to the undulating dose distribution. When a 2° angular increment was used in both studies, a 4-mm grid size changed the dose variation by up to 3–4% (50 cGy) for the heart and the spinal cord, while a 3-mm grid size produced a dose difference of <1% (12 cGy) in all tested OARs. When a 3-mm grid size was employed, angular increments of 6° and 10° caused maximum dose variations of 3% (23 cGy) and 10% (61 cGy) in the spinal cord, respectively, while a 4° increment resulted in a dose difference of <1% (8 cGy) in all cases except for that of one patient. The 3-mm grid size and 4° angular increment enabled a 78% savings in computation time without making any critical sacrifices to dose accuracy. CONCLUSIONS: A parameter set with a 3-mm grid size and a 4° angular increment is found to be appropriate for predicting patient dose distributions with a dose difference below 1% while reducing the computation time by more than half for lung SBRT using DCAT.
format Online
Article
Text
id pubmed-3909323
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-39093232014-02-02 Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy Park, Ji-Yeon Kim, Siyong Park, Hae-Jin Lee, Jeong-Woo Kim, Yeon-Sil Suh, Tae-Suk Radiat Oncol Research PURPOSE: To recommend the optimal plan parameter set of grid size and angular increment for dose calculations in treatment planning for lung stereotactic body radiation therapy (SBRT) using dynamic conformal arc therapy (DCAT) considering both accuracy and computational efficiency. MATERIALS AND METHODS: Dose variations with varying grid sizes (2, 3, and 4 mm) and angular increments (2°, 4°, 6°, and 10°) were analyzed in a thorax phantom for 3 spherical target volumes and in 9 patient cases. A 2-mm grid size and 2° angular increment are assumed sufficient to serve as reference values. The dosimetric effect was evaluated using dose–volume histograms, monitor units (MUs), and dose to organs at risk (OARs) for a definite volume corresponding to the dose–volume constraint in lung SBRT. The times required for dose calculations using each parameter set were compared for clinical practicality. RESULTS: Larger grid sizes caused a dose increase to the structures and required higher MUs to achieve the target coverage. The discrete beam arrangements at each angular increment led to over- and under-estimated OARs doses due to the undulating dose distribution. When a 2° angular increment was used in both studies, a 4-mm grid size changed the dose variation by up to 3–4% (50 cGy) for the heart and the spinal cord, while a 3-mm grid size produced a dose difference of <1% (12 cGy) in all tested OARs. When a 3-mm grid size was employed, angular increments of 6° and 10° caused maximum dose variations of 3% (23 cGy) and 10% (61 cGy) in the spinal cord, respectively, while a 4° increment resulted in a dose difference of <1% (8 cGy) in all cases except for that of one patient. The 3-mm grid size and 4° angular increment enabled a 78% savings in computation time without making any critical sacrifices to dose accuracy. CONCLUSIONS: A parameter set with a 3-mm grid size and a 4° angular increment is found to be appropriate for predicting patient dose distributions with a dose difference below 1% while reducing the computation time by more than half for lung SBRT using DCAT. BioMed Central 2014-01-04 /pmc/articles/PMC3909323/ /pubmed/24387278 http://dx.doi.org/10.1186/1748-717X-9-5 Text en Copyright © 2014 Park et al.; licensee BioMed Central Ltd. http://creativecommons.org/licenses/by/2.0 This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Park, Ji-Yeon
Kim, Siyong
Park, Hae-Jin
Lee, Jeong-Woo
Kim, Yeon-Sil
Suh, Tae-Suk
Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title_full Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title_fullStr Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title_full_unstemmed Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title_short Optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
title_sort optimal set of grid size and angular increment for practical dose calculation using the dynamic conformal arc technique: a systematic evaluation of the dosimetric effects in lung stereotactic body radiation therapy
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3909323/
https://www.ncbi.nlm.nih.gov/pubmed/24387278
http://dx.doi.org/10.1186/1748-717X-9-5
work_keys_str_mv AT parkjiyeon optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy
AT kimsiyong optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy
AT parkhaejin optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy
AT leejeongwoo optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy
AT kimyeonsil optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy
AT suhtaesuk optimalsetofgridsizeandangularincrementforpracticaldosecalculationusingthedynamicconformalarctechniqueasystematicevaluationofthedosimetriceffectsinlungstereotacticbodyradiationtherapy