Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males,...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914879/ https://www.ncbi.nlm.nih.gov/pubmed/24505345 http://dx.doi.org/10.1371/journal.pone.0087996 |
_version_ | 1782302484741488640 |
---|---|
author | Liu, Xinting Shen, Meixiao Huang, Shenghai Leng, Lin Zhu, Dexi Lu, Fan |
author_facet | Liu, Xinting Shen, Meixiao Huang, Shenghai Leng, Lin Zhu, Dexi Lu, Fan |
author_sort | Liu, Xinting |
collection | PubMed |
description | PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males, 8 females; 24 to 32 years old) were enrolled. Imaging was performed with a custom built ultra-high resolution OCT instrument (UHR-OCT, ∼3 µm resolution) and a commercial RTVue100 OCT (∼5 µm resolution) instrument. An automated algorithm was developed to segment the macular retina into eight layers and quantitate the thickness of each layer. The right eye of each subject was imaged two times by the first examiner using each instrument to assess intra-observer repeatability and once by the second examiner to assess inter-observer reproducibility. The intraclass correlation coefficient (ICC) and coefficients of repeatability and reproducibility (COR) were analyzed to evaluate the reliability. RESULTS: The ICCs for the intra-observer repeatability and inter-observer reproducibility of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers, except the photoreceptor inner segments, which ranged from 0.051 to 0.643, and the outer segments, which ranged from 0.709 to 0.959. The CORs were less than 6.73% for the total retina and all intra-retinal layers. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. However, the ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the inner segment and outer segment layers. CONCLUSIONS: Thickness measurements of the intra-retinal layers determined by the automated algorithm are reliable when applied to images acquired by the UHR-OCT and RTVue100 instruments. |
format | Online Article Text |
id | pubmed-3914879 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-39148792014-02-06 Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments Liu, Xinting Shen, Meixiao Huang, Shenghai Leng, Lin Zhu, Dexi Lu, Fan PLoS One Research Article PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males, 8 females; 24 to 32 years old) were enrolled. Imaging was performed with a custom built ultra-high resolution OCT instrument (UHR-OCT, ∼3 µm resolution) and a commercial RTVue100 OCT (∼5 µm resolution) instrument. An automated algorithm was developed to segment the macular retina into eight layers and quantitate the thickness of each layer. The right eye of each subject was imaged two times by the first examiner using each instrument to assess intra-observer repeatability and once by the second examiner to assess inter-observer reproducibility. The intraclass correlation coefficient (ICC) and coefficients of repeatability and reproducibility (COR) were analyzed to evaluate the reliability. RESULTS: The ICCs for the intra-observer repeatability and inter-observer reproducibility of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers, except the photoreceptor inner segments, which ranged from 0.051 to 0.643, and the outer segments, which ranged from 0.709 to 0.959. The CORs were less than 6.73% for the total retina and all intra-retinal layers. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. However, the ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the inner segment and outer segment layers. CONCLUSIONS: Thickness measurements of the intra-retinal layers determined by the automated algorithm are reliable when applied to images acquired by the UHR-OCT and RTVue100 instruments. Public Library of Science 2014-02-05 /pmc/articles/PMC3914879/ /pubmed/24505345 http://dx.doi.org/10.1371/journal.pone.0087996 Text en © 2014 Liu et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Liu, Xinting Shen, Meixiao Huang, Shenghai Leng, Lin Zhu, Dexi Lu, Fan Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title | Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title_full | Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title_fullStr | Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title_full_unstemmed | Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title_short | Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments |
title_sort | repeatability and reproducibility of eight macular intra-retinal layer thicknesses determined by an automated segmentation algorithm using two sd-oct instruments |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914879/ https://www.ncbi.nlm.nih.gov/pubmed/24505345 http://dx.doi.org/10.1371/journal.pone.0087996 |
work_keys_str_mv | AT liuxinting repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments AT shenmeixiao repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments AT huangshenghai repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments AT lenglin repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments AT zhudexi repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments AT lufan repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments |