Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments

PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males,...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Xinting, Shen, Meixiao, Huang, Shenghai, Leng, Lin, Zhu, Dexi, Lu, Fan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914879/
https://www.ncbi.nlm.nih.gov/pubmed/24505345
http://dx.doi.org/10.1371/journal.pone.0087996
_version_ 1782302484741488640
author Liu, Xinting
Shen, Meixiao
Huang, Shenghai
Leng, Lin
Zhu, Dexi
Lu, Fan
author_facet Liu, Xinting
Shen, Meixiao
Huang, Shenghai
Leng, Lin
Zhu, Dexi
Lu, Fan
author_sort Liu, Xinting
collection PubMed
description PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males, 8 females; 24 to 32 years old) were enrolled. Imaging was performed with a custom built ultra-high resolution OCT instrument (UHR-OCT, ∼3 µm resolution) and a commercial RTVue100 OCT (∼5 µm resolution) instrument. An automated algorithm was developed to segment the macular retina into eight layers and quantitate the thickness of each layer. The right eye of each subject was imaged two times by the first examiner using each instrument to assess intra-observer repeatability and once by the second examiner to assess inter-observer reproducibility. The intraclass correlation coefficient (ICC) and coefficients of repeatability and reproducibility (COR) were analyzed to evaluate the reliability. RESULTS: The ICCs for the intra-observer repeatability and inter-observer reproducibility of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers, except the photoreceptor inner segments, which ranged from 0.051 to 0.643, and the outer segments, which ranged from 0.709 to 0.959. The CORs were less than 6.73% for the total retina and all intra-retinal layers. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. However, the ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the inner segment and outer segment layers. CONCLUSIONS: Thickness measurements of the intra-retinal layers determined by the automated algorithm are reliable when applied to images acquired by the UHR-OCT and RTVue100 instruments.
format Online
Article
Text
id pubmed-3914879
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-39148792014-02-06 Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments Liu, Xinting Shen, Meixiao Huang, Shenghai Leng, Lin Zhu, Dexi Lu, Fan PLoS One Research Article PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments. METHODS: Twenty normal subjects (12 males, 8 females; 24 to 32 years old) were enrolled. Imaging was performed with a custom built ultra-high resolution OCT instrument (UHR-OCT, ∼3 µm resolution) and a commercial RTVue100 OCT (∼5 µm resolution) instrument. An automated algorithm was developed to segment the macular retina into eight layers and quantitate the thickness of each layer. The right eye of each subject was imaged two times by the first examiner using each instrument to assess intra-observer repeatability and once by the second examiner to assess inter-observer reproducibility. The intraclass correlation coefficient (ICC) and coefficients of repeatability and reproducibility (COR) were analyzed to evaluate the reliability. RESULTS: The ICCs for the intra-observer repeatability and inter-observer reproducibility of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers, except the photoreceptor inner segments, which ranged from 0.051 to 0.643, and the outer segments, which ranged from 0.709 to 0.959. The CORs were less than 6.73% for the total retina and all intra-retinal layers. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. However, the ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the inner segment and outer segment layers. CONCLUSIONS: Thickness measurements of the intra-retinal layers determined by the automated algorithm are reliable when applied to images acquired by the UHR-OCT and RTVue100 instruments. Public Library of Science 2014-02-05 /pmc/articles/PMC3914879/ /pubmed/24505345 http://dx.doi.org/10.1371/journal.pone.0087996 Text en © 2014 Liu et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Liu, Xinting
Shen, Meixiao
Huang, Shenghai
Leng, Lin
Zhu, Dexi
Lu, Fan
Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title_full Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title_fullStr Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title_full_unstemmed Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title_short Repeatability and Reproducibility of Eight Macular Intra-Retinal Layer Thicknesses Determined by an Automated Segmentation Algorithm Using Two SD-OCT Instruments
title_sort repeatability and reproducibility of eight macular intra-retinal layer thicknesses determined by an automated segmentation algorithm using two sd-oct instruments
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914879/
https://www.ncbi.nlm.nih.gov/pubmed/24505345
http://dx.doi.org/10.1371/journal.pone.0087996
work_keys_str_mv AT liuxinting repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments
AT shenmeixiao repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments
AT huangshenghai repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments
AT lenglin repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments
AT zhudexi repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments
AT lufan repeatabilityandreproducibilityofeightmacularintraretinallayerthicknessesdeterminedbyanautomatedsegmentationalgorithmusingtwosdoctinstruments