Cargando…

Myocardial Defect Detection Using PET-CT: Phantom Studies

It is expected that both noise and activity distribution can have impact on the detectability of a myocardial defect in a cardiac PET study. In this work, we performed phantom studies to investigate the detectability of a defect in the myocardium for different noise levels and activity distributions...

Descripción completa

Detalles Bibliográficos
Autores principales: Mananga, Eugene S., El Fakhri, Georges, Schaefferkoetter, Joshua, Bonab, Ali A., Ouyang, Jinsong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914931/
https://www.ncbi.nlm.nih.gov/pubmed/24505429
http://dx.doi.org/10.1371/journal.pone.0088200
Descripción
Sumario:It is expected that both noise and activity distribution can have impact on the detectability of a myocardial defect in a cardiac PET study. In this work, we performed phantom studies to investigate the detectability of a defect in the myocardium for different noise levels and activity distributions. We evaluated the performance of three reconstruction schemes: Filtered Back-Projection (FBP), Ordinary Poisson Ordered Subset Expectation Maximization (OP–OSEM), and Point Spread Function corrected OSEM (PSF–OSEM). We used the Channelized Hotelling Observer (CHO) for the task of myocardial defect detection. We found that the detectability of a myocardial defect is almost entirely dependent on the noise level and the contrast between the defect and its surroundings.