Cargando…

Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations

High-energy linacs produce secondary particles such as neutrons (photoneutron production). The neutrons have the important role during treatment with high energy photons in terms of protection and dose escalation. In this work, neutron dose equivalents of 18 MV Varian and Elekta accelerators are mea...

Descripción completa

Detalles Bibliográficos
Autores principales: Nedaie, Hassan Ali, Darestani, Hoda, Banaee, Nooshin, Shagholi, Negin, Mohammadi, Kheirollah, Shahvar, Arjang, Bayat, Esmaeel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3931222/
https://www.ncbi.nlm.nih.gov/pubmed/24600167
http://dx.doi.org/10.4103/0971-6203.125476
_version_ 1782304626183241728
author Nedaie, Hassan Ali
Darestani, Hoda
Banaee, Nooshin
Shagholi, Negin
Mohammadi, Kheirollah
Shahvar, Arjang
Bayat, Esmaeel
author_facet Nedaie, Hassan Ali
Darestani, Hoda
Banaee, Nooshin
Shagholi, Negin
Mohammadi, Kheirollah
Shahvar, Arjang
Bayat, Esmaeel
author_sort Nedaie, Hassan Ali
collection PubMed
description High-energy linacs produce secondary particles such as neutrons (photoneutron production). The neutrons have the important role during treatment with high energy photons in terms of protection and dose escalation. In this work, neutron dose equivalents of 18 MV Varian and Elekta accelerators are measured by thermoluminescent dosimeter (TLD) 600 and TLD700 detectors and compared with the Monte Carlo calculations. For neutron and photon dose discrimination, first TLDs were calibrated separately by gamma and neutron doses. Gamma calibration was carried out in two procedures; by standard 60Co source and by 18 MV linac photon beam. For neutron calibration by (241)Am-Be source, irradiations were performed in several different time intervals. The Varian and Elekta linac heads and the phantom were simulated by the MCNPX code (v. 2.5). Neutron dose equivalent was calculated in the central axis, on the phantom surface and depths of 1, 2, 3.3, 4, 5, and 6 cm. The maximum photoneutron dose equivalents which calculated by the MCNPX code were 7.06 and 2.37 mSv.Gy(-1) for Varian and Elekta accelerators, respectively, in comparison with 50 and 44 mSv.Gy(-1) achieved by TLDs. All the results showed more photoneutron production in Varian accelerator compared to Elekta. According to the results, it seems that TLD600 and TLD700 pairs are not suitable dosimeters for neutron dosimetry inside the linac field due to high photon flux, while MCNPX code is an appropriate alternative for studying photoneutron production.
format Online
Article
Text
id pubmed-3931222
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-39312222014-03-05 Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations Nedaie, Hassan Ali Darestani, Hoda Banaee, Nooshin Shagholi, Negin Mohammadi, Kheirollah Shahvar, Arjang Bayat, Esmaeel J Med Phys Original Article High-energy linacs produce secondary particles such as neutrons (photoneutron production). The neutrons have the important role during treatment with high energy photons in terms of protection and dose escalation. In this work, neutron dose equivalents of 18 MV Varian and Elekta accelerators are measured by thermoluminescent dosimeter (TLD) 600 and TLD700 detectors and compared with the Monte Carlo calculations. For neutron and photon dose discrimination, first TLDs were calibrated separately by gamma and neutron doses. Gamma calibration was carried out in two procedures; by standard 60Co source and by 18 MV linac photon beam. For neutron calibration by (241)Am-Be source, irradiations were performed in several different time intervals. The Varian and Elekta linac heads and the phantom were simulated by the MCNPX code (v. 2.5). Neutron dose equivalent was calculated in the central axis, on the phantom surface and depths of 1, 2, 3.3, 4, 5, and 6 cm. The maximum photoneutron dose equivalents which calculated by the MCNPX code were 7.06 and 2.37 mSv.Gy(-1) for Varian and Elekta accelerators, respectively, in comparison with 50 and 44 mSv.Gy(-1) achieved by TLDs. All the results showed more photoneutron production in Varian accelerator compared to Elekta. According to the results, it seems that TLD600 and TLD700 pairs are not suitable dosimeters for neutron dosimetry inside the linac field due to high photon flux, while MCNPX code is an appropriate alternative for studying photoneutron production. Medknow Publications & Media Pvt Ltd 2014 /pmc/articles/PMC3931222/ /pubmed/24600167 http://dx.doi.org/10.4103/0971-6203.125476 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Nedaie, Hassan Ali
Darestani, Hoda
Banaee, Nooshin
Shagholi, Negin
Mohammadi, Kheirollah
Shahvar, Arjang
Bayat, Esmaeel
Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title_full Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title_fullStr Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title_full_unstemmed Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title_short Neutron dose measurements of Varian and Elekta linacs by TLD600 and TLD700 dosimeters and comparison with MCNP calculations
title_sort neutron dose measurements of varian and elekta linacs by tld600 and tld700 dosimeters and comparison with mcnp calculations
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3931222/
https://www.ncbi.nlm.nih.gov/pubmed/24600167
http://dx.doi.org/10.4103/0971-6203.125476
work_keys_str_mv AT nedaiehassanali neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT darestanihoda neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT banaeenooshin neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT shagholinegin neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT mohammadikheirollah neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT shahvararjang neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations
AT bayatesmaeel neutrondosemeasurementsofvarianandelektalinacsbytld600andtld700dosimetersandcomparisonwithmcnpcalculations