Correlation between microleakage and screw loosening at implant-abutment connection

PURPOSE: This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests. MATERIALS AND METHODS: Three different abutment types (Intenal hex titan...

Descripción completa

Detalles Bibliográficos
Autores principales: Sahin, Cem, Ayyildiz, Simel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Prosthodontics 2014
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3942525/
https://www.ncbi.nlm.nih.gov/pubmed/24605204
http://dx.doi.org/10.4047/jap.2014.6.1.35
Descripción
Sumario:PURPOSE: This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests. MATERIALS AND METHODS: Three different abutment types (Intenal hex titanium, internal hex zirconium, morse tapered titaniuım) with different geometries were connected to its own implant fixture. All the abutments were tightened with a standard torque value then the composition was connected to the modified fluid filtration system. After the measurements of leakage removal torque values were re-measured. Kruskal-wallis test was performed for non-parametric and one-way ANOVA was performed for parametric data. The correlation was evaluated using Spearman Correlation Test (α=0.05). RESULTS: Significantly higher microleakage was found at the connection of implant-internal hex zirconium abutment. Observed mean torque value loss was also significantly higher than other connection geometries. Spearman tests revealed a significant correlation between microleakage and screw loosening. CONCLUSION: Microleakage may provoke screw loosening. Removing torque values rationally decrease with the increase of microleakage.