Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples
INTRODUCTION: For nucleic acid testing (NAT) of blood donations, either the blood samples can be pooled together in a batch of six or eight prior to testing (mini-pool-NAT [MP-NAT]), or the tests can be run on every individual sample (individual donor-NAT [ID-NAT]). It has been debated in various st...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3943140/ https://www.ncbi.nlm.nih.gov/pubmed/24678169 http://dx.doi.org/10.4103/0973-6247.126684 |
_version_ | 1782479184299294720 |
---|---|
author | Chatterjee, Kabita Agarwal, Nitin Coshic, Poonam Borgohain, Mayuri Chakroborty, Sourit |
author_facet | Chatterjee, Kabita Agarwal, Nitin Coshic, Poonam Borgohain, Mayuri Chakroborty, Sourit |
author_sort | Chatterjee, Kabita |
collection | PubMed |
description | INTRODUCTION: For nucleic acid testing (NAT) of blood donations, either the blood samples can be pooled together in a batch of six or eight prior to testing (mini-pool-NAT [MP-NAT]), or the tests can be run on every individual sample (individual donor-NAT [ID-NAT]). It has been debated in various studies whether pooling of samples results in decreased sensitivity of detection as the volume of individual samples gets lesser in a pool. The objective of this study was to investigate the effect of dilution on the sensitivity of tests. MATERIALS AND METHODS: The study was performesd on nine plasma samples which were hepatitis B reactive exclusively by Procleix Ultrio Plus and not by Procleix Ultrio or serology. These nine exclusive UltrioPlus ID-NAT yield samples were diluted in 1:2, 1:4. 1:6 and 1:8 dilutions using previously tested negative plasma and each dilution of every sample along with archived undiluted sample were retested in three replicates with Procleix Ultrio Plus Assay. RESULTS: Among NAT yield samples, 88.88% of the samples were detected when retested in ID-NAT in undiluted form. Samples with higher viral load (sample 5 and 6) were detected by all dilutions. When samples with viral load below 20 IU/mL were tested in dilutions of 1:6 or 1:8, only 9 out of 27 replicates (33.33%) were detected. This means that more than 67% of low viral load samples were missed by MP-NAT of 1:6 or 1:8 dilution out of total NAT yield samples. CONCLUSION: Individual Donor NAT is ideal methodology for NAT as dilution due to pooling may miss samples with low viral load as evident in this study. |
format | Online Article Text |
id | pubmed-3943140 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-39431402014-03-27 Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples Chatterjee, Kabita Agarwal, Nitin Coshic, Poonam Borgohain, Mayuri Chakroborty, Sourit Asian J Transfus Sci Original Article INTRODUCTION: For nucleic acid testing (NAT) of blood donations, either the blood samples can be pooled together in a batch of six or eight prior to testing (mini-pool-NAT [MP-NAT]), or the tests can be run on every individual sample (individual donor-NAT [ID-NAT]). It has been debated in various studies whether pooling of samples results in decreased sensitivity of detection as the volume of individual samples gets lesser in a pool. The objective of this study was to investigate the effect of dilution on the sensitivity of tests. MATERIALS AND METHODS: The study was performesd on nine plasma samples which were hepatitis B reactive exclusively by Procleix Ultrio Plus and not by Procleix Ultrio or serology. These nine exclusive UltrioPlus ID-NAT yield samples were diluted in 1:2, 1:4. 1:6 and 1:8 dilutions using previously tested negative plasma and each dilution of every sample along with archived undiluted sample were retested in three replicates with Procleix Ultrio Plus Assay. RESULTS: Among NAT yield samples, 88.88% of the samples were detected when retested in ID-NAT in undiluted form. Samples with higher viral load (sample 5 and 6) were detected by all dilutions. When samples with viral load below 20 IU/mL were tested in dilutions of 1:6 or 1:8, only 9 out of 27 replicates (33.33%) were detected. This means that more than 67% of low viral load samples were missed by MP-NAT of 1:6 or 1:8 dilution out of total NAT yield samples. CONCLUSION: Individual Donor NAT is ideal methodology for NAT as dilution due to pooling may miss samples with low viral load as evident in this study. Medknow Publications & Media Pvt Ltd 2014 /pmc/articles/PMC3943140/ /pubmed/24678169 http://dx.doi.org/10.4103/0973-6247.126684 Text en Copyright: © Asian Journal of Transfusion Science http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Chatterjee, Kabita Agarwal, Nitin Coshic, Poonam Borgohain, Mayuri Chakroborty, Sourit Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title | Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title_full | Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title_fullStr | Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title_full_unstemmed | Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title_short | Sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis B nucleic acid testing yield samples |
title_sort | sensitivity of individual and mini-pool nucleic acid testing assessed by dilution of hepatitis b nucleic acid testing yield samples |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3943140/ https://www.ncbi.nlm.nih.gov/pubmed/24678169 http://dx.doi.org/10.4103/0973-6247.126684 |
work_keys_str_mv | AT chatterjeekabita sensitivityofindividualandminipoolnucleicacidtestingassessedbydilutionofhepatitisbnucleicacidtestingyieldsamples AT agarwalnitin sensitivityofindividualandminipoolnucleicacidtestingassessedbydilutionofhepatitisbnucleicacidtestingyieldsamples AT coshicpoonam sensitivityofindividualandminipoolnucleicacidtestingassessedbydilutionofhepatitisbnucleicacidtestingyieldsamples AT borgohainmayuri sensitivityofindividualandminipoolnucleicacidtestingassessedbydilutionofhepatitisbnucleicacidtestingyieldsamples AT chakrobortysourit sensitivityofindividualandminipoolnucleicacidtestingassessedbydilutionofhepatitisbnucleicacidtestingyieldsamples |