Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subject...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Academy of Rehabilitation Medicine
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3953369/ https://www.ncbi.nlm.nih.gov/pubmed/24639931 http://dx.doi.org/10.5535/arm.2014.38.1.84 |
_version_ | 1782307341715111936 |
---|---|
author | Kim, Won Kuel Kim, Don-Kyu Seo, Kyung Mook Kang, Si Hyun |
author_facet | Kim, Won Kuel Kim, Don-Kyu Seo, Kyung Mook Kang, Si Hyun |
author_sort | Kim, Won Kuel |
collection | PubMed |
description | OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subjects were tested in seated and supine position using three measurement methods: isometric knee extension by isokinetic dynamometer, non-fixed HHD, and fixed HHD. During the measurement, the knee joints of subjects were fixed at a 35° angle from the extended position. The fixed HHD measurement was conducted with the HHD fixed to distal tibia with a Velcro strap; non-fixed HHD was performed with a hand-held method without Velcro fixation. All the measurements were repeated three times and among them, the maximum values of peak torque were used for the analysis. RESULTS: The data from the fixed HHD method showed higher validity than the non-fixed method compared with the results of the isokinetic dynamometer. Pearson correlation coefficients (r) between fixed HHD and isokinetic dynamometer method were statistically significant (supine-right: r=0.806, p<0.05; seating-right: r=0.473, p<0.05; supine-left: r=0.524, p<0.05), whereas Pearson correlation coefficients between non-fixed dynamometer and isokinetic dynamometer methods were not statistically significant, except for the result of the supine position of the left leg (r=0.384, p<0.05). Both fixed and non-fixed HHD methods showed excellent inter-rater reliability. However, the fixed HHD method showed a higher reliability than the non-fixed HHD method by considering the intraclass correlation coefficient (fixed HHD, 0.952-0.984; non-fixed HHD, 0.940-0.963). CONCLUSION: Fixation of HHD during measurement in the supine position increases the reliability and validity in measuring the quadriceps strength. |
format | Online Article Text |
id | pubmed-3953369 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Korean Academy of Rehabilitation Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-39533692014-03-17 Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study Kim, Won Kuel Kim, Don-Kyu Seo, Kyung Mook Kang, Si Hyun Ann Rehabil Med Original Article OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subjects were tested in seated and supine position using three measurement methods: isometric knee extension by isokinetic dynamometer, non-fixed HHD, and fixed HHD. During the measurement, the knee joints of subjects were fixed at a 35° angle from the extended position. The fixed HHD measurement was conducted with the HHD fixed to distal tibia with a Velcro strap; non-fixed HHD was performed with a hand-held method without Velcro fixation. All the measurements were repeated three times and among them, the maximum values of peak torque were used for the analysis. RESULTS: The data from the fixed HHD method showed higher validity than the non-fixed method compared with the results of the isokinetic dynamometer. Pearson correlation coefficients (r) between fixed HHD and isokinetic dynamometer method were statistically significant (supine-right: r=0.806, p<0.05; seating-right: r=0.473, p<0.05; supine-left: r=0.524, p<0.05), whereas Pearson correlation coefficients between non-fixed dynamometer and isokinetic dynamometer methods were not statistically significant, except for the result of the supine position of the left leg (r=0.384, p<0.05). Both fixed and non-fixed HHD methods showed excellent inter-rater reliability. However, the fixed HHD method showed a higher reliability than the non-fixed HHD method by considering the intraclass correlation coefficient (fixed HHD, 0.952-0.984; non-fixed HHD, 0.940-0.963). CONCLUSION: Fixation of HHD during measurement in the supine position increases the reliability and validity in measuring the quadriceps strength. Korean Academy of Rehabilitation Medicine 2014-02 2014-02-25 /pmc/articles/PMC3953369/ /pubmed/24639931 http://dx.doi.org/10.5535/arm.2014.38.1.84 Text en Copyright © 2014 by Korean Academy of Rehabilitation Medicine http://creativecommons.org/licenses/by-nc/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Kim, Won Kuel Kim, Don-Kyu Seo, Kyung Mook Kang, Si Hyun Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title | Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title_full | Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title_fullStr | Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title_full_unstemmed | Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title_short | Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study |
title_sort | reliability and validity of isometric knee extensor strength test with hand-held dynamometer depending on its fixation: a pilot study |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3953369/ https://www.ncbi.nlm.nih.gov/pubmed/24639931 http://dx.doi.org/10.5535/arm.2014.38.1.84 |
work_keys_str_mv | AT kimwonkuel reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy AT kimdonkyu reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy AT seokyungmook reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy AT kangsihyun reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy |