Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study

OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subject...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Won Kuel, Kim, Don-Kyu, Seo, Kyung Mook, Kang, Si Hyun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Academy of Rehabilitation Medicine 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3953369/
https://www.ncbi.nlm.nih.gov/pubmed/24639931
http://dx.doi.org/10.5535/arm.2014.38.1.84
_version_ 1782307341715111936
author Kim, Won Kuel
Kim, Don-Kyu
Seo, Kyung Mook
Kang, Si Hyun
author_facet Kim, Won Kuel
Kim, Don-Kyu
Seo, Kyung Mook
Kang, Si Hyun
author_sort Kim, Won Kuel
collection PubMed
description OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subjects were tested in seated and supine position using three measurement methods: isometric knee extension by isokinetic dynamometer, non-fixed HHD, and fixed HHD. During the measurement, the knee joints of subjects were fixed at a 35° angle from the extended position. The fixed HHD measurement was conducted with the HHD fixed to distal tibia with a Velcro strap; non-fixed HHD was performed with a hand-held method without Velcro fixation. All the measurements were repeated three times and among them, the maximum values of peak torque were used for the analysis. RESULTS: The data from the fixed HHD method showed higher validity than the non-fixed method compared with the results of the isokinetic dynamometer. Pearson correlation coefficients (r) between fixed HHD and isokinetic dynamometer method were statistically significant (supine-right: r=0.806, p<0.05; seating-right: r=0.473, p<0.05; supine-left: r=0.524, p<0.05), whereas Pearson correlation coefficients between non-fixed dynamometer and isokinetic dynamometer methods were not statistically significant, except for the result of the supine position of the left leg (r=0.384, p<0.05). Both fixed and non-fixed HHD methods showed excellent inter-rater reliability. However, the fixed HHD method showed a higher reliability than the non-fixed HHD method by considering the intraclass correlation coefficient (fixed HHD, 0.952-0.984; non-fixed HHD, 0.940-0.963). CONCLUSION: Fixation of HHD during measurement in the supine position increases the reliability and validity in measuring the quadriceps strength.
format Online
Article
Text
id pubmed-3953369
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Korean Academy of Rehabilitation Medicine
record_format MEDLINE/PubMed
spelling pubmed-39533692014-03-17 Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study Kim, Won Kuel Kim, Don-Kyu Seo, Kyung Mook Kang, Si Hyun Ann Rehabil Med Original Article OBJECTIVE: To determine the reliability and validity of hand-held dynamometer (HHD) depending on its fixation in measuring isometric knee extensor strength by comparing the results with an isokinetic dynamometer. METHODS: Twenty-seven healthy female volunteers participated in this study. The subjects were tested in seated and supine position using three measurement methods: isometric knee extension by isokinetic dynamometer, non-fixed HHD, and fixed HHD. During the measurement, the knee joints of subjects were fixed at a 35° angle from the extended position. The fixed HHD measurement was conducted with the HHD fixed to distal tibia with a Velcro strap; non-fixed HHD was performed with a hand-held method without Velcro fixation. All the measurements were repeated three times and among them, the maximum values of peak torque were used for the analysis. RESULTS: The data from the fixed HHD method showed higher validity than the non-fixed method compared with the results of the isokinetic dynamometer. Pearson correlation coefficients (r) between fixed HHD and isokinetic dynamometer method were statistically significant (supine-right: r=0.806, p<0.05; seating-right: r=0.473, p<0.05; supine-left: r=0.524, p<0.05), whereas Pearson correlation coefficients between non-fixed dynamometer and isokinetic dynamometer methods were not statistically significant, except for the result of the supine position of the left leg (r=0.384, p<0.05). Both fixed and non-fixed HHD methods showed excellent inter-rater reliability. However, the fixed HHD method showed a higher reliability than the non-fixed HHD method by considering the intraclass correlation coefficient (fixed HHD, 0.952-0.984; non-fixed HHD, 0.940-0.963). CONCLUSION: Fixation of HHD during measurement in the supine position increases the reliability and validity in measuring the quadriceps strength. Korean Academy of Rehabilitation Medicine 2014-02 2014-02-25 /pmc/articles/PMC3953369/ /pubmed/24639931 http://dx.doi.org/10.5535/arm.2014.38.1.84 Text en Copyright © 2014 by Korean Academy of Rehabilitation Medicine http://creativecommons.org/licenses/by-nc/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kim, Won Kuel
Kim, Don-Kyu
Seo, Kyung Mook
Kang, Si Hyun
Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title_full Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title_fullStr Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title_full_unstemmed Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title_short Reliability and Validity of Isometric Knee Extensor Strength Test With Hand-Held Dynamometer Depending on Its Fixation: A Pilot Study
title_sort reliability and validity of isometric knee extensor strength test with hand-held dynamometer depending on its fixation: a pilot study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3953369/
https://www.ncbi.nlm.nih.gov/pubmed/24639931
http://dx.doi.org/10.5535/arm.2014.38.1.84
work_keys_str_mv AT kimwonkuel reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy
AT kimdonkyu reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy
AT seokyungmook reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy
AT kangsihyun reliabilityandvalidityofisometrickneeextensorstrengthtestwithhandhelddynamometerdependingonitsfixationapilotstudy