The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants

BACKGROUND: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of...

Descripción completa

Detalles Bibliográficos
Autores principales: Fayaz, Ali, Mahshid, Minoo, Saboury, Aboulfazl, Sadr, Seyed Jalil, Ansari, Ghassem
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3955319/
https://www.ncbi.nlm.nih.gov/pubmed/24688564
_version_ 1782307555770368000
author Fayaz, Ali
Mahshid, Minoo
Saboury, Aboulfazl
Sadr, Seyed Jalil
Ansari, Ghassem
author_facet Fayaz, Ali
Mahshid, Minoo
Saboury, Aboulfazl
Sadr, Seyed Jalil
Ansari, Ghassem
author_sort Fayaz, Ali
collection PubMed
description BACKGROUND: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. MATERIALS AND METHODS: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). RESULTS: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). CONCLUSION: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.
format Online
Article
Text
id pubmed-3955319
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-39553192014-03-31 The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants Fayaz, Ali Mahshid, Minoo Saboury, Aboulfazl Sadr, Seyed Jalil Ansari, Ghassem Dent Res J (Isfahan) Original Article BACKGROUND: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. MATERIALS AND METHODS: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). RESULTS: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). CONCLUSION: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles. Medknow Publications & Media Pvt Ltd 2014 /pmc/articles/PMC3955319/ /pubmed/24688564 Text en Copyright: © Dental Research Journal http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Fayaz, Ali
Mahshid, Minoo
Saboury, Aboulfazl
Sadr, Seyed Jalil
Ansari, Ghassem
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_full The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_fullStr The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_full_unstemmed The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_short The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_sort effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3955319/
https://www.ncbi.nlm.nih.gov/pubmed/24688564
work_keys_str_mv AT fayazali theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT mahshidminoo theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT sabouryaboulfazl theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT sadrseyedjalil theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT ansarighassem theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT fayazali effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT mahshidminoo effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT sabouryaboulfazl effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT sadrseyedjalil effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT ansarighassem effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants