Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency

OBJECTIVE: To determine the in vivo efficiency of monopolar radiofrequency ablation (RFA) using a dual-switching (DS) system and a separable clustered (SC) electrode to create coagulation in swine liver. MATERIALS AND METHODS: Thirty-three ablation zones were created in nine pigs using a DS system a...

Descripción completa

Detalles Bibliográficos
Autores principales: Yoon, Jeong Hee, Lee, Jeong Min, Hwang, Eui Jin, Hwang, In Pyung, Baek, Jeehyun, Han, Joon Koo, Choi, Byung Ihn
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Radiology 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3955790/
https://www.ncbi.nlm.nih.gov/pubmed/24643383
http://dx.doi.org/10.3348/kjr.2014.15.2.235
_version_ 1782307624216166400
author Yoon, Jeong Hee
Lee, Jeong Min
Hwang, Eui Jin
Hwang, In Pyung
Baek, Jeehyun
Han, Joon Koo
Choi, Byung Ihn
author_facet Yoon, Jeong Hee
Lee, Jeong Min
Hwang, Eui Jin
Hwang, In Pyung
Baek, Jeehyun
Han, Joon Koo
Choi, Byung Ihn
author_sort Yoon, Jeong Hee
collection PubMed
description OBJECTIVE: To determine the in vivo efficiency of monopolar radiofrequency ablation (RFA) using a dual-switching (DS) system and a separable clustered (SC) electrode to create coagulation in swine liver. MATERIALS AND METHODS: Thirty-three ablation zones were created in nine pigs using a DS system and an SC electrode in the switching monopolar mode. The pigs were divided into two groups for two experiments: 1) preliminary experiments (n = 3) to identify the optimal inter-electrode distances (IEDs) for dual-switching monopolar (DSM)-RFA, and 2) main experiments (n = 6) to compare the in vivo efficiency of DSM-RFA with that of a single-switching monopolar (SSM)-RFA. RF energy was alternatively applied to one of the three electrodes (SSM-RFA) or concurrently applied to a pair of electrodes (DSM-RFA) for 12 minutes in in vivo porcine livers. The delivered RFA energy and the shapes and dimensions of the coagulation areas were compared between the two groups. RESULTS: No pig died during RFA. The ideal IEDs for creating round or oval coagulation area using the DSM-RFA were 2.0 and 2.5 cm. DSM-RFA allowed more efficient RF energy delivery than SSM-RFA at the given time (23.0 ± 4.0 kcal vs. 16.92 ± 2.0 kcal, respectively; p = 0.0005). DSM-RFA created a significantly larger coagulation volume than SSM-RFA (40.4 ± 16.4 cm(3) vs. 20.8 ± 10.7 cm(3); p < 0.001). Both groups showed similar circularity of the ablation zones (p = 0.29). CONCLUSION: Dual-switching monopolar-radiofrequency ablation using an SC electrode is feasible and can create larger ablation zones than SSM-RFA as it allows more RF energy delivery at a given time.
format Online
Article
Text
id pubmed-3955790
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher The Korean Society of Radiology
record_format MEDLINE/PubMed
spelling pubmed-39557902014-03-18 Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency Yoon, Jeong Hee Lee, Jeong Min Hwang, Eui Jin Hwang, In Pyung Baek, Jeehyun Han, Joon Koo Choi, Byung Ihn Korean J Radiol Gastrointestinal Imaging OBJECTIVE: To determine the in vivo efficiency of monopolar radiofrequency ablation (RFA) using a dual-switching (DS) system and a separable clustered (SC) electrode to create coagulation in swine liver. MATERIALS AND METHODS: Thirty-three ablation zones were created in nine pigs using a DS system and an SC electrode in the switching monopolar mode. The pigs were divided into two groups for two experiments: 1) preliminary experiments (n = 3) to identify the optimal inter-electrode distances (IEDs) for dual-switching monopolar (DSM)-RFA, and 2) main experiments (n = 6) to compare the in vivo efficiency of DSM-RFA with that of a single-switching monopolar (SSM)-RFA. RF energy was alternatively applied to one of the three electrodes (SSM-RFA) or concurrently applied to a pair of electrodes (DSM-RFA) for 12 minutes in in vivo porcine livers. The delivered RFA energy and the shapes and dimensions of the coagulation areas were compared between the two groups. RESULTS: No pig died during RFA. The ideal IEDs for creating round or oval coagulation area using the DSM-RFA were 2.0 and 2.5 cm. DSM-RFA allowed more efficient RF energy delivery than SSM-RFA at the given time (23.0 ± 4.0 kcal vs. 16.92 ± 2.0 kcal, respectively; p = 0.0005). DSM-RFA created a significantly larger coagulation volume than SSM-RFA (40.4 ± 16.4 cm(3) vs. 20.8 ± 10.7 cm(3); p < 0.001). Both groups showed similar circularity of the ablation zones (p = 0.29). CONCLUSION: Dual-switching monopolar-radiofrequency ablation using an SC electrode is feasible and can create larger ablation zones than SSM-RFA as it allows more RF energy delivery at a given time. The Korean Society of Radiology 2014 2014-03-07 /pmc/articles/PMC3955790/ /pubmed/24643383 http://dx.doi.org/10.3348/kjr.2014.15.2.235 Text en Copyright © 2014 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Gastrointestinal Imaging
Yoon, Jeong Hee
Lee, Jeong Min
Hwang, Eui Jin
Hwang, In Pyung
Baek, Jeehyun
Han, Joon Koo
Choi, Byung Ihn
Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title_full Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title_fullStr Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title_full_unstemmed Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title_short Monopolar Radiofrequency Ablation Using a Dual-Switching System and a Separable Clustered Electrode: Evaluation of the In Vivo Efficiency
title_sort monopolar radiofrequency ablation using a dual-switching system and a separable clustered electrode: evaluation of the in vivo efficiency
topic Gastrointestinal Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3955790/
https://www.ncbi.nlm.nih.gov/pubmed/24643383
http://dx.doi.org/10.3348/kjr.2014.15.2.235
work_keys_str_mv AT yoonjeonghee monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT leejeongmin monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT hwangeuijin monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT hwanginpyung monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT baekjeehyun monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT hanjoonkoo monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency
AT choibyungihn monopolarradiofrequencyablationusingadualswitchingsystemandaseparableclusteredelectrodeevaluationoftheinvivoefficiency