Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report

Here we report a patient in whom there was a myopic shift after combined cataract surgery and pars plana vitrectomy against the epiretinal membrane, related to axial measurement estimation error caused by a hidden double-peak appearance on partial coherence interferometry measurement. A 52-year-old...

Descripción completa

Detalles Bibliográficos
Autores principales: Kitaguchi, Yoshiyuki, Yano, Shinsaku, Gomi, Fumi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3958499/
https://www.ncbi.nlm.nih.gov/pubmed/24648720
http://dx.doi.org/10.2147/OPTH.S59309
_version_ 1782307883185078272
author Kitaguchi, Yoshiyuki
Yano, Shinsaku
Gomi, Fumi
author_facet Kitaguchi, Yoshiyuki
Yano, Shinsaku
Gomi, Fumi
author_sort Kitaguchi, Yoshiyuki
collection PubMed
description Here we report a patient in whom there was a myopic shift after combined cataract surgery and pars plana vitrectomy against the epiretinal membrane, related to axial measurement estimation error caused by a hidden double-peak appearance on partial coherence interferometry measurement. A 52-year-old female presented with epiretinal membrane and underwent combined cataract surgery and pars plana vitrectomy. Axial length was measured with partial coherence interferometry. Although the signal curve in the summary display showed a single peak, a 1.6 diopter myopic shift occurred. Viewed retrospectively, six of 20 individual signal curves showed a double peak. Most of them showed a higher anterior peak, with only one having a higher posterior peak. The other 14 curves showed a single peak at a similar distance to an anterior peak. The anterior peak appeared to be derived from the epiretinal membrane. The possibility of a double peak should always be considered in patients with epiretinal membrane even if the summary display of the partial coherence interferometry measurement shows a single peak. Checking all signal curves would reduce the risk of missing a hidden double peak.
format Online
Article
Text
id pubmed-3958499
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-39584992014-03-19 Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report Kitaguchi, Yoshiyuki Yano, Shinsaku Gomi, Fumi Clin Ophthalmol Case Report Here we report a patient in whom there was a myopic shift after combined cataract surgery and pars plana vitrectomy against the epiretinal membrane, related to axial measurement estimation error caused by a hidden double-peak appearance on partial coherence interferometry measurement. A 52-year-old female presented with epiretinal membrane and underwent combined cataract surgery and pars plana vitrectomy. Axial length was measured with partial coherence interferometry. Although the signal curve in the summary display showed a single peak, a 1.6 diopter myopic shift occurred. Viewed retrospectively, six of 20 individual signal curves showed a double peak. Most of them showed a higher anterior peak, with only one having a higher posterior peak. The other 14 curves showed a single peak at a similar distance to an anterior peak. The anterior peak appeared to be derived from the epiretinal membrane. The possibility of a double peak should always be considered in patients with epiretinal membrane even if the summary display of the partial coherence interferometry measurement shows a single peak. Checking all signal curves would reduce the risk of missing a hidden double peak. Dove Medical Press 2014-03-13 /pmc/articles/PMC3958499/ /pubmed/24648720 http://dx.doi.org/10.2147/OPTH.S59309 Text en © 2014 Kitaguchi et al. This work is published by Dove Medical Press Limited, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Case Report
Kitaguchi, Yoshiyuki
Yano, Shinsaku
Gomi, Fumi
Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title_full Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title_fullStr Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title_full_unstemmed Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title_short Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
title_sort axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3958499/
https://www.ncbi.nlm.nih.gov/pubmed/24648720
http://dx.doi.org/10.2147/OPTH.S59309
work_keys_str_mv AT kitaguchiyoshiyuki axiallengthestimationerrorcausedbyhiddendoublepeakonpartialcoherenceinterferometryinaneyewithepiretinalmembraneacasereport
AT yanoshinsaku axiallengthestimationerrorcausedbyhiddendoublepeakonpartialcoherenceinterferometryinaneyewithepiretinalmembraneacasereport
AT gomifumi axiallengthestimationerrorcausedbyhiddendoublepeakonpartialcoherenceinterferometryinaneyewithepiretinalmembraneacasereport