A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions

Gafchromic films are used as dosimeter for in vivo and in phantom dose measurements. The dose response of Gafchromic EBT2 film under single and repeated exposure conditions is compared in this study to analyze the usability of Gafchromic EBT2 films in cumulative dose measurements. The post-irradiati...

Descripción completa

Detalles Bibliográficos
Autores principales: Ganapathy, K., Kurup, P.G.G., Murali, V., Muthukumaran, M., Velmurugan, J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3958996/
https://www.ncbi.nlm.nih.gov/pubmed/24672151
http://dx.doi.org/10.4103/0971-6203.121194
_version_ 1782307980092375040
author Ganapathy, K.
Kurup, P.G.G.
Murali, V.
Muthukumaran, M.
Velmurugan, J.
author_facet Ganapathy, K.
Kurup, P.G.G.
Murali, V.
Muthukumaran, M.
Velmurugan, J.
author_sort Ganapathy, K.
collection PubMed
description Gafchromic films are used as dosimeter for in vivo and in phantom dose measurements. The dose response of Gafchromic EBT2 film under single and repeated exposure conditions is compared in this study to analyze the usability of Gafchromic EBT2 films in cumulative dose measurements. The post-irradiation change in response of the film is studied for up to 4 days after irradiation. The effect of repeated exposure to scanner light on the response of the film is also studied. To check usability of Gafchromic EBT2 films in cumulative dose measurements, three EBT2 films were exposed to a daily fraction dose of 100 cGy, 150 cGy and 200 cGy, respectively, for 4 days. The dose response of the films exposed to cumulative irradiation was compared with the dose measured from films exposed to the same dose but in a single exposure. It is observed that the post-irradiation darkening of the film does not saturate and continue to take place even 4 days after irradiation. The dose measured from the EBT2 films after 4 days from irradiation was around 2% higher than the dose measured from the same films at 24 hours post-irradiation. It was also observed that the repeated exposure to scanner light does not produce any significant change in the film response. The dose response of films exposed to cumulative irradiation agrees with the dose response of films exposed to the same dose in a single irradiation with less than 3% difference. Gafchromic EBT2 films can be used to measure the cumulative dose delivered over multiple fractions, when the delivered dose is uniform across the film.
format Online
Article
Text
id pubmed-3958996
institution National Center for Biotechnology Information
language English
publishDate 2013
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-39589962014-03-26 A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions Ganapathy, K. Kurup, P.G.G. Murali, V. Muthukumaran, M. Velmurugan, J. J Med Phys Original Article Gafchromic films are used as dosimeter for in vivo and in phantom dose measurements. The dose response of Gafchromic EBT2 film under single and repeated exposure conditions is compared in this study to analyze the usability of Gafchromic EBT2 films in cumulative dose measurements. The post-irradiation change in response of the film is studied for up to 4 days after irradiation. The effect of repeated exposure to scanner light on the response of the film is also studied. To check usability of Gafchromic EBT2 films in cumulative dose measurements, three EBT2 films were exposed to a daily fraction dose of 100 cGy, 150 cGy and 200 cGy, respectively, for 4 days. The dose response of the films exposed to cumulative irradiation was compared with the dose measured from films exposed to the same dose but in a single exposure. It is observed that the post-irradiation darkening of the film does not saturate and continue to take place even 4 days after irradiation. The dose measured from the EBT2 films after 4 days from irradiation was around 2% higher than the dose measured from the same films at 24 hours post-irradiation. It was also observed that the repeated exposure to scanner light does not produce any significant change in the film response. The dose response of films exposed to cumulative irradiation agrees with the dose response of films exposed to the same dose in a single irradiation with less than 3% difference. Gafchromic EBT2 films can be used to measure the cumulative dose delivered over multiple fractions, when the delivered dose is uniform across the film. Medknow Publications & Media Pvt Ltd 2013 /pmc/articles/PMC3958996/ /pubmed/24672151 http://dx.doi.org/10.4103/0971-6203.121194 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Ganapathy, K.
Kurup, P.G.G.
Murali, V.
Muthukumaran, M.
Velmurugan, J.
A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title_full A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title_fullStr A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title_full_unstemmed A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title_short A study on comparison of Gafchromic EBT2 film response under single and cumulative exposure conditions
title_sort study on comparison of gafchromic ebt2 film response under single and cumulative exposure conditions
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3958996/
https://www.ncbi.nlm.nih.gov/pubmed/24672151
http://dx.doi.org/10.4103/0971-6203.121194
work_keys_str_mv AT ganapathyk astudyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT kuruppgg astudyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT muraliv astudyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT muthukumaranm astudyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT velmuruganj astudyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT ganapathyk studyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT kuruppgg studyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT muraliv studyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT muthukumaranm studyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions
AT velmuruganj studyoncomparisonofgafchromicebt2filmresponseundersingleandcumulativeexposureconditions