Arterial pressure: agreement between a brachial cuff-based device and radial tonometry

OBJECTIVES: Aortic (central) blood pressure (BP) differs from brachial BP and may be a superior predictor of cardiovascular events. However, its measurement is currently restricted to research settings, owing to a moderate level of operator dependency. We tested a new noninvasive device in a large U...

Descripción completa

Detalles Bibliográficos
Autores principales: Park, Chloe M., Korolkova, Olga, Davies, Justin E., Parker, Kim H., Siggers, Jennifer H., March, Katherine, Tillin, Therese, Chaturvedi, Nish, Hughes, Alun D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3966921/
https://www.ncbi.nlm.nih.gov/pubmed/24379000
http://dx.doi.org/10.1097/HJH.0000000000000082
Descripción
Sumario:OBJECTIVES: Aortic (central) blood pressure (BP) differs from brachial BP and may be a superior predictor of cardiovascular events. However, its measurement is currently restricted to research settings, owing to a moderate level of operator dependency. We tested a new noninvasive device in a large UK cohort. The device estimates central BP using measurements obtained with an upper arm cuff inflated to suprasystolic pressure. We compared these estimates with those obtained using radial tonometry as well as with invasively acquired measurements of aortic BP in a limited number of individuals. METHODS: Consecutive cuff-based and tonometry-based estimates of the pressure waveform and the central BP were obtained from 1107 individuals (70 ± 6 years). Short-term and long-term reproducibility studies were performed on 28 individuals. Simultaneous cuff-based and invasively measured pressure traces were acquired and compared in an additional six individuals (65 ± 20 years). RESULTS: Central systolic BP, as estimated by the cuff-based device, was found to be highly reproducible (coefficient of variation 4 and 8% for short and long-term reproducibility, respectively) and was comparable to that estimated by tonometry (average difference 3 ± 6 mmHg, intraclass correlation coefficient = 0.91). The cuff-based pressure waveforms were similar to those acquired invasively (cross-correlation coefficient 0.93), and the difference in the estimated central systolic BP was −5 ± 8 mmHg (P = 0.2). CONCLUSION: Cuff-based devices show promise to simplify the measurement of central BP, whilst maintaining a similar fidelity to tonometry. This could lead to improved adoption of estimates of central BP in clinical practice.