Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections

PURPOSE: Patient-specific dosimetry of lutetium-177 ((177)Lu)-DOTATATE treatment in neuroendocrine tumours is important, because uptake differs across patients. Single photon emission computer tomography (SPECT)-based dosimetry requires a conversion factor between the obtained counts and the activit...

Descripción completa

Detalles Bibliográficos
Autores principales: de Nijs, Robin, Lagerburg, Vera, Klausen, Thomas L., Holm, Søren
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3969156/
https://www.ncbi.nlm.nih.gov/pubmed/24525900
http://dx.doi.org/10.1097/MNM.0000000000000079
_version_ 1782309235874332672
author de Nijs, Robin
Lagerburg, Vera
Klausen, Thomas L.
Holm, Søren
author_facet de Nijs, Robin
Lagerburg, Vera
Klausen, Thomas L.
Holm, Søren
author_sort de Nijs, Robin
collection PubMed
description PURPOSE: Patient-specific dosimetry of lutetium-177 ((177)Lu)-DOTATATE treatment in neuroendocrine tumours is important, because uptake differs across patients. Single photon emission computer tomography (SPECT)-based dosimetry requires a conversion factor between the obtained counts and the activity, which depends on the collimator type, the utilized energy windows and the applied scatter correction techniques. In this study, energy window subtraction-based scatter correction methods are compared experimentally and quantitatively. MATERIALS AND METHODS: (177)Lu SPECT images of a phantom with known activity concentration ratio between the uniform background and filled hollow spheres were acquired for three different collimators: low-energy high resolution (LEHR), low-energy general purpose (LEGP) and medium-energy general purpose (MEGP). Counts were collected in several energy windows, and scatter correction was performed by applying different methods such as effective scatter source estimation (ESSE), triple-energy and dual-energy window, double-photopeak window and downscatter correction. The intensity ratio between the spheres and the background was measured and corrected for the partial volume effect and used to compare the performance of the methods. RESULTS: Low-energy collimators combined with 208 keV energy windows give rise to artefacts. For the 113 keV energy window, large differences were observed in the ratios for the spheres. For MEGP collimators with the ESSE correction technique, the measured ratio was close to the real ratio, and the differences between spheres were small. CONCLUSION: For quantitative (177)Lu imaging MEGP collimators are advised. Both energy peaks can be utilized when the ESSE correction technique is applied. The difference between the calculated and the real ratio is less than 10% for both energy windows.
format Online
Article
Text
id pubmed-3969156
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-39691562014-03-31 Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections de Nijs, Robin Lagerburg, Vera Klausen, Thomas L. Holm, Søren Nucl Med Commun Original Articles PURPOSE: Patient-specific dosimetry of lutetium-177 ((177)Lu)-DOTATATE treatment in neuroendocrine tumours is important, because uptake differs across patients. Single photon emission computer tomography (SPECT)-based dosimetry requires a conversion factor between the obtained counts and the activity, which depends on the collimator type, the utilized energy windows and the applied scatter correction techniques. In this study, energy window subtraction-based scatter correction methods are compared experimentally and quantitatively. MATERIALS AND METHODS: (177)Lu SPECT images of a phantom with known activity concentration ratio between the uniform background and filled hollow spheres were acquired for three different collimators: low-energy high resolution (LEHR), low-energy general purpose (LEGP) and medium-energy general purpose (MEGP). Counts were collected in several energy windows, and scatter correction was performed by applying different methods such as effective scatter source estimation (ESSE), triple-energy and dual-energy window, double-photopeak window and downscatter correction. The intensity ratio between the spheres and the background was measured and corrected for the partial volume effect and used to compare the performance of the methods. RESULTS: Low-energy collimators combined with 208 keV energy windows give rise to artefacts. For the 113 keV energy window, large differences were observed in the ratios for the spheres. For MEGP collimators with the ESSE correction technique, the measured ratio was close to the real ratio, and the differences between spheres were small. CONCLUSION: For quantitative (177)Lu imaging MEGP collimators are advised. Both energy peaks can be utilized when the ESSE correction technique is applied. The difference between the calculated and the real ratio is less than 10% for both energy windows. Lippincott Williams & Wilkins 2014-05 2014-04-03 /pmc/articles/PMC3969156/ /pubmed/24525900 http://dx.doi.org/10.1097/MNM.0000000000000079 Text en © 2014 Wolters Kluwer Health | Lippincott Williams & Wilkins http://creativecommons.org/licenses/by-nc-nd/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivitives 3.0 License, where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially.
spellingShingle Original Articles
de Nijs, Robin
Lagerburg, Vera
Klausen, Thomas L.
Holm, Søren
Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title_full Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title_fullStr Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title_full_unstemmed Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title_short Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections
title_sort improving quantitative dosimetry in (177)lu-dotatate spect by energy window-based scatter corrections
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3969156/
https://www.ncbi.nlm.nih.gov/pubmed/24525900
http://dx.doi.org/10.1097/MNM.0000000000000079
work_keys_str_mv AT denijsrobin improvingquantitativedosimetryin177ludotatatespectbyenergywindowbasedscattercorrections
AT lagerburgvera improvingquantitativedosimetryin177ludotatatespectbyenergywindowbasedscattercorrections
AT klausenthomasl improvingquantitativedosimetryin177ludotatatespectbyenergywindowbasedscattercorrections
AT holmsøren improvingquantitativedosimetryin177ludotatatespectbyenergywindowbasedscattercorrections