A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in th...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi Publishing Corporation
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3971571/ https://www.ncbi.nlm.nih.gov/pubmed/24716061 http://dx.doi.org/10.1155/2014/209461 |
_version_ | 1782309494934470656 |
---|---|
author | Sakamoto, Masaaki Watanabe, Hitoshi Higashi, Hidetaka Kubosawa, Hitoshi |
author_facet | Sakamoto, Masaaki Watanabe, Hitoshi Higashi, Hidetaka Kubosawa, Hitoshi |
author_sort | Sakamoto, Masaaki |
collection | PubMed |
description | A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion. Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed. Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself. Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD. |
format | Online Article Text |
id | pubmed-3971571 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Hindawi Publishing Corporation |
record_format | MEDLINE/PubMed |
spelling | pubmed-39715712014-04-08 Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty Sakamoto, Masaaki Watanabe, Hitoshi Higashi, Hidetaka Kubosawa, Hitoshi Case Rep Orthop Case Report A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion. Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed. Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself. Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD. Hindawi Publishing Corporation 2014 2014-03-04 /pmc/articles/PMC3971571/ /pubmed/24716061 http://dx.doi.org/10.1155/2014/209461 Text en Copyright © 2014 Masaaki Sakamoto et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Sakamoto, Masaaki Watanabe, Hitoshi Higashi, Hidetaka Kubosawa, Hitoshi Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title | Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title_full | Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title_fullStr | Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title_full_unstemmed | Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title_short | Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty |
title_sort | titanium alloy stem as a cause for adverse reaction to metal debris after bipolar hemiarthroplasty |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3971571/ https://www.ncbi.nlm.nih.gov/pubmed/24716061 http://dx.doi.org/10.1155/2014/209461 |
work_keys_str_mv | AT sakamotomasaaki titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty AT watanabehitoshi titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty AT higashihidetaka titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty AT kubosawahitoshi titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty |