Cargando…

Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty

A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in th...

Descripción completa

Detalles Bibliográficos
Autores principales: Sakamoto, Masaaki, Watanabe, Hitoshi, Higashi, Hidetaka, Kubosawa, Hitoshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3971571/
https://www.ncbi.nlm.nih.gov/pubmed/24716061
http://dx.doi.org/10.1155/2014/209461
_version_ 1782309494934470656
author Sakamoto, Masaaki
Watanabe, Hitoshi
Higashi, Hidetaka
Kubosawa, Hitoshi
author_facet Sakamoto, Masaaki
Watanabe, Hitoshi
Higashi, Hidetaka
Kubosawa, Hitoshi
author_sort Sakamoto, Masaaki
collection PubMed
description A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion. Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed. Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself. Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD.
format Online
Article
Text
id pubmed-3971571
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-39715712014-04-08 Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty Sakamoto, Masaaki Watanabe, Hitoshi Higashi, Hidetaka Kubosawa, Hitoshi Case Rep Orthop Case Report A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion. Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed. Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself. Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD. Hindawi Publishing Corporation 2014 2014-03-04 /pmc/articles/PMC3971571/ /pubmed/24716061 http://dx.doi.org/10.1155/2014/209461 Text en Copyright © 2014 Masaaki Sakamoto et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Sakamoto, Masaaki
Watanabe, Hitoshi
Higashi, Hidetaka
Kubosawa, Hitoshi
Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title_full Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title_fullStr Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title_full_unstemmed Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title_short Titanium Alloy Stem as a Cause for Adverse Reaction to Metal Debris after Bipolar Hemiarthroplasty
title_sort titanium alloy stem as a cause for adverse reaction to metal debris after bipolar hemiarthroplasty
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3971571/
https://www.ncbi.nlm.nih.gov/pubmed/24716061
http://dx.doi.org/10.1155/2014/209461
work_keys_str_mv AT sakamotomasaaki titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty
AT watanabehitoshi titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty
AT higashihidetaka titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty
AT kubosawahitoshi titaniumalloystemasacauseforadversereactiontometaldebrisafterbipolarhemiarthroplasty