Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure
AIM OF THE STUDY: In many countries, Low Level Disinfection (LLD) of covered transvaginal ultrasound probes is recommended between patients' examinations. The aim of this study was to evaluate the antimicrobial efficacy of LLD under routine conditions on a range of microorganisms. MATERIALS AND...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3973690/ https://www.ncbi.nlm.nih.gov/pubmed/24695371 http://dx.doi.org/10.1371/journal.pone.0093368 |
_version_ | 1782479360196870144 |
---|---|
author | M'Zali, Fatima Bounizra, Carole Leroy, Sandrine Mekki, Yahia Quentin-Noury, Claudine Kann, Michael |
author_facet | M'Zali, Fatima Bounizra, Carole Leroy, Sandrine Mekki, Yahia Quentin-Noury, Claudine Kann, Michael |
author_sort | M'Zali, Fatima |
collection | PubMed |
description | AIM OF THE STUDY: In many countries, Low Level Disinfection (LLD) of covered transvaginal ultrasound probes is recommended between patients' examinations. The aim of this study was to evaluate the antimicrobial efficacy of LLD under routine conditions on a range of microorganisms. MATERIALS AND METHODS: Samples were taken over a six month period in a private French Radiology Center. 300 specimens derived from endovaginal ultrasound probes were analyzed after disinfection of the probe with wipes impregnated with a quaternary ammonium compound and chlorhexidine. Human papillomavirus (HPV) was sought in the first set of s100 samples, Chlamydia trachomatis and mycoplasmas were searched in the second set of 100 samples, bacteria and fungi in the third 100 set samples. HPV, C. trachomatis and mycoplasmas were detected by PCR amplification. PCR positive samples were subjected to a nuclease treatment before an additional PCR assay to assess the likely viable microorganisms. Bacteria and fungi were investigated by conventional methods. RESULTS: A substantial persistence of microorganisms was observed on the disinfected probes: HPV DNA was found on 13% of the samples and 7% in nuclease-resistant form. C. trachomatis DNA was detected on 20% of the probes by primary PCR but only 2% after nuclease treatment, while mycoplasma DNA was amplified in 8% and 4%, respectively. Commensal and/or environmental bacterial flora was present on 86% of the probes, occasionally in mixed culture, and at various levels (10->3000 CFU/probe); Staphylococcus aureus was cultured from 4% of the probes (10-560 CFU/probe). No fungi were isolated. CONCLUSION: Our findings raise concerns about the efficacy of impregnated towels as a sole mean for disinfection of ultrasound probes. Although the ultrasound probes are used with disposable covers, our results highlight the potential risk of cross contamination between patients during ultrasound examination and emphasize the need for reviewing the disinfection procedure. |
format | Online Article Text |
id | pubmed-3973690 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-39736902014-04-04 Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure M'Zali, Fatima Bounizra, Carole Leroy, Sandrine Mekki, Yahia Quentin-Noury, Claudine Kann, Michael PLoS One Research Article AIM OF THE STUDY: In many countries, Low Level Disinfection (LLD) of covered transvaginal ultrasound probes is recommended between patients' examinations. The aim of this study was to evaluate the antimicrobial efficacy of LLD under routine conditions on a range of microorganisms. MATERIALS AND METHODS: Samples were taken over a six month period in a private French Radiology Center. 300 specimens derived from endovaginal ultrasound probes were analyzed after disinfection of the probe with wipes impregnated with a quaternary ammonium compound and chlorhexidine. Human papillomavirus (HPV) was sought in the first set of s100 samples, Chlamydia trachomatis and mycoplasmas were searched in the second set of 100 samples, bacteria and fungi in the third 100 set samples. HPV, C. trachomatis and mycoplasmas were detected by PCR amplification. PCR positive samples were subjected to a nuclease treatment before an additional PCR assay to assess the likely viable microorganisms. Bacteria and fungi were investigated by conventional methods. RESULTS: A substantial persistence of microorganisms was observed on the disinfected probes: HPV DNA was found on 13% of the samples and 7% in nuclease-resistant form. C. trachomatis DNA was detected on 20% of the probes by primary PCR but only 2% after nuclease treatment, while mycoplasma DNA was amplified in 8% and 4%, respectively. Commensal and/or environmental bacterial flora was present on 86% of the probes, occasionally in mixed culture, and at various levels (10->3000 CFU/probe); Staphylococcus aureus was cultured from 4% of the probes (10-560 CFU/probe). No fungi were isolated. CONCLUSION: Our findings raise concerns about the efficacy of impregnated towels as a sole mean for disinfection of ultrasound probes. Although the ultrasound probes are used with disposable covers, our results highlight the potential risk of cross contamination between patients during ultrasound examination and emphasize the need for reviewing the disinfection procedure. Public Library of Science 2014-04-02 /pmc/articles/PMC3973690/ /pubmed/24695371 http://dx.doi.org/10.1371/journal.pone.0093368 Text en © 2014 M'Zali et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article M'Zali, Fatima Bounizra, Carole Leroy, Sandrine Mekki, Yahia Quentin-Noury, Claudine Kann, Michael Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title | Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title_full | Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title_fullStr | Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title_full_unstemmed | Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title_short | Persistence of Microbial Contamination on Transvaginal Ultrasound Probes despite Low-Level Disinfection Procedure |
title_sort | persistence of microbial contamination on transvaginal ultrasound probes despite low-level disinfection procedure |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3973690/ https://www.ncbi.nlm.nih.gov/pubmed/24695371 http://dx.doi.org/10.1371/journal.pone.0093368 |
work_keys_str_mv | AT mzalifatima persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure AT bounizracarole persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure AT leroysandrine persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure AT mekkiyahia persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure AT quentinnouryclaudine persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure AT kannmichael persistenceofmicrobialcontaminationontransvaginalultrasoundprobesdespitelowleveldisinfectionprocedure |