Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study
PURPOSE: To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults. SUBJECTS AND METHODS:...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3974695/ https://www.ncbi.nlm.nih.gov/pubmed/24729678 http://dx.doi.org/10.2147/OPTH.S56135 |
_version_ | 1782479501799718912 |
---|---|
author | Heiner, Peter Ligabue, Edoardo Fan, Alex Lam, Dennis |
author_facet | Heiner, Peter Ligabue, Edoardo Fan, Alex Lam, Dennis |
author_sort | Heiner, Peter |
collection | PubMed |
description | PURPOSE: To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults. SUBJECTS AND METHODS: This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure. RESULTS: In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery. CONCLUSION: This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit. |
format | Online Article Text |
id | pubmed-3974695 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-39746952014-04-11 Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study Heiner, Peter Ligabue, Edoardo Fan, Alex Lam, Dennis Clin Ophthalmol Original Research PURPOSE: To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults. SUBJECTS AND METHODS: This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure. RESULTS: In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery. CONCLUSION: This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit. Dove Medical Press 2014-03-27 /pmc/articles/PMC3974695/ /pubmed/24729678 http://dx.doi.org/10.2147/OPTH.S56135 Text en © 2014 Heiner et al. This work is published by Dove Medical Press Ltd, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Ltd, provided the work is properly attributed. |
spellingShingle | Original Research Heiner, Peter Ligabue, Edoardo Fan, Alex Lam, Dennis Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title | Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title_full | Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title_fullStr | Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title_full_unstemmed | Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title_short | Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study |
title_sort | safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (envista®) – results of a european and asian-pacific study |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3974695/ https://www.ncbi.nlm.nih.gov/pubmed/24729678 http://dx.doi.org/10.2147/OPTH.S56135 |
work_keys_str_mv | AT heinerpeter safetyandeffectivenessofasinglepiecehydrophobicacrylicintraocularlensenvistaresultsofaeuropeanandasianpacificstudy AT ligabueedoardo safetyandeffectivenessofasinglepiecehydrophobicacrylicintraocularlensenvistaresultsofaeuropeanandasianpacificstudy AT fanalex safetyandeffectivenessofasinglepiecehydrophobicacrylicintraocularlensenvistaresultsofaeuropeanandasianpacificstudy AT lamdennis safetyandeffectivenessofasinglepiecehydrophobicacrylicintraocularlensenvistaresultsofaeuropeanandasianpacificstudy |